Table 3. Overview of MRI data.
MRI sequences | Acquisition parameters | Rationale and outcome indices |
Magnetisation-prepared rapid gradient-echo (MP-RAGE) | 1 mm3 resolution, FOV: 256×256, TR=2300 ms, TE=2 ms, TI=857 ms, flip angle: 9; Duration~7 min | 3-dimensional (3D) T1-weighted anatomical image
|
Multi-shell High Angular Resolution Diffusion Imaging (msHARDI)33 | 2 mm3 resolution; FOV: 220×200; matrix size: 110×110 x 66; TE/TR=59/3000 ms; δ/Δ: 7/24 ms; b-values=0 (14 volumes), 500 (30 directions), 1200 (30 directions), 2400 (60 directions), 4000 (60 directions) and 6000 (60 directions) s/mm2; Duration ~20 min | Multishell diffusion weighted imaging data
|
Quantitative magnetisation transfer (qMT)55 56 | 1.72 mm3 resolution; FOV 220×220×179 mm; matrix size 128×128 x 104; turbo flash sequence, turbo factor 4, non-selective excitation MT pulse duration: 15.3 ms, 11 MT-weighted volumes and 1 vol without MT-weighting, 11 Frequency offsets (Hz) and 11 flip angles (degrees): 47 180 (628°); 56 360 (332°); 12060, (628°); 1000 (332°);1000 (333°); 2750 (628°); 2770 (628°); 2790 (628°); 2890 (628°); 1000 (628°); 1000 (628°); Duration ~25 min | qMT data to model the macromolecular proton fraction as an estimate of white matter myelin content36. |
Multicomponent-driven equilibrium single pulse observation of T1 and T2 (mcDESPOT)37 | 1.72 mm3 resolution; FOV 220×220×179 mm; matrix size 128×128×104;T1-weighted 3D spoiled gradient-recalled echo sequence (SPGR) with TR=4 ms; TE=1.9 ms; 8 flip angles (3°, 4°, 5°, 6°, 7°, 9°, 13° and 18°); Inversion recovery-prepped spoiled gradient-recalled echo sequence with TR=4 ms; TE=1.9 ms; flip angle 5°; steady-state free precession with TR=4.54 ms, TE=2.27 ms, 8 flip angles (10°, 13.33°, 16.67°, 20°, 23.33°, 30°, 43.33° and 60°), duration ~14 min | mcDespot data to model
|
FOV, field of view; mcDESPOT, Multicomponent Driven Equilibrium Single Pulse Observation of T1 and T2; MT, magnetisation transfer; TE, echo time; TI, inversion time; TR, repetition time