Cemented total hip arthroplasty with 5 different articulations: pairwise comparison of differences in 1-to-10-year mean PE wear in mm with 95% confidence interval results from the linear mixed model

1-to-10-year mean PE wear, mm Difference
Reference Comparison b comparison–reference
Steel/CPE a 0.34 (0.12 to 0.56) CoCr/CPE 1.35 (1.16 to 1.55) 1.01 (0.71 to 1.31)
Steel/CPE a 0.34 (0.12 to 0.56) OxZr/CPE 1.68 (1.44 to 1.92) 1.34 (1.01 to 1.67)
Steel/CPE a 0.34 (0.12 to 0.56) CoCr/XLPE 0.08 (–0.11 to 0.26) –0.27 (–0.55 to 0.02)
Steel/CPE a 0.34 (0.12 to 0.56) OxZr/XLPE 0.06 (–0.14 to 0.26) –0.28 (–0.58 to 0.02)
CoCr/CPE b 1.35 (1.16 to 1.55) OxZr/CPE 1.68 (1.44 to 1.92) 0.33 (0.02 to 0.64)
CoCr/CPE b 1.35 (1.16 to 1.55) CoCr/XLPE 0.08 (–0.11 to 0.26) –1.28 (–1.54 to –1.01)
CoCr/CPE b 1.35 (1.16 to 1.55) OxZr/XLPE 0.06 (–0.14 to 0.26) –1.29 (–1.57 to –1.01)
OxZr/CPE b 1.68 (1.44 to 1.92) CoCr/XLPE 0.08 (–0.11 to 0.26) –1.60 (–1.90 to –1.30)
OxZr/CPE b 1.68 (1.44 to 1.92) OxZr/XLPE 0.06 (–0.14 to 0.26) –1.62 (–1.93 to –1.30)
CoCr/XLPE b 0.08 (–0.11 to 0.26) OxZr/XLPE 0.06 (–0.14 to 0.26) –0.01 (–0.29 to 0.26)