Table 1. Optical properties from repeated measurements on human tissue at different wavelengths for SD and CWC methods. Values are reported as means, standard deviations (std), and coefficients of variation (CV).

  µa µs
Mean ± std (mm-1) CV (%) Mean ± std (mm-1) CV (%)
685 nm SD 0.0055 ± 0.0005 9.7 0.67 ± 0.02 2.3
CWC 0.0055 ± 0.0004 7.8 0.56 ± 0.01 2.4
852 nm SD 0.0064 ± 0.0003 4.3 0.53 ± 0.01 1.0
CWC 0.0067 ± 0.0001 1.1 0.433 ± 0.004 0.8
915 nm SD 0.0068 ± 0.0004 6.0 0.61 ± 0.01 1.3
CWC 0.0068 ± 0.0002 3.3 0.510 ± 0.004 0.8
980 nm SD 0.0137 ± 0.0003 2.4 0.52 ± 0.01 1.1
CWC 0.0147 ± 0.0002 1.7 0.423 ± 0.002 0.6