Fig. 5.
Isolevel distances, L, from the laser line. (b) Piecewise linear regression between the coagulation depth modelled with the wedge thickness, h, and static isolevels squared, L2. When estimating h, The 30% isolevel is used as long as ≥ 3 mm2, then 20% isolevel as long as ≥ 0.8 mm2, and finally the 15% when < 0.8 mm2. The curves end when the wedge is so thick that the static scattering estimate never becomes as low as the corresponding isolevel at any distance. (c) Least squares fit between h and the static scattering S at the center of the laser line using a logarithmic function .