Skip to main content
. 2024 Sep 19;15:224. doi: 10.1186/s13244-024-01797-3

Table 4.

Comparison of SNRs and CNRs from three methods with mDIXON coronary MRA

Conventional CS DL-CS
Mean SNR 58.88 ± 16.98 59.24 ± 14.71 76.06 ± 15.93a,b
 RCA 62.42 ± 20.38 61.73 ± 16.50 79.65 ± 18.77a,b
 LAD 56.19 ± 14.89 57.48 ± 17.08 73.40 ± 17.07a,b
 LCX 58.02 ± 20.92 58.51 ± 15.77 75.12 ± 17.88a,b
Mean CNR 17.61 ± 7.71 19.00 ± 7.87 29.70 ± 8.67a,b
 RCA 21.10 ± 11.31 22.36 ± 10.34 34.61 ± 12.95a,b
 LAD 15.85 ± 10.13 16.91 ± 10.90 26.02 ± 11.38a,b
 LCX 15.88 ± 11.51 17.73 ± 9.68 28.48 ± 11.37a,b

CNR contrast-to-noise ratio, CS compressed sensing, DL deep learning, LAD left anterior descending artery, LCX left circumflex, RCA right coronary artery, SNR signal-to-noise ratio

a DL-CS vs conventional, p < 0.001

b DL-CS vs CS, p < 0.001