Table 1.
Group 1 |
Group 2 - Sequential scan |
Group 3 - Helical scan |
|||||||
---|---|---|---|---|---|---|---|---|---|
CT Philips | CT GE | T-test | CT Philips | CT GE | T-test | CT Philips | CT GE | T-test | |
Scan parameters | |||||||||
Tube voltage (kV) | 120 | 120 | 120 | 120 | |||||
Tube current | 378 mAs (432 mA) | 350 mA | 423 mAs (299 mA) | 80-300 (smart mA) | |||||
Slice thickness (mm) | 1.25 | 1.25 | 1 | 1.25 | |||||
Rotation time (s) | 0.75 | 1 | 0.4 | 0.5 | |||||
Pitch |
– | – | 0.399 | 0.516:1 | |||||
Strength of IR | Level 3 | 60 % | Level 4 | 50 % | |||||
Number of patients | 100 | 100 | 50 | 50 | 20 | 20 | |||
ED (mSv) | |||||||||
min | 1.76 | 1.21 | 1.76 | 1.33 | 2.46 | 1.66 | |||
max | 2.84 | 2.14 | 2.84 | 2.14 | 2.84 | 2.06 | |||
mean (SD) | 2.09 (0.31) | 1.74 (0.28) | p < 0.01a | 1.99 (0.17) | 1.98 (0.15) | p=0.15 | 2.63 (0.11) | 1.81 (0.15) | p < 0.01a |
median | 1.89 | 1.76 | 2.01 | 1.95 | 2.64 | 1.79 | |||
DLP (mGy.cm) | |||||||||
min | 840.2 | 574.52 | 840.2 | 631.48 | 1172 | 656.61 | |||
max | 1353 | 1017.66 | 1353 | 1017.7 | 1353 | 979.63 | |||
mean(SD) | 993.28 (146.18) | 828.14 (130.06) | p < 0.01a | 949.42 (80.87) | 942.13 (73.05) | p=0.64 | 1252.2 (52.11) | 859.66 (72.04) | p < 0.01a |
median | 900.55 | 839.43 | 959.1 | 928.56 | 1255 | 850.24 | |||
third quartile | 1019.75 | 928.52 | 960 | 1017.57 | 1291.75 | 901.41 | |||
CTDIvol(mGy) | |||||||||
min | 54.6 | 29.27 | 56.2 | 35.08 | 54.6 | 32.91 | |||
max | 71.9 | 59.96 | 71.9 | 59.96 | 54.6 | 41.12 | |||
mean (SD) | 57.61 (2.89) | 43.66 (11.15) | p < 0.01a | 58.63 (3.33) | 57.25 (3.4) | p < 0.05a | 54.6 (0.0) | 37.18 (2.52) | p < 0.01a |
median | 57.4 | 37.8 | 58 | 58.03 | 54.6 | 37.78 | |||
third quartile | 58.6 | 56.54 | 58.93 | 58.03 | 54.6 | 39.29 |
Statistically significant.