Skip to main content
. 2024 Sep 28;14:22406. doi: 10.1038/s41598-024-72431-x

Table 4.

Results of delivered dose uniformity testing for determining Ipratropium and Fenoterol in Atrovent® comp HFA inhaler using the proposed spectrophotometric methods.

Atrovent® comp HFA meter dose no Label claim (%)
IPR FEN
ICS method IDW method IAM method ICS method IDW method IAM method
1 99.99 99.89 100.54 96.88 97.47 100.09
2 103.63 100.26 100.32 98.23 98.44 100.02
3 100.04 98.89 100.97 97.75 102.18 102.23
4 98.37 101.25 100.63 97.26 98.81 100.13
5 102.73 99.52 100.45 98.71 99.03 101.12
6 99.46 103.97 101.44 97.02 103.06 102.31
7 101.32 101.99 95.58 100.16 98.66 95.87
8 103.05 102.26 96.61 99.92 99.42 97.85
9 99.59 103.38 99.94 100.40 100.61 98.53
10 103.05 102.98 102.29 99.92 101.95 101.67
Mean 101.12 101.44 99.90 98.62 99.97 99.98
SD 1.78 1.66 2.0 1.31 1.78 1.95
RSD% 1.76 1.64 2.0 1.33 1.78 1.95
AV * 4.26 3.98 4.87 3.15 4.26 4.69

*Acceptance value = 2.4 × SD with maximum allowed level (L1) is 15.