Failure reasons of retrieved plates from revision surgery.

Failure information Magnitude
Healing status Delayed union 7
Non-union 37
Others 5
Materials Pure Ti 38
Ti6Al4V 6
Stainless steel 5
Fracture mode Fatigue fracture 46
Quasi-cleavage fracture 3