Table 6.

Agreement of parameters measured using the scansys and pentacam.

Group Parameters Mean ± SD P-valuea r 95% LoA
Scansys   Pentacam
NL CCT 547.73 ± 29.94 552.20 ± 29.19 < 0.001a 0.915b − 19.43 to 28.36
TCT 542.39 ± 29.01 549.29 ± 29.17 < 0.001a 0.920b − 15.91 to 29.70
Kmax 44.29 ± 1.56 44.19 ± 1.55 0.125 0.915b − 1.35 to 1.16
PTE 4.85 ± 4.34 4.67 ± 2.49 0.595 0.615b − 6.89 to 6.53
KCS CCT 527.38 ± 30.41 533.44 ± 31.95 < 0.001a 0.971b − 8.99 to 21.10
TCT 520.31 ± 30.92 527.82 ± 32.09 < 0.001a 0.972b − 7.28 to 22.29
Kmax 45.82 ± 2.08 45.61 ± 2.16 0.007a 0.958b − 1.413 to 1.01
PTE 12.155 ± 5.98 12.37 ± 4.24 0.658 0.742b − 7.65 to 8.07
KC CCT 464.61 ± 47.86 462.55 ± 44.09 0.335 0.949b − 31.67 to 27.56
TCT 450.94 ± 47.178 455.04 ± 42.59 0.042a 0.956b − 23.42 to 31.62
Kmax 58.81 ± 11.75 58.86 ± 10.32 0.908 0.983b − 4.83 to 4.91
PTE 62.30 ± 36.98 51.63 ± 24.56 < 0.001a 0.874b − 48.99 to 27.65

Data are shown as mean ± SD.

NL normal controls, KCS keratoconus suspects, KC keratoconus, CCT central corneal thickness, TCT thinnest corneal thickness, Kmax maximum keratometry, PTE posterior elevation at the thinnest point, LoA limits of agreement, SD standard deviation.

The aP-values are for the comparison of the Scansys and Pentacam parameters (paired t-test, aP < 0.05 indicates a statistically significant difference); r, Pearson’s correlation coefficient; bP < 0.05 indicates a statistically significant correlation.