Figure 3.

Figure 3.

Transverse magnetization response as a function of the RF offset to various water suppression schemes used in 2D [1H,1H] NOESY experiments. For simplicity, the offset is reported in ppm. (A) 3–9-19 sequence with an interpulse delay of 200 μs. (B) 3–9-19 sequence with an interpulse delay of 75 μs. (C) WATERGATE sequence. (D) modified WATERGATE with water flip-back pulses. (E). Excitation sculpting sequence. (F) WADE water suppression sequence. The green lines show the Mz magnetization that on-resonance reaches a maximum for the WATERGATE with water flip-back (D) and WADE (F) sequences.