Table 2.

Receiver operating characteristic analysis for each parameter in differentiating NL from KC.

AUROC 95% CI Cutoff Sensitivity (%) Specificity (%)
Scansys
 KCP 1.000 0.976–1.000 > 16.4 100 98.11
Pentacam
 BAD-D 1.000 0.977–1.000 > 3.12 100 100
 IS-value 0.986 0.962–1.000 > 1.455 96.08 100
 KISA 0.992 0.980–1.000 > 44.11 94.12 99.06
 ISV 1.000 0.977–1.000 > 35.5 100 100
 IVA 1.000 0.977–1.000 > 0.255 100 100
 KI 1.000 0.977–1.000 > 1.075 100 99.06
 CKI 0.974 0.947–1.000 > 1.025 90.2 100
 IHA 0.912 0.855–0.969 > 11.5 82.35 90.57
 IHD 1.000 0.977–1.000 > 0.027 100 100
Corvis ST
 CBI 0.998 0.971–1.000 > 0.566 98.04 99.06
 TBI 1.000 0.977–1.000 > 0.938 100 100
 SPA1 0.945 0.907–0.983 ≤ 86.953 88.24 94.34
 DA ratio (2 mm) 0.926 0.873–0.979 > 4.755 86.27 90.57
 Integrated radius 0.964 0.934–0.994 > 9.546 84.31 96.23

NL normal controls, KC keratoconus, AUROC area under the receiver operating characteristic curve, CI confidence interval.