Table 2.

General recommendations on RMN-guided catheter ablation.

Recommendation Level of consensus Symbol References
Staff
It is recommended to work with a dedicated clinical team to operate the RMN system and care for the patient 24/27 (89%) graphic file with name fcvm-11-1431396-i006.jpg 11
It is recommended that the clinical team is specifically trained to operate the RMN system 26/27 (96%) graphic file with name fcvm-11-1431396-i007.jpg 11
The clinical team may consist both off registered nurses, as well as registered technicians 16/27 (59%) graphic file with name fcvm-11-1431396-i057.jpg EO
Manufacturer's support may be considered to assist the staff to operate the RMN system. 18/27 (67%) graphic file with name fcvm-11-1431396-i058.jpg EO
Preparation and positioning of the patient
It is recommended that the patient is set up on the table by either dedicated nurses and/or technicians 27/27 (100%) graphic file with name fcvm-11-1431396-i008.jpg EO
Arm boards are recommended to support of the arms of the patient during the procedure 22/27 (81%) graphic file with name fcvm-11-1431396-i009.jpg EO
Other types of arm support may be considered (e.g., soft Velcro wraps, fixation with blanket, angled arm support and gel pads) 14/27 (52%) graphic file with name fcvm-11-1431396-i059.jpg EO
Positioning of patients for left-sided ventricular targets is not more challenging than for other targets. 14/17 (82%) graphic file with name fcvm-11-1431396-i010.jpg EO
It may be considered to iso-center the chamber of interest before catheters are inserted 17/27 (63%) graphic file with name fcvm-11-1431396-i060.jpg EO
It may be considered to perform pre-procedural imaging (either with CT or MRI) for anatomical and/or substrate representation, and image integration 21/27 (78%) graphic file with name fcvm-11-1431396-i061.jpg 9, 10, 13, 14
It is recommended to use a mechanical ventilator that is MRI compatible, or otherwise place it outside the 5 Gauss zone 22/27 (81%), 5/5 (100%), respectively graphic file with name fcvm-11-1431396-i011.jpg EO
Fluoroscopy
The Siemens and Philips fluoroscopy systems are compatible with RMN 27/27 (100%) graphic file with name fcvm-11-1431396-i012.jpg EO
It may be considered to take and store fluoroscopy images for use in Navigant software of the RMN system 17/27 (63%) graphic file with name fcvm-11-1431396-i062.jpg EO
It may be considered to use catheter overlay fluoroscopy images to reduce radiofrequency exposure 16/27 (59%) graphic file with name fcvm-11-1431396-i063.jpg EO
Point annotation on fluoroscopy is used by a minority to visualize structures or equipment 13/27 (48%) graphic file with name fcvm-11-1431396-i081.jpg EO

EO, expert opinion; MRI, magnetic resonance imaging; RMN, robotic magnetic navigation.