Skip to main content
. 2024 Sep 2;10(17):e37299. doi: 10.1016/j.heliyon.2024.e37299

Table 4.

Known LVEF values for the 4D-XCAT models and calculated LVEF values obtained with the four commercial software programs: (AN: Alfa Nuclear; SM: Siemens, GE-X; General Electric Xeleris; M-TT: Mediso Tera-Tomo) presented as mean ± SD (n = 64).

LVEF 4D-XCAT AN SM GE-X M-TT
Mean (%) 53.7 50.8 50.7 48.9 52.4
Standard Deviation (%) 13.0 14.5 15.3 15.8 15.5