Standard liner jump height, liner stability ratio, and angle of coverage of different reverse total shoulder arthroplasty systems.

Reverse total shoulder arthroplasty system Cup/glenosphere size Jump height, mm Stability ratio, % Angle of coverage, °
Affinis Inverse (Ceramys) Inlay (Mathys) 36 7.5 139 109
39 8.6 148 112
42 9.9 160 116
Affinis Inverse (CoCrMo) Inlay (Mathys) 36 7.9 147 111
39 9.0 156 115
42 10.4 170 119
Comprehensive Shoulder Arthroplasty (Zimmer Biomet) 36 9.5 187 124
41 9.6 160 116
δ XTEND (DePuy Synthes) 38 8.3 147 112
42 9.4 151 113
Embrace Shoulder System - Enduro-S (CoCrMo) Inlay (Link) 39 8.9 155 114
42 8.8 141 109
Embrace Shoulder System UHMWPE Inlay (Link) 36 8.9 169 119
39 8.9 155 114
42 9.0 144 110
Equinoxe (Exactech) 38 8.2 144 110
42 8.4 134 107
46 8.7 126 103
Reverse Shoulder Prosthesis (DJO/Encore) 32 8.7 193 125
36 9.9 197 126
40 10.9 194 126
44 12.0 197 126
Reverse Shoulder System (Medacta) 32 6.4 134 107
36 7.1 132 106
39 7.8 134 106
42 8.4 133 106
SMR Reverse shoulder (LimaCorporate) 36 10.4 214 130
SMR Reverse shoulder - HP (LimaCorporate) 40 8.9 150 113
44 8.8 134 107
Trabecular Metal Reverse (Zimmer Biomet) 36 8.1 153 114
40 7.7 128 104
Tornier Perform Reversed (Stryker) 33 7.3 150 113
36 8.0 149 112
39 8.5 146 111
42 9.9 160 116
Unic (Evolutis) 33 7.9 164 117
37 8.4 154 114
Univers Reverse Humeral Prosthesis (Arthrex) 36 9.7 193 125
39 10.6 194 125
42 11.4 194 125
Verso shoulder system (Innovative Design Orthopaedics) 36 8.0 149 112
41 10.4 175 121