Skip to main content
. Author manuscript; available in PMC: 2024 Sep 17.
Published in final edited form as: J Nucl Cardiol. 2018 Nov 7;27(2):562–572. doi: 10.1007/s12350-018-01505-x

Table 6.

The mean and standard-deviation values of global motion scores (SMS), global thickening scores (STS), and ejection fraction (EF) obtained with FBP and OS-EM reconstruction

100% Dose 50% Dose 25% Dose 12.5% Dose Personalized

FBP
 SMS 5.42 ± 8.63 5.55 ± 8.59 5.81 ± 8.48 7.66 ± 9.27 * 5.44 ± 8.56
 STS 3.50 ±6.16 3.68 ± 6.19 4.26 ± 6.54 5.69 ± 7.16 * 3.51 ± 6.13
 EF 51.22 ± 12.26 51.35 ± 12.27 51.54 ± 12.24 50.32 ± 12.39 * 51.30 ± 12.30
OS-EM
 SMS 5.89 ± 9.39 6.15 ± 9.59 6.51 ± 9.69 7.93 ± 10.59 * 5.92 ± 9.44
 STS 3.92 ± 6.24 4.37 ± 6.57 4.79 ± 6.66 6.04 ± 7.64 * 4.12 ± 6.48
 EF 49.31 ± 11.88 49.39 ± 11.97 49.95 ± 12.10 50.31 ± 13.68 * 49.35 ± 11.95

The asterisk (*) signifies a statistically significant difference from full dose at level < 0.001 (paired t test)