Table III.

Constrained liner jump height, liner stability ratio, and angle of coverage of different reverse total shoulder arthroplasty systems.

Reverse total shoulder arthroplasty system Cup/glenosphere size Jump height, mm Stability ratio, % Angle of coverage, °
Comprehensive Shoulder Arthroplasty (Zimmer Biomet) 36 12.6 315 145
41 11.7 209 129
δ XTEND (DePuy Synthes) 38 12.0 251 137
42 13.2 248 136
Equinoxe (Exactech) 38 11.6 235 134
42 12.2 217 131
46 12.8 201 127
Reverse Shoulder Prosthesis (DJO/Encore) 32 9.9 244 135
36 11.1 243 135
40 12.4 245 136
44 13.7 244 135
SMR Reverse shoulder - HP (LimaCorporate) 40 9.8 169 119
44 9.9 151 113
Trabecular Metal Reverse (Zimmer Biomet) 36 9.7 193 125
40 9.1 153 114
Tornier Perform Reversed (Stryker) 33 9.6 216 130
36 9.8 194 126
39 10.5 191 125
42 11.0 184 123
Unic (Evolutis) 37 11.9 264 139
Univers Reverse Humeral Prosthesis (Arthrex) 36 12.3 300 143
39 13.0 281 141
42 13.9 277 140
Verso shoulder system (Innovative Design Orthopaedics) 36 13.9 426 154
41 16.3 479 156