Table 5. Representative research results of skin hydration measurement devices.
Study | Measurement location | Device | Principle | Correlation coefficient |
---|---|---|---|---|
Fluhr et al. (1999)10 | Ventral forearm | Corneometer CM820 | Capacitance | CM820/CM825 SCC: 0.8901 |
Dorsal forearm | Corneometer CM825 | Capacitance | CM820/Skicon 200 SCC: 0.8996 | |
Dorsal hand | Skicon 200 | Conductance | CM820/Nova DPM SCC: 0.8165 | |
Front | Nova DPM | Impedance | CM820/DermaLab SCC: 0.8760 | |
Cheek | DermaLab | Conductance | CM825/Skicon 200 SCC: 0.8927 | |
Back (C4) | CM825/Nova DPM SCC: 0.7928 | |||
Leg | CM825/DermaLab SCC: 0.8964 | |||
Skicon 200/Nova DPM SCC: 0.8719 | ||||
Skicon 200/DermaLab SCC: 0.9358 | ||||
Nova DPM/DermaLab SCC: 0.8876 | ||||
Alanen et al. (2004)19 | Ventral forearm | Corneometer CM820 | Capacitance | CM820/SC-2 SCC: 0.75 (Ventral forearm) |
MoistureMeter SC-2 | Capacitance | |||
Clarys et al. (2012)8 | Sole of foot | Corneometer CM825 | Capacitance | CM825/Skicon 200 PCC: 0.97 |
Calf | Skicon 200 | Conductance | ||
Knee | ||||
Upper leg | ||||
Palm | ||||
Dorsal hand | ||||
Ventral forearm | ||||
Dorsal forearm | ||||
Upper arm | ||||
Forehead | ||||
Cheek | ||||
Abdomen | ||||
Hua et al. (2017)21 | Face | Corneometer CM825 | Capacitance | CM825/HP PCC: 0.808 |
Forearm | HP | Conductance | ||
Westermann et al. (2020)20 | Forearm | Corneometer CM825 | Capacitance | CM825/SkinUP PCC: 0.804 (Forearm) |
Forehead | SkinUp | Impedance | CM825/SkinUP PCC: 0.730 (Forehead) | |
Cheeks | CM825/SkinUP PCC: 0.696 (Cheeks) | |||
Choi et al. (2021)12 | Ventral forearm | HP | Conductance | HP/Biodisplay PCC: 0.601 (Ventral forearm) |
Biodisplay | Capacitance | |||
Present study | Ventral forearm | Corneometer CM820 | Capacitance | CM820/HP PCC: 0.72 (Ventral forearm) |
Dorsal forearm | HP | Conductance | CM820/HP PCC: 0.74 (Dorsal forearm) | |
Shin | CM820/HP PCC: 0.75 (Shin) | |||
CM820/HP PCC: 0.74 (Entire arm) | ||||
CM820/HP PCC: 0.75 (Entire body) |
SCC: Spearman’s correlation coefficient, HP: hydration probe, PCC: Pearson’s correlation coefficient.