Table IV.
Stability ratio for standard liners across their own different sizes.
Arthroplasty system | Trend |
---|---|
Stable stability ratio (< 5%) | |
δ XTEND (DePuy Synthes) | ↔ |
Reverse Shoulder Prosthesis (DJO/Encore) | ↔ |
Reverse Shoulder System (Medacta) | ↔ |
Univers Reverse Humeral Prosthesis (Arthrex) | ↔ |
Slight inconsistencies (5% to 10%) | |
Unic (Evolutis) | ↘ |
Notable inconsistencies (> 10%) | |
Affinis Inverse Ceramys Inlay (Mathys) | ↗ |
Affinis Inverse CoCrMo Inlay (Mathys) | ↗ |
Comprehensive Shoulder (Zimmer Biomet) | ↘ |
Embrace Shoulder System; Enduro-S Inlay (Link) | ↘ |
Embrace Shoulder System; UHMWPE Inlay (Link) | ↘ |
Equinoxe (Exactech) | ↘ |
SMR Reverse shoulder HP liners (LimaCorporate) | ↘ |
Trabecular Metal Reverse Arthroplasty (Zimmer Biomet) | ↘ |
Tornier Perform Reversed (Stryker) | ↘↗ |
Verso shoulder system (IDO) | ↗ |