Table 2.
Receiver operating characteristic analysis for each parameter in differentiating NL from KC.
| AUROC | 95% CI | Cutoff | Sensitivity (%) | Specificity (%) | |
|---|---|---|---|---|---|
| Scansys | |||||
| KCP | 1.000 | 0.976–1.000 | > 16.4 | 100 | 98.11 |
| Pentacam | |||||
| BAD-D | 1.000 | 0.977–1.000 | > 3.12 | 100 | 100 |
| IS-value | 0.986 | 0.962–1.000 | > 1.455 | 96.08 | 100 |
| KISA | 0.992 | 0.980–1.000 | > 44.11 | 94.12 | 99.06 |
| ISV | 1.000 | 0.977–1.000 | > 35.5 | 100 | 100 |
| IVA | 1.000 | 0.977–1.000 | > 0.255 | 100 | 100 |
| KI | 1.000 | 0.977–1.000 | > 1.075 | 100 | 99.06 |
| CKI | 0.974 | 0.947–1.000 | > 1.025 | 90.2 | 100 |
| IHA | 0.912 | 0.855–0.969 | > 11.5 | 82.35 | 90.57 |
| IHD | 1.000 | 0.977–1.000 | > 0.027 | 100 | 100 |
| Corvis ST | |||||
| CBI | 0.998 | 0.971–1.000 | > 0.566 | 98.04 | 99.06 |
| TBI | 1.000 | 0.977–1.000 | > 0.938 | 100 | 100 |
| SPA1 | 0.945 | 0.907–0.983 | ≤ 86.953 | 88.24 | 94.34 |
| DA ratio (2 mm) | 0.926 | 0.873–0.979 | > 4.755 | 86.27 | 90.57 |
| Integrated radius | 0.964 | 0.934–0.994 | > 9.546 | 84.31 | 96.23 |
NL normal controls, KC keratoconus, AUROC area under the receiver operating characteristic curve, CI confidence interval.