Table 1.
Influence of reducing washing agent volume on purification performance of SepPak C18 light (in-tandem alignment connected back-to-back).
Treatment (Wash & Rinse) |
Activity Retention (%LA) | Activity Retention (%LA) Post-Wash |
Recovery Efficiency (%RA) | RCP (%) |
---|---|---|---|---|
1 mL PBS (n = 9) | 99.7 ± 0.2 | 85.8 ± 6.9 | 60.7 ± 12.7 | 100 |
0.4 mL PBS (n = 6) | 99.8 ± 0.1 | 94.9 ± 6.2 | 83.5 ± 7.6 | 100 |
The SepPak C18 tandem setting was pre-conditioned with 10 mL of EtOH and 10 mL of H2O. Product activity was recovered from the cartridge via slow elution with 1 mL of 5% v/v EtOH (aqua) solution. Before purification, the RCP was determined via HPLC analysis of a crude reaction solution of ≥95%. %LA = percentage of the total loaded radioactivity from the starting radioactivity, with %LA measured (MBq) on the cartridges after the loading step pre- and post-wash. %RA = percentage of the total recovered radioactivity from the starting radioactivity with RA measured (MBq) after the activity elution (recovery) step.