Towards a sensing model using a random laser combined with diffuse reflectance spectroscopy

Dongqin Ni 1,2,*, Florian Klämpfl 1,2, Michael Schmidt 1,2, Martin Hohmann 1,2
PMCID: PMC11427212  PMID: 39346981

Abstract

The previous research proves that the random laser emission reflects not only the scattering properties but also the absorption properties. The random laser is therefore considered a potential tool for optical properties sensing. Although the qualitative sensing using the random laser is extensively investigated, a quantitative measurement of optical properties is still rare. In this study, a generalized mathematical quantitative model using random laser combined with diffuse reflectance spectroscopy is proposed for optical sensing in turbid media. This model describes the gain effect of the active medium and the optical properties effect of the passive medium separately. Rhodamine 6G is used as the active medium. Intralipid and ink are employed to demonstrate the effect of the scattering and absorption, respectively. The peak wavelength shift of the random laser is proved to be an ideal sensing parameter for this sensing model. It is also revealed that the scaling parameters in the sensing model are interrelated and can be simplified to one. With this combined model, the direct sensing of optical properties in diverse turbid media is promising.


The understanding of light-matter interactions and the characterization of optical properties (OPs) are of paramount importance in various fields of research [1,2]. In biological and biomedical fields, a profound understanding of OPs is essential for the interpretation of diagnostic and therapeutic measurements, as well as for the development of medical technologies [2]. The characterization of OPs in turbid media, such as biological tissues, typically involves parameters such as absorption coefficient μa , reduced scattering coefficient μs , and anisotropy factor g [2]. While absorption and scattering strength are described by μa and μs , respectively, g denotes the scattering angle dependence [2]. The relation between the scattering parameters of μs and g is expressed by the equation:

μs=μs(1g), (1)

where μs represents the scattering coefficient.

The use of a random laser (RL) is emerging as a promising approach for the direct characterization of OPs in turbid media [3]. Within a RL system, the optical feedback for lasing stems from multiple scattering among random scatterers [4]. Consequently, the emission properties of RLs reflect the scattering properties of the random media, offering potential applicability in optical sensing in turbid media, particularly in biological tissues [4]. For example, RLs have demonstrated success in tissue differentiation [5], cancerous tissue screening [6,7], and the detection of optomechanical strain in tissues [8]. These achievements are noteworthy, although the aforementioned sensing applications are found solely on qualitative measurements. Some quantitative analysis of the RL emission are performed to correlate the strain [9,10], chemicals concentration [11,12] and pH value [13] in biological samples. However, a common sensing target such as the OPs of samples is still rare, thereby limiting their generalizability to diverse turbid systems. There is a need for a quantitative model using RL emission properties to assess the universal OPs in various turbid media.

Researchers have attempted to formulate a quantitative OPs sensing model based on RL emission [3,14]. The sensing parameters include the RL emission intensity, peak wavelength, linewidth, lasing modes and lasing threshold [4]. In particular, the work of Tommasi et al. [14] demonstrated the measurement of RL emission intensity to characterize the constant value of μs arising from microspheres with different diameters. They separated the active gain medium from the passive scattering medium by using an isolated transparent spherical cell. Both pump and emission light were guided through one fiber. In the work from Hohmann et al. [3], the active gain medium was mixed with the scattering medium. They reported the RL dependence on changes in μs , not only from the RL intensity, but also from the spectral peak wavelength and linewidth. The responses to changes in μs are consistent for all three parameters. Furthermore, it was empirically found that critical alterations in the RL emission variations are induced by μs rather than μs . Specifically, the RL emission exhibited maximum intensity when the laser cavity length was an integer multiple of the scattering mean free path ls (equivalent to 1/μs ), rather than the reduced scattering mean free path ls (equivalent to 1/μs ) [3]. Consequently, direct measurement employing RLs may provide a solution to the precise characterization of μs . The question remains in the optical characterization of μa . It was explored in a previous study [15], wherein an increase in μa resulted in a decrease in RL intensity, a broadening of the linewidth, a blueshift of the peak wavelength, and an increase in the lasing threshold [15]. It has been demonstrated that the RL emission exhibits an inverse behavior on μa in comparison to μs or μs [15]. The above three studies showed the feasibility of the RL emission to sense the OPs of μs , μs and μa . However, a mathematical representation of the sensing model which combines the three parameters is still lacking.

In fact, the aforementioned RL behavior is analogous to the well-known diffuse reflectance: diffuse reflectance intensity is directly proportional to μs and inversely proportional to μa [16]. The phenomenon that the RL responds similarly to the diffuse reflectance in the context of OPs is not fortuitous. Pioneering RL research has evidenced that RL generated from a turbid medium with scattering strength in the diffuse regime can be modeled as light diffusion with gain [17]. In this study, when observed in the backscattered direction, the RL can be hypothesized to be light diffuse reflectance with gain. Since the modeling of the diffuse reflectance spectroscopy (DRS) signal is well established in literature [16,18], the RL emission has high potential to be also modeled and applied for optical sensing.

In addition, the utilization of RL emission for sensing compared to the diffuse reflectance alone may facilitate deeper detection, since the typical sensing depth using diffuse reflectance is limited, for example, 0.5 - 1.9 mm in depth using a light source operating in the wavelength range of 350 - 1919 nm and applying a source-detector separation of 2500 µm [19]. This limitation will be overcome in RLs by cascaded gain amplification along the light path, leading to an extended active light path length in turbid media [4]. Furthermore, this effect may improve the sensitivity of RL-based OPs sensing due to the enhanced laser emission compared to the lamp typically used in the diffuse reflectance measurements. Therefore, the prospect of constructing a mathematical RL-based OPs sensing model, adapted from the diffuse reflectance model, with a higher sensitivity of a deeper sensing target, seems very promising. The empirical comparison between DRS and RL is summarized in Table 1. One can refer to the previous review paper for a more detailed discussion of the properties and applications of RLs [4].

Table 1. An empirical comparison between DRS and RL. The comparison is based on the nature of different techniques, but not fully experimentally evidenced.

DRS DRS-RL
Light source broadband lamps or LEDs narrowband laser which is generated along the light propagation inside turbid media
Detection instrumentation collection fiber collection optics with focusing lens
angle 0 degree to the incident light source no specific angle, but the angle alters RL intensity
separated from light source yes, separation affects the sensing depth NOT investigated yet in literature
Advantages simple instrumentation higher sensitivity and deeper sensing depth due to the cascaded light amplification when the light propagates inside the active medium
Disadvantages limited sensing depth fluorescent dye is required for lasing; limited wavelength range

The aim of this study is therefore to use the RL spectral properties to construct a mathematical DRS-RL model for quantitative OPs sensing in turbid media. In particular, for the first time, this RL model is able to separately describe the gain effect in addition to the OPs effect. This model can be adapted to various turbid media, but in this study it is validated only in the simplest turbid system consisting of Intralipid (IL) as scatterer and black ink as absorber.

1. Theoretical consideration

1.1. DRS model

The diffuse reflectance is indeed a function of μs , μa , phase function p(θ) (or g factor in the multiple scattering regime in this study) and geometry G [20]:

Rd=f(μs,μa,g,G). (2)

The phase function p(θ) describes the scattering angle dependence in a single scattering event, analogous to the function of the g factor in the multiple scattering condition. One study demonstrated the non-negligible influence of the phase function p(θ) on the diffuse reflectance in certain probe geometry configurations, e.g. when the light collection is close to the source [21]. The probe geometry G includes the effect of the source-detector separation and the effect of the numerical aperture of the fiber probe [22,23]. The symbol f() denotes a nonspecific function for the argument in bracket. The same function notation is applied in the following equations in this study.

Among various applied mathematical models [22], the Zonios’s model of diffuse reflectance in semi-infinite turbid media with fiber probes is the most practical one [22]. The diffuse reflectance Rd in the Zonios’s model reads:

Rd=μsk1+k2μa, (3)

where k1 and k2 are the scaling parameters depending on the optical probe geometry G [22]. The phase function was not formulated in this model.

Nevertheless, Eq. (2) and (3) indicate that a direct measurement of the OPs of μs , μs and μa is promising ( μs can be calculated with known μs and g according to the Eq. (1)). The difficulty lies in the measurement of g . Although g is a relevant parameter of OPs, a mathematical representation that separates its effect in Eq. (3) is still lacking in the DRS literature. A future study including the influence of the g factor into modeling is interesting, but it is not the scope of this study.

1.2. DRS-RL sensing principle: sensing parameter and sensing curve

The emission from the RL sample is either non-lasing or lasing emission. Only the RL signal in the lasing regime is utilized for sensing, because the RL emission in the lasing regime is more stable and stronger due to the dominant stimulated emission. To separate the non-lasing and lasing emissions, the peak wavelength of the RL is employed. As sketched in Fig. 1 (a), the peak wavelength λp always changes from blueshift to redshift at the lasing threshold when the pump energy E is increased. This trend of the peak wavelength shift above the lasing threshold is universal and independent of the OPs of the samples [15]. Moreover, this non-monotonic change of the peak wavelength makes it a better lasing indicator than the other RL parameters (e.g. FWHM and peak intensity) which only show monotonic changes at the lasing threshold [15]. Hence, the RL peak wavelength is chosen as the lasing threshold indicator, as well as the sensing parameter of the model.

Fig. 1.

Fig. 1.

Sensing principle: (a) raw data showing the response of the RL peak wavelength λp on the pump energy E . The lasing threshold is determined upon the peak wavelength changes from blueshift to redshift. Different colors represent the samples with different OPs of μs and μa . (b) pre-processed data to separate the non-lasing and lasing regime. λp and E are replaced by the relative peak wavelength Δλp and the relative pump energy ΔE , both taking the values at the lasing threshold as a reference. Only the data representing the lasing emission on the non-negative axes are further utilized for sensing. (c) converged sensing curve. The Δλp of all samples after the correction of the Rd effect, i.e. Δλp/Rd , responds identically to ΔE , due to the fact that the remaining gain effect is identical in all samples.

In practice, to separate the non-lasing and lasing regimes, the peak wavelength and pump energy are replaced by a relative peak wavelength Δλp=λλth and a relative pump energy ΔE=EEth , both of which take the values at the lasing threshold as a reference. In other words, both Δλp and ΔE are zero at the lasing threshold, as shown in Fig. 1 (b). Only the lasing regime where the Δλp and ΔE are non-negative is used for the further sensing application. Meanwhile, the nonlinearity of the lasing effect at the threshold is also excluded by using Δλp instead of λp as the sensing parameter, which makes the further formulation of the gain effect much easier.

The hypothesis that the RL emission is the light diffuse reflectance with gain is then mathematically expressed as:

Δλp=f(Rd,ΔE). (4)

Since the trend of the peak wavelength response to the pump energy is OPs or diffuse reflectance independent, the gain effect induced by ΔE , i.e. f(ΔE) can be formulated separately:

Δλp=Rdf(ΔE). (5)

When the effect of Rd is excluded from Δλp by division, it is assumed that the Δλp/Rd depends only on the ΔE , i.e. the gain effect. Mathematically, this relationship can be expressed as:

Δλp/Rd=f(ΔE) (6)

and by applying the Zonios’s model from Eq. (3):

Δλpk1+k2μaμs=f(ΔE). (7)

The right side of the Eq. (6) and (7) can be interpreted as the "OPs-independent gain effect" that can be measured when the sample has no scatterer and no absorber, but only a gain medium, i.e. a transparent active medium. In this study, the measurement of the "OPs-independent gain effect" was performed on the active medium of the Rhodamine 6G (R6G) water solution, as shown in Fig. 2.

Fig. 2.

Fig. 2.

Response of peak wavelength to the pump energy applied to a R6G gain medium.

Since the gain effect of the same gain medium is identical, the response of Δλp/Rd to ΔE is also identical for each sample. This identical response is sketched in Fig. 1 (c), where all RL peak wavelength after correcting for the OPs effect converge to an identical curve. This curve is referred to the "OPs-independent sensing curve" in this study.

Since the "OPs-independent gain effect" shows an exponential response in Fig. 2, the fit function of the "OPs-independent sensing curve" is predefined as an exponential function too:

Δλpk1init+k2initμaμs=aebΔE+c, (8)

where a , b , c are the fitting parameters, and k1init , k2init are the initial assumptions of the scaling parameters of k1 , k2 .

In a short summary, the RL emission detected from the backscattered direction is assumed to be the diffuse reflectance with gain. The RL peak wavelength shift after correcting for the diffuse reflectance effect is assumed to be identical due to the identical gain medium. Such a response can be expressed mathematically as Eq. (8) and illustrated graphically as the identical sensing curve in Fig. 1 (c). Given the sensing curve, the measured Δλp and ΔE , and the optimized scaling parameters k1 and k2 , the OPs of μs and μa can be derived from Eq. (8).

2. Experimental validation

2.1. RL samples and experimental setup

The RL sample prepared in this study is a water-based liquid medium, consisting of Rhodamine 6G (R6G; Sigma Aldrich, Germany) as the laser gain and Intralipid (IL; Fresenius Kabi, Germany) as the scatterer. Indian black ink (Royal Talens, The Netherland) was added as an external absorber. While the IL and ink concentrations were varied to change the scattering and absorption strength of the turbid media, the R6G concentration was kept at 2× 104 g/ml to achieve the optimal gain efficiency.

The method for calculating the OPs of scatters and absorbers was described in the previous study [15,24]. Two concentrations of IL scatterers of 5% and 9% v/v were selected, leading to μs of 63.78 and 106.13 cm1 , respectively. Since μs is wavelength dependent, the above values took the average of the individual values calculated at different peak wavelengths of the RL. The ink concentrations ranged from 0 to 0.09% v/v, corresponding to μa from 0 to 8.64 cm1 . The OPs values of μs and μa were selected according to the OPs of biological tissues [2]. Meanwhile, μs is at least 10 times larger than μa : the values satisfy the diffusion approximation where μs is much larger than μa .

The experimental setup is shown in Fig. 3 [15]. The liquid sample with a volume of 75 mL was filled into a beaker, resulting in a cylindrical shape with a diameter of 50 mm and a height of 38 mm. The pump light from a pulsed Nd:YAG laser (Q-smart 450, Quantel) was focused onto the sample, and the backscattered light was collected and directed to a spectrograph (Mechelle Me5000 Echelle, Andor). For each measurement, a laser pulse with a wavelength of 532 nm and a pulse duration of 5 ns was generated, and was focused on the sample surface with a spot diameter of 0.25 mm. The emission light from the same position of the excitation spot was detected with an angle of 45 . The detection angle was not specifically selected in this study, but only for the good placement of the collection optics. Despite the nonspecific selection, the detection angle of 45 still results in the collection of more than half of the maximum intensity regardless of distinct RL systems in literature [12,25,26]. The spectrograph has a spectral range from 200 to 975 nm with a spectral resolving power ( λ / Δλ ) of 6,000, i.e. a spectral resolution of 0.1 nm at the wavelength of 600 nm. For each sample, the laser pump energy E was varied from 1.82 mJ to 40.86 mJ in a trend of half-Gaussian distribution to probe both the non-lasing and lasing regime of the RL emission. Each measurement was repeated five times, and the averaged values were used.

Fig. 3.

Fig. 3.

RL experimental setup from our previous study [15]. The pump light from a pulsed Nd:YAG laser is focused onto the sample. The back-scattered emission light is collected from the excitation spot with an angle of 45 , and then guided by the optical fiber to a spectrograph. M: mirror; L: lens; NF: notch filter; NDF: neutral density filter.

2.2. Optimization of scaling parameters k1 and k2

The optimization of the scaling parameters k1 and k2 is estimated by the convergence of the data points to the identical sensing curve. To assess the convergence, the R2 value, which estimates the goodness of the fit, was applied as the optimization function. In practice, since the minimization algorithms are more commonly available than maximization algorithms, the minimum value of 1R2 was evaluated instead of the maximum value of R2 . Three scales were chosen as the initial assumption for k1 and k2 : 10, 1 and 0.1. These scales are comparable or smaller than the values of μs and μa in Eq. (8), so that the OPs rather than the scaling parameters of k1 and k2 dominate the diffuse reflectance. The tested values of k1 and k2 that lead to a minimum value of the objective function of 1R2 were returned as the optimal values.

Two different optimization methods: direct search algorithm Nelder-Mead and Markov chain Monte Carlo (MCMC) algorithm Metropolis-Hastings were applied to avoid the method-induced optimization bias. The former one was implemented by using the Python package of the "scipy.optimize.minimize" with "Nelder-Mead" as the optimizer, 1R2 as the objective function and k1init , k2init as the initial assumption. In the latter one, a Gaussian distribution near the initial assumption k was proposed ( k includes k1 and k2 ). A random point k from the proposed distribution was selected, and accepted if p(k)p(k)α , where p() denotes the objective function of 1R2 and α denotes a uniform random number between 0 and 1.

3. Results

3.1. Data preprocessing

The response of the RL peak wavelength λp on the pump energy E is shown in Fig. 4 (a). The right graph of Fig. 4 (a) shows a good agreement of peak intensity, FWHM and peak wavelength for lasing threshold determination. In the left graph of Fig. 4 (a): for samples with the same IL concentration, increasing ink concentration induces the higher lasing threshold Eth as well as the blueshift of the peak wavelength λp in the lasing regime. For samples with the same ink concentration, increasing the IL concentration leads to an inverse response: a lower lasing threshold and a redshift of the peak wavelength in the lasing regime.

Fig. 4.

Fig. 4.

Data preprocessing for the DRS-RL sensing. (a) response of the RL peak wavelength to the pump energy that is applied to the samples. The peak wavelength changes from blueshift to redshift at the lasing threshold determined by the peak intensity and FWHM, showing that the peak wavelength shift is a good indicator of the lasing threshold. (b) preprocessed data in the lasing regime. Different colors represent the samples with different OPs. (c) data cloud after the correction of the OPs effect. The data points are converged when dividing by Rd , the value of which is calculated taking the arbitrary numbers of 8 and 2 for k1 and k2 .

Figure 4 (b) shows the preprocessed data in the lasing regime. The peak wavelength of each RL sample with different OPs increases with increasing pump energy. The identical exponential response proves that the gain effect is identical and independent of the OPs. In Fig. 4 (c), after correcting for the OPs effect by dividing the diffuse reflectance Rd , the data points representing the peak wavelength shift are remarkably converged. It is noteworthy that the k1 and k2 in Rd take the arbitrary numbers of 8 and 2 to show the convergence. The convergence can be further improved by adjusting the scaling parameters of k1 and k2 .

3.2. Model parameter optimization: k1 and k2

Table 2 summarizes the optimization results of the two optimizers applying three different scales of initial assumptions of k1 and k2 . Although the returned optimal values of k1 and k2 are different, the values of the fitting goodness R2 are the same (the R2 values differ only in the 9th digit after the decimal). The bias caused by the selection of the optimizers or initial assumptions is thus eliminated. Moreover, the optimal convergence of the data points to the sensing curve is 94%, which is independent of the optimizer and initial assumption.

Table 2. Optimization of the scaling parameters of k1 and k2 .

k1init,k2init 10, 10 1, 1 0.1, 0.1
Optimal k1,k2R2 Optimal k1,k2R2 Optimal k1,k2R2
Nelder-Mead 4.22,1.520.94 0.43,0.150.94 0.04,0.020.94
Metropolis Hastings 8.17,2.940.94 0.54,0.190.94 0.14,0.050.94

Furthermore, the relationship between the optimal k1 and k2 derived from both optimizers is illustrated in Fig. 5 for a comparison. The results from both cases reveal the same linearity between k1 and k2 :

k2=0.36k1. (9)

Fig. 5.

Fig. 5.

Linearity of the k1 and k2 optimal values derived by the optimizer (a) Nelder-Mead and (b) Metropolis Hastings.

Although the revealed linearity helps to reduce the dimension of the scaling parameters, whether the constant slope of 0.36 has a physical meaning is unknown and requires further investigation. However, this is not the purpose of this study.

3.3. DRS-RL sensing curve

Figure 6 shows the DRS-RL sensing curve after the normalization. The sensing data points fit well to the exponential sensing curve with R20.94 . The outlines located outside of the 95% prediction interval may result from measurement errors.

Fig. 6.

Fig. 6.

The normalized DRS-RL sensing curve.

The sensing curve is formulated by applying the optimal values of k1 and k2 in Table 2 to the fitting function in Eq. (8). Although the optimal values of k1 and k2 vary widely using different optimizers and different initial assumptions, a unique fitting function is returned after normalizing Δλp/Rd to its maximum, as shown in Fig. 6. Taking into account the Eq. (9), the sensing curve is expressed as

Δλpk1(1+0.36μa)μs|norm=1.13(1e0.034ΔE). (10)

A general scaling factor k is further introduced to simplify the above equation:

kΔλp(1+0.36μa)μs=1e0.034ΔE, (11)

k=9.59 applying the measured ( Δλp , ΔE ), the known ( μs , μa ) values and the fitting parameters, and taking the average of the k values calculated for each data point. Therefore, the mathematical expression of the sensing curve is finally formulated as:

9.59Δλp(1+0.36μa)μs=1e0.034ΔE. (12)

From Eq. (12), the RL sensing parameter of Δλp is derived:

Δλp=19.59μs1+0.36μa(1e0.034ΔE). (13)

This equation indicates that the RL peak wavelength shift in the lasing regime can be represented by three combined components: (1) a general scaling factor, (2) the OPs and (3) the gain effect induced by the pump energy. Furthermore, the calculated values of Δλp from Eq. (13) are approximate to the original values measured experimentally, with an average R2=0.9323 as shown in Fig. 7.

Fig. 7.

Fig. 7.

The values of Δλp calculated from the sensing curve or from the Eq. (13) (represented by the curves), compared with the experimentally measured values of Δλp (represented by the dots).

The deviation is larger at pump energy well above the lasing threshold, especially for samples with lower values of μa/μs . The deviation might be reduced if the setup geometry is optimized. Another plausible explanation for the deviation is that some other nonlinear gain effect, such as gain saturation, happens at high pump energy, causing the data points to deviate from the exponential response of the sensing curve. Normally, the gain saturation happens when the pump energy is well above the lasing threshold. For samples with smaller μa/μs (i.e. less absorption loss and stronger optical feedback for lasing), the lasing threshold is lower. This means that for the same pump energy, the condition of being well above the lasing threshold is more easily reached in the samples with smaller μa/μs due to the lower lasing threshold, i.e. the gain saturation is more likely to occur and the data points are more likely to deviate. The proposed DRS-RL model is able to exclude the nonlinear gain effect at the lasing threshold, but not the nonlinear gain saturation well above the lasing threshold. A feasible approach is to exclude the data points at pump energies well above the lasing threshold (e.g. E>20Eth ) from the modeling. Despite this deviation, the sensing curve and equation are representative for the experimental measurements and can be applied for optical sensing.

3.4. Discussion and conclusion

Apart from the gain effect, RL in the backscattering direction responds similarly as the diffuse reflectance changes on optical properties. This study proposes a mathematical model of optical sensing that adapts the features of RLs to the common DRS model, the Zonios’s model. A linear relationship between the two scaling parameters in the Zonios’s model was revealed, and therefore the two scaling parameters can be simplified to one general scaling factor. In addition, it was shown that the pump energy induced gain effect in RL is universal and independent of the OPs of the samples investigated. The RL gain effect follows an exponential response of the transparent gain medium to the pump energy. For the first time, the OPs effect and the gain effect of the RL can be represented separately in a mathematical model for optical sensing.

The proposed optical sensing model was also experimentally validated. The reconstruction of the RL emission quantity using the sensing model can achieve an average R2 of 93.23%. However, the validation only included the turbid samples with two scattering coefficients. More samples with more scattering coefficients are needed to test the sensing model. Besides, the Zonios’s model was used because it is simplified and also representative of diffuse reflectance. A limitation is that the phase function p(Θ) or the g factor is not considered, although it has been shown that the p(Θ) or g factor has an influence on the diffuse reflectance, especially when the source and detector separation is very close. Whether these factors also affect the RL quantities is still questionable. Therefore, further investigations such as the influence of p(Θ) or g factor, the influence of the setup geometry (e.g. the detection angle, source-detection separation and pump spot size) and their interrelationship are needed to complete the proposed DRS-RL sensing model. Furthermore, the transformation of the proposed sensing model from the liquid sample to a solid tissue phantom, which is inharmonious, needs to be investigated for a more practical application.

Despite further investigation, the proposed DRS-RL sensing model combines the advantages of DRS for turbid medium sensing and lasing for deep medium sensing. The applications are not limited to the biomedical fields such as the quantification of cancerous degrees, but also extended to other fields such as the quantification of smoke particles in combustion filed, or the quantification of thermal degradation of coating in material filed.

Acknowledgment

This work was funded by the Deutsche Forschungsgemeinschaft (German Research Foundation–DFG 414732368). In addition, the authors gratefully acknowledge the funding of SAOT by the Bavarian State Ministry for Science and Art.

Funding

Erlangen Graduate School of Advanced Optical Technologies10.13039/501100007216; Deutsche Forschungsgemeinschaft10.13039/501100001659 (414732368).

Disclosures

The authors declare no conflicts of interest.

Data availability

The data used to support the findings of this study are available from the corresponding authors upon request.

References

  • 1.Flory F., “Optical properties of nanostructured materials: a review,” J. Nanophotonics 5(1), 052502 (2011). 10.1117/1.3609266 [DOI] [Google Scholar]
  • 2.Jacques S. L., “Optical properties of biological tissues: A review,” Phys. Med. Biol. 58(14), 5007–5008 (2013). 10.1088/0031-9155/58/14/5007 [DOI] [PubMed] [Google Scholar]
  • 3.Hohmann M., Späth M., Ni D., et al. , “Random laser as a potential tool for the determination of the scattering coefficient,” Biomed. Opt. Express 12(9), 5439 (2021). 10.1364/BOE.432052 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ni D., Späth M., Klämpfl F., et al. , “Properties and applications of random lasers as emerging light sources and optical sensors: a review,” IEEE Sens. J. 23(24), C1–29 (2023). 10.1109/JSEN.2023.3338435 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Hohmann M., Dörner D., Mehari F., et al. , “Investigation of random lasing as a feedback mechanism for tissue differentiation during laser surgery,” Biomed. Opt. Express 10(2), 807 (2019). 10.1364/BOE.10.000807 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Polson R. C., Vardeny Z. V., “Random lasing in human tissues,” Appl. Phys. Lett. 85(7), 1289–1291 (2004). 10.1063/1.1782259 [DOI] [Google Scholar]
  • 7.Mogharari N., Sajad B., “Random Laser Emission Spectra of the Normal and Cancerous Thyroid Tissues,” Iran J Sci Technol Trans Sci 43(4), 2055–2060 (2019). 10.1007/s40995-019-00691-8 [DOI] [Google Scholar]
  • 8.Song Q., Xu Z., Choi S. H., et al. , “Detection of nanoscale structural changes in bone using random lasers,” Biomed. Opt. Express 1(5), 1401 (2010). 10.1364/BOE.1.001401 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Briones-Herrera J. C., Cuando-Espitia N., Sánchez-Arévalo F. M., et al. , “Evaluation of mechanical behavior of soft tissue by means of random laser emission,” Rev. Sci. Instrum. 84(10), 104301 (2013). 10.1063/1.4823783 [DOI] [PubMed] [Google Scholar]
  • 10.Cuando-Espitia N., Sánchez-Arévalo F., Hernández-Cordero J., “Random laser imaging of bovine pericardium under the uniaxial tensile test,” Biomed. Opt. Express 9(8), 3523 (2018). 10.1364/BOE.9.003523 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Wan Ismail W. Z., Liu G., Zhang K., et al. , “Dopamine sensing and measurement using threshold and spectral measurements in random lasers,” Opt. Express 24(2), A85–A91 (2016). 10.1364/OE.24.000A85 [DOI] [PubMed] [Google Scholar]
  • 12.Shi X., Ge K., Tong J.-H., et al. , “Low-cost biosensors based on a plasmonic random laser on fiber facet,” Opt. Express 28(8), 12233 (2020). 10.1364/OE.392661 [DOI] [PubMed] [Google Scholar]
  • 13.Gaio M., Caixeiro S., Marelli B., et al. , “Gain-Based Mechanism for pH Sensing Based on Random Lasing,” Phys. Rev. Appl. 7(3), 034005 (2017). 10.1103/PhysRevApplied.7.034005 [DOI] [Google Scholar]
  • 14.Tommasi F., Ignesti E., Fini L., et al. , “Random laser based method for direct measurement of scattering properties,” Opt. Express 26(21), 27615–27627 (2018). 10.1364/OE.26.027615 [DOI] [PubMed] [Google Scholar]
  • 15.Ni D., Späth M., Klämpfl F., et al. , “Threshold behavior and tunability of a diffusive random laser,” Opt. Express 31(16), 25747 (2023). 10.1364/OE.491475 [DOI] [PubMed] [Google Scholar]
  • 16.Moy A. J., Tunnell J. W., Diffuse Reflectance Spectroscopy and Imaging (Elsevier Inc., 2016). [Google Scholar]
  • 17.Wiersma D. S., Lagendijk A., “Light diffusion with gain and random lasers,” Phys. Rev. E 54(4), 4256–4265 (1996). 10.1103/PhysRevE.54.4256 [DOI] [PubMed] [Google Scholar]
  • 18.Akter S., Hossain M. G., Nishidate I., et al. , “Medical applications of reflectance spectroscopy in the diffusive and sub-diffusive regimes,” J. Near Infrared Spectrosc. 26(6), 337–350 (2018). 10.1177/0967033518806637 [DOI] [Google Scholar]
  • 19.Nogueira M. S., Maryam S., Amissah M., et al. , “Evaluation of wavelength ranges and tissue depth probed by diffuse reflectance spectroscopy for colorectal cancer detection,” Sci. Rep. 11(1), 798 (2021). 10.1038/s41598-020-79517-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Calabro K. W., Bigio I. J., “Influence of the phase function in generalized diffuse reflectance models: review of current formalisms and novel observations,” J. Biomed. Opt. 19(7), 075005 (2014). 10.1117/1.JBO.19.7.075005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kanick S. C., Krishnaswamy V., Gamm U. A., et al. , “Scattering phase function spectrum makes reflectance spectrum measured from Intralipid phantoms and tissue sensitive to the device detection geometry,” Biomed. Opt. Express 3(5), 1086 (2012). 10.1364/BOE.3.001086 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Zonios G., Dimou A., “Modeling diffuse reflectance from semi-infinite turbid media: application to the study of skin optical properties,” Opt. Express 14(19), 8661 (2006). 10.1364/OE.14.008661 [DOI] [PubMed] [Google Scholar]
  • 23.Hennessy R., Goth W., Sharma M., et al. , “Effect of probe geometry and optical properties on the sampling depth for diffuse reflectance spectroscopy,” J. Biomed. Opt. 19(10), 107002 (2014). 10.1117/1.JBO.19.10.107002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Aernouts B., Van Beers R., Watté R., et al. , “Dependent scattering in Intralipid® phantoms in the 600-1850 nm range,” Opt. Express 22(5), 6086 (2014). 10.1364/OE.22.006086 [DOI] [PubMed] [Google Scholar]
  • 25.Tenopala-Carmona F., García-Segundo C., Cuando-Espitia N., et al. , “Angular distribution of random laser emission,” Opt. Lett. 39(3), 655 (2014). 10.1364/OL.39.000655 [DOI] [PubMed] [Google Scholar]
  • 26.Dey A., Pramanik A., Kumbhakar P., et al. , “Manoeuvring a natural scatterer system in random lasing action and a demonstration of speckle free imaging,” OSA Continuum 4(6), 1712 (2021). 10.1364/OSAC.422806 [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data used to support the findings of this study are available from the corresponding authors upon request.


Articles from Biomedical Optics Express are provided here courtesy of Optica Publishing Group