Skip to main content
. 2024 Aug 7;36(5):275–281. doi: 10.5021/ad.23.103

Table 2. Test-retest reliability of the Corneometer and hydration probe.

Measurement location Device ICC 95% confidence interval
Ventral forearm Corneometer 0.971 0.963–0.978
Hydration probe 0.983 0.979–0.987
Dorsal forearm Corneometer 0.970 0.962–0.977
Hydration probe 0.984 0.980–0.988
Shin Corneometer 0.954 0.941–0.965
Hydration probe 0.980 0.974–0.984
Entire arm Corneometer 0.971 0.965–0.975
Hydration probe 0.984 0.981–0.986
Entire body Corneometer 0.966 0.961–0.971
Hydration probe 0.986 0.984–0.988

Entire arm means ventral and dorsal forearm, and entire body means all 3 sites measured.

ICC: intraclass correlation coefficient.