Regression parameters for the determination of NIC and CAF using the proposed green chromatographic method

Parameter NIC CAF
Linearity range (μg/mL) 0.1–20.0 0.2–40.0
Correlation coefficient (r) 0.9999 0.9999
Slope ± SDa 317.13 ± 3.17 24.65 ± 0.49
Intercept ± SDa 31.58 ± 0.33 18.82 ± 0.03
LODb (μg/mL) 0.03 0.07
SDa of residuals 6.61 1.02
% RSDc 1.24 1.25
% Error 0.44 0.44

a Standard deviation

b Limit of detection

c Relative standard deviation