Abstract
Background
The presence of contamination and microorganisms at any stage of processing renders a method unsafe, leading to a high risk of cross-transmission and cross-infection.
Objective
The objective of this study was to assess the cleaning quality of aspirator instruments used in neurosurgical procedures.
Methods
The experimental study was conducted at the materials and sterilization center, as well as the microbiology laboratory, of a philanthropic hospital in Brazil. A study protocol was implemented, which involved the analysis of 10 samples of Yasargil aspirators with varying dimensions. The samples were subjected to protein tests to detect the presence of organic matter and microbiological analysis. Descriptive statistics were used to analyze the data.
Results
The results indicated that 40% of the instruments tested positive for protein after manual cleaning. Furthermore, after automated cleaning, samples showed an increased microbiological load, with Escherichia coli accounting for 20% and Klebsiella aerogenes for 10% of the identified microorganisms.
Conclusion
This study provides evidence of failures in the cleaning process of healthcare products and highlights the presence of biofilm in the pipes, thereby compromising the drinking water quality standard.
Keywords: surgical instruments, sterilization; neurosurgery, patient safety, water microbiology
Background
Technological advances in equipment and surgical instruments have become a reality for health professionals working in the Materials and Sterilization Center (MSC). This evolution requires periodic training, an adequate physical structure, and well-established routines to maintain healthcare quality.
The mission of the MSC is to supply Health Products (HPs) to surgical, healthcare, and diagnostic services, transforming highly critical contaminated materials into products suitable for processing (Kothekar and Kulkarni, 2020). This method involves several interconnected steps that are fundamental to ensure safe patient care: cleaning, preparation, disinfection/sterilization, storage, and distribution (Rutala and Weber, 2019).
In this study, we highlight the cleaning step, defined as the removal of both organic and inorganic contamination and the reduction of the microbial load present in the HPs, using water, detergents, cleaning products, and accessories, through mechanical action. This process acts on internal and external surfaces to make the product safe for handling and prepared for disinfection or sterilization (Anvisa, 2012; Croke, 2020).
It is noteworthy that the cleaning methods used by health professionals, whether manual or automated, should follow a Standard Operating Procedure (SOP) to ensure proper conditions for the use of contamination-free HPs (Anvisa, 2012). However, the presence of contamination and microorganisms at any stage of processing makes the method unsafe, contributing to the high risk of cross-transmission that promotes Surgical Site Infections (SSIs) (Croke, 2020; Kothekar and Kulkarni, 2020).
In this context, several chemical tests are available to validate the cleaning process. The adenosine triphosphate (ATP), protein tests, and other chemical indicators, commonly associated with a visual inspection, are noteworthy. Complementary microbiological analysis contributes to detecting microorganisms present in HPs. However, current national legislation does not mandate the implementation of these practices to reduce SSIs (Alvim et al., 2019; Anvisa, 2012; Rutala and Weber, 2019).
The scope of this study is to assess the quality of cleaning of HPs in sterilization centers, especially the suction instruments used in neurosurgical procedures. This focus is justifiable because these instruments are considered critical and complex, with internal spaces that are difficult to access for direct cleaning. These instruments are used in surgical procedures on sterile or decontaminated tissues, where failures in manual and/or automated cleaning can compromise patient safety (Patel et al., 2019).
Thus, infections in neurosurgery negatively impact morbidity and mortality, prolong hospital stays, and increase care costs (Jiménez-Martínez et al., 2019; Patel et al., 2019). The literature shows that the incidence of SSIs in neurosurgical procedures ranges from 1.8% to 15.3%, necessitating preventive measures and well-established processes in the MSC. However, research is still developing to verify the quality of suction instrument cleaning. Therefore, investigating this step using chemical tests and microbiological analysis becomes fundamental and necessary, as it is the purpose of this study (Jiménez-Martínez et al., 2019).
The objective is to verify the quality of cleaning of suction instruments used in neurosurgical procedures.
Methods
A pilot study at the laboratory level was conducted in a medium-sized philanthropic hospital in Brazil. All methodological steps were guided by the guidelines described in the Standards for Quality Improvement Reporting Excellence (SQUIRE 2.0) (Ogrinc et al., 2015).
The health service is responsible for the care of several medical specialties, including emergency care, a surgical center, an inpatient unit, and intensive care. The institution was inaugurated more than a century ago and has undergone several renovations to adapt the physical structure to the current health norms and legislation.
The MSC of the study site comprises ten healthcare professionals, including one nurse who supervises, eight nursing technicians, and a receptionist. The main activities performed in the sector are described in 39 operational procedures and three systemic protocols, which are available for electronic consultation and in printed copies. The equipment installed in this sector was acquired less than a year before conducting this study, and underwent validated calibration.
The study included ten neurosurgical procedures performed from January to March 2020, representing 100% of the volume of surgeries performed by the specialty. In the institution, selective surgical procedures were suspended due to the COVID-19 pandemic starting in April of the same year.
Protocol of study
Different sizes of Yasargil aspirators, in boxes and trays, were randomly selected for this study because they are considered critical Health Products (HPs) of complex conformation. These aspirators feature lumens smaller than five millimeters or have blind bottoms, internal spaces inaccessible to direct friction, recesses, or valves (Anvisa, 2012). The sector coordination provided the surgical chart and checked it the day before the procedure.
A total of 10 Yasargil samples were selected for the assessment of cleaning quality. Data collection was performed by the researchers themselves in three stages for each sample obtained from the Yasargil aspirator: (a) in the operating room, immediately after the completion of the procedure, following verbal and written authorization from the surgeon; (b) at the purge location of the MSC study, where surgical instruments are cleaned manually; and (c) at the same purge location, following automated cleaning using a Baumer® ultrasonic washer, equipped with connection points for lumen cleaning, validated using the W110® effectiveness test. At each stage, a swab was used to perform the protein test, and lavage samples were collected for microbiological analysis.
The presence of significant organic matter was detected using the STERIS Resi-Test™ swab protein test. This indicator was read qualitatively, considering the positive or negative result on the surface of the selected sample, to identify possible failures in the manual and/or automated cleaning process. The swab was rubbed on the internal surface of the suction instruments, areas not touched by hands, thus reducing contamination bias caused by skin microorganisms.
Lavage samples were collected using the flush method, wherein 20 mL of sterile bidistilled water was aseptically injected and deposited into a sterile vial. Immediately after collection, the samples were transported to the institutional laboratory. The sterile vials underwent sonication and agitation for five seconds at a frequency of 42 kilohertz. A calibrated loop was used to replicate the samples, with a dilution of 1/100, onto MacConkey dual medium blood plates.
The plates were then incubated at 37°C, under room atmosphere, for 24 to 48 hours to facilitate bacterial growth. The isolated colonies were subjected to Gram staining and biochemical tests for microbial classification. The Vitek® automation system was utilized for the identification of genus and species. Following cultivation and bacterial growth, the number of colonies was counted and multiplied by 100 to quantify the results in terms of Colony Forming Units (CFU/ml).
Data were transcribed to a form containing the following study variables: surgical procedure; surgery time; type of sample; dimensions of the suction instruments; and the results of the samples, considering the result of the protein test (positive or negative) and the microbiological analysis.
The information was entered into a spreadsheet and analyzed using Epi Info 7® software with descriptive statistics to present absolute and relative values related to the description of the surgeries, the HP samples, and the values of the experiments. Measures of central tendency (mean) and standard deviation were used to assess the surgical time of the procedures included in this study.
This study did not involve research with human beings, thus dispensing with the need for an opinion from the Research Ethics Committee.
Results
Most neurosurgical procedures attributed to spinal arthrodesis (40%), with a duration time from 120 to 360 minutes, and a mean of 250.5 (±79.6). The main HPS analyzed were yasargil aspirators with dimensions of 19 cm/4 mm (40%) and 15 cm/5 mm (30%) located mainly in skull instrument cases (50%) (Table 1).
Table 1.
Description of Surgical Procedures and Samples of Health Products, 2020 (n = 10).
| Variables | n | % |
|---|---|---|
| Procedure | ||
| Spinal arthrodesis | 4 | 40 |
| Aneurysm clipping | 2 | 20 |
| Expansive lesions IC | 2 | 20 |
| Lumbar hernia | 1 | 10 |
| Craniectomy | 1 | 10 |
| Sample | ||
| Aspirators yasargil (19 cm/4 mm) | 4 | 40 |
| Aspirators yasargil (15 cm/5 mm) | 3 | 30 |
| Aspirators yasargil (18 cm/2 mm) | 2 | 20 |
| Aspirators yasargil (17 cm/5,5 mm) | 1 | 10 |
| Sample location | ||
| Skull instrument case | 5 | 50 |
| Lumbar spine box | 4 | 40 |
| External ventricular drain | 1 | 10 |
In assessing the quality of HPS cleaning, the protein test was positive for 4 (40%) samples collected after manual cleaning, and there was no detectable organic matter after automated cleaning. In the microbiological evaluation, 2 (20%) samples had the presence of E. coli, and 1 (10%) identified K. aerogenes. All positive results remained above 100,000 CFU/ml (Table 2).
Table 2.
Assessment of the quality of cleaning of health products included in the study, 2020 (n = 10).
| Sample | Protein test (stages) | Microbiological analysis (stages) | Identified microorganism (stages) | ||||||
|---|---|---|---|---|---|---|---|---|---|
| (a) | (b) | (c) | (a) | (b) | (c) | (a) | (b) | (c) | |
| 1. Aspirators yasargil (19 cm/4 mm) | Pos | Neg | Neg | NHC | NHC | NHC | Neg | Neg | Neg |
| 2. Aspirators yasargil (19 cm/4 mm) | Pos | Pos | Neg | NHC | >100.000* | >100.000* | Neg | E. coli | E. coli |
| 3. Aspirators yasargil (19 cm/4 mm) | Pos | Pos | Neg | NHC | 14.000* | >100.000* | Neg | E. coli | E. coli |
| 4. Aspirators yasargil (19 cm/4 mm) | Pos | Neg | Neg | NHC | NHC | NHC | Neg | Neg | Negativo |
| 5. Aspirators yasargil (15 cm/5 mm) | Pos | Neg | Neg | NHC | NHC | NHC | Neg | Neg | Neg |
| 6. Aspirators yasargil (15 cm/5 mm) | Pos | Pos | Neg | NHC | NHC | NHC | Neg | Neg | Neg |
| 7. Aspirators yasargil (15 cm/5 mm) | Pos | Neg | Neg | NHC | NHC | >100.000* | Neg | Neg | K. aerogenes |
| 8. Aspirators yasargil (18 cm/2 mm) | Pos | Neg | Neg | NHC | NHC | NHC | Neg | Neg | Neg |
| 9. Aspirators yasargil (18 cm/2 mm) | Pos | Pos | Neg | NHC | NHC | NHC | Neg | Neg | Neg |
| 10. Aspirators yasargil (17 cm/5,5 mm) | Pos | Neg | Neg | NHC | NHC | NHC | Neg | Neg | Neg |
Note: Pos. = Positive test; Neg. = Negative test; E. coli = Escherichia coli; K. aerogenes = Klebsiella aerogenes; NHC = No growth; * = CFU/ml; Stages: (a) in the operating room; (b) after manual cleaning; and (c) after automated cleaning.
Discussion
This study demonstrates that the protein test is an essential method for evaluating the quality of Health Products (HPs) cleaning. Despite being qualitative, the indicator’s results revealed shortcomings in the manual cleaning process. The data underscore that the combination of automated cleaning can aid in the complete removal of organic matter. Moreover, the detection of microorganisms above 100 CFU/ml, which are not typically associated with neurosurgical procedures, raises concerns about the possible presence of biofilms in the water pipes of the MSC.
Cleaning is undeniably crucial in reducing the microbial load of HPs. Researchers assert that sterilization can be jeopardized when dirt removal is inadequately performed in the initial stages (Yoo, 2018). If conducted superficially, it increases the rate of rework and the likelihood of adverse events (Washburn and Pietsch, 2018). Therefore, healthcare professionals should be trained to use magnifying glasses and chemical tests. This education should adhere to a structured Standard Operating Procedure (SOP) developed based on the best scientific evidence and parameters that ensure excellence in the process (Shreckengost et al., 2022). Task lighting is also necessary to enhance visibility during the cleaning process.
The protein test, which evaluates surfaces, is used in MSCs for the rapid detection of contaminant residues. International studies assessing protein load on surgical instruments found that cleaning processes were not standardized across the services evaluated, compromising the safety of the subsequent steps (Kothekar and Kulkarni, 2020; Yoo, 2018). Supporting this, the data from this study further confirm that 40% of the tested suction instrument samples were positive for protein, posing a significant risk for SSIs. As a result, a double-check procedure for cleaning all suction devices was implemented at the study site to ensure the cleaning stage’s quality (Washburn and Pietsch, 2018).
It is important to note that national legislation does not recommend a specific, rapidly readable chemical test to evaluate cleaning quality (Anvisa, 2012). Such standardization is vital for maintaining the quality of disinfection and/or sterilization of HPs (Alvim et al., 2019; Croke, 2020; Shreckengost et al., 2022). Internationally, the Association for the Advancement of Medical Instrumentation (AAMI) recommends standardizing a rapid methodology to assess the effectiveness of manual and/or automated cleaning (American National Standard, 2017).
The discovery of positive samples for E. coli and K. aerogenes in the study was unexpected, as there is no clear explanation in the literature about the relationship between these microorganisms and the contamination of suction instruments used in neurosurgical procedures. A plausible explanation may involve the formation of biofilm in the water pipes (Blaustein et al., 2016).
A significant finding of this study is the lack of periodic physicochemical and microbiological analysis of the water by the MSC. Coupled with the fact that the infrastructure is old—where the machinery was updated but the pipes that supply osmosis water to the ultrasonic washer have not been modified—these aspects could have directly impacted the microbiological results after manual and automated cleaning. Previous research has identified biofilm association caused by E. coli in irrigation systems providing drinking water (Almeida et al., 2019; Blaustein et al., 2016).
tBrazilian regulations currently do not mandate microbiological tests to evaluate the quality of cleaning nor define methodologies for laboratory evaluation (Anvisa, 2012). However, the data presented in this study underscore the need for the creation and enforcement of comprehensive regulations that establish a careful routine of instrumental evaluation. This is especially pertinent as the MSC frequently uses pressurized water guns and potable and/or purified water for cleaning and rinsing HPs, supplied by pipes at risk of biofilm formation (American National Standard, 2017; Rutala and Weber, 2019; Shreckengost et al., 2022).
A biofilm is defined as an extracellular polymeric matrix that protects microorganisms adhered to a surface in a coordinated manner (Kremer and McDonnell, 2020). Its formation is common in pipes and water reservoirs due to continuous water contact with surfaces (Blaustein et al., 2016; Kreme and McDonnell, 2020). The MSC is a sector with constant water handling in all stages of HP processing. However, the study site does not use reverse osmosis in this process, and the detection of uncommon microorganisms in the sector makes the research data concerning.
The researchers collected microbiological samples and conducted molecular biology analyses on the internal part of the washer-disinfector and the water pipes, confirming the presence of biofilms. The microorganisms identified were consistent with those detected in the confirmation samples collected.
This study represents a significant advancement for teaching and research, as it was unprecedented concerning the theme studied. It also facilitated the understanding of the importance of assessing the quality of cleaning of HPs, especially suction instruments. The results highlight the water pipes responsible for cleaning and rinsing HPs, which are often not prioritized in the MSC and may contribute to biofilm formation.
As a limitation, although the study included all neurosurgical procedures performed during the period, the sample size is considered limited. Nonetheless, it did not hinder achieving the study’s objective.
Conclusion
This study aimed to assess the cleaning quality of healthcare products used in neurosurgical procedures. It revealed notable deficiencies in the cleaning process of suction instruments, as evidenced by positive results from the protein test and the presence of bacteria with epidemiological significance in the collected lavage samples for microbiological analysis.
The identification of E. coli and K. aerogenes indicates compromised water quality, exceeding the recommended thresholds documented in the literature. These findings provide evidence of biofilm formation within the pipes of the research institution’s sterilization center, thereby disqualifying the water for potability purposes.
Footnotes
Author contributions: Conceptualization: ALSA. Methodology: ALSA. Validation: ALSA, AAV, EM, MAR, AMF, NBD, MOD, ALBG, LC, ESAV, AFLS, MC, and DA. Formal analysis: ALSA and EM. Investigation: ALSA, AAV, EM. Resources: ALSA, AAV, EM, MAR, AMF, NBD, MOD, ALBG, LC, ESAV, AFLS, MC, and DA. Data curation: ALSA. Writing—original draft preparation: ALSA, AAV, and EM. Writing-review and editing: ALSA, AAV, EM, MAR, AMF, NBD, MOD, ALBG, LC, ESAV, AFLS, MC, and DA. All authors have read and agreed to the published version of the manuscript.
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.
ORCID iDs
Andre Luiz Silva Alvim https://orcid.org/0000-0001-6119-6762
Adriely de Abreu Varoto https://orcid.org/0000-0002-4143-6834
Elaine Martins https://orcid.org/0000-0001-6302-9837
Adriano Menis Ferreira https://orcid.org/0000-0002-4054-768X
Alessandra Lyrio Barbosa Giroti https://orcid.org/0000-0002-8335-2030
Liliane Moretti Carneiro https://orcid.org/0000-0003-3195-8767
Emileide Dos Santos Almeida Vaz https://orcid.org/0000-0002-7489-1278
Alvaro Francisco Lopes de Sousa https://orcid.org/0000-0003-2710-2122
Data availability statement
The dataset analyzed during the current study is available from the corresponding author upon reasonable request.
References
- Almeida CJ, Romano UA, Damião FA, et al. (2019) Controle do biofilme em reservatórios e tubulações de água potável. Atas saúde ambiente 7: 1–15. [Google Scholar]
- Alvim AL, Ramos MS, Durão PMS. (2019) Monitoramento da limpeza de produtos para saúde com teste adenosina trifosfato. Revista SOBECC 24(2): 57–61. [Google Scholar]
- American National Standard (2017) Advancing safety in healthcare technology. Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities. Arlington: American National Standard/Advancing Safety in Healthcare Technology. [Google Scholar]
- Anvisa (2012) Resolução - RDC Nº 15/2012. Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Brasília: Diário Oficial da União. [Google Scholar]
- Blaustein RA, Shelton DR, Kessel JASV, et al. (2016) Irrigation waters and pipe-based biofilms as sources for antibiotic-resistant bactéria. Environmental Monitoring and Assessment 188(1): 56. [DOI] [PubMed] [Google Scholar]
- Croke L. (2020) Guideline for care and cleaning of surgical instruments. AORN Journal 112(3): 9–11. [DOI] [PubMed] [Google Scholar]
- Jiménez-Martínez E, Cuervo G, Hornero A, et al. (2019) Risk factors for surgical site infection after craniotomy: a prospective cohort study. Antimicrobial Resistance and Infection Control 8: 69. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kothekar AT, Kulkarni AP. (2020) Basic principles of disinfection and sterilization in intensive care and anesthesia and their applications during COVID-19 pandemic. Indian Journal of Critical Care Medicine 24(11): 1114–1124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kremer TA, McDonnell G. (2020) Sterilization central: water quality in medical device processing: the power of prevention. Biomedical Instrumentation & Technology 54(4): 304–309. [DOI] [PubMed] [Google Scholar]
- Ogrinc G, Davies L, Goodman D, et al. (2015) SQUIRE 2.0 (Standards for QUality Improvement Reporting Excellence): revised publication guidelines from a detailed consensus process. American Journal of Medical Quality 30(6): 543–549. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Patel S, Thompson D, Innocent S, et al. (2019) Risk factors for surgical site infections in neurosurgery. Annals of the Royal College of Surgeons of England 101(3): 220–225. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rutala WA, Weber DJ. (2019) Disinfection, sterilization, and antisepsis: an overview. American Journal of Infection Control 47S; A3–A9. [DOI] [PubMed] [Google Scholar]
- Shreckengost CSH, Starr N, Mammo TN, et al. (2022) Clean and confident: impact of sterile instrument processing workshops on knowledge and confidence in five low- and middle-income countries. Surgical Infections 3(2): 183–190. [DOI] [PubMed] [Google Scholar]
- Washburn RE, Pietsch JJ. (2018) Assessment of test methods for evaluating effectiveness of cleaning flexible endoscopes. American Journal of Infection Control 46(6): 685–688. [DOI] [PubMed] [Google Scholar]
- Yoo JH. (2018) Review of disinfection and sterilization – back to the basics. Infection Chemotherapy 50(2): 101–109. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The dataset analyzed during the current study is available from the corresponding author upon reasonable request.