Commercially available keratorefractive lenticule extraction (KLEx) platforms.

Parameters SMILE CLEAR SILK SmartSight
Femtosecond Laser Platform ZEISS VISUMAX 500 Ziemer Z8 Elita SCHWIND ATOS
Approval CE/FDA CE CE CE
Possible refractive cuts Flap/Lenticule Flap/Lenticule/Arcuate Lenticule Flap/Lenticule
Refractive correction range −0.5 to −10 DS
0 to -5DC
−0.5 to −10 DS
0 to -5DC
−0.5 to −12 DS
0 to -6DC
−0.5 to −12 DS
0 to -6DC
FS Energy (Nanojoules) 110–150 <100 40–50 75–135
Patient Interface Machine fixed Handheld Machine fixed Machine fixed
Automated Pupil Detection Absent Present Present Present
Cyclotorsion Compensation Absent Present Present Present