Table 2. Intraday and Interday Accuracy and Precision of LLOQ and QC Sample of HHC in Mice Plasma (Mean ± Standard Deviation (SD)).

item nominal conc. (ng/mL) observed conc. (ng/mL) accuracy (%) precision (CV%)
accuracy and precision of LLOQ sample of HHC
intra- and interday
calibration curve: y = 1.5341 × 10–3x – 1.8465 × 10–4; R2 = 0.9959
intraday (n = 3) 5.00 5.30 ± 0.30 106.04 5.66
interday (n = 9) 5.00 4.97 ± 0.66 99.47 13.19
accuracy and precision of QC sample of HHC
intraday
calibration curve: y = 1.5341 × 10–3x – 1.8465 × 10–4; R2 = 0.9959
interday
calibration curve day1: y = 1.5341 × 10–3x – 1.8465 × 10–4; R2 = 0.9959
calibration curve day2: y = 1.6891 × 10–3x – 2.2302 × 10–4; R2 = 0.9932
calibration curve day3: y = 1.7797 × 10–3x – 1.5662 × 10–3; R2 = 0.9979
intraday (n = 3) 15.00 14.80 ± 0.49 98.68 3.28
150.00 157.60 ± 10.75 105.07 6.82
375.00 382.78 ± 33.28 102.08 8.69
interday (n = 9) 15.00 14.39 ± 1.11 95.97 7.73
150.00 143.38 ± 8.50 95.59 8.89
375.00 356.74 ± 0.30 95.13 9.26