Table 1.
Regression parameters for the determination of NIC and CAF using the proposed green chromatographic method
Parameter | NIC | CAF |
---|---|---|
Linearity range (μg/mL) | 0.1–20.0 | 0.2–40.0 |
Correlation coefficient (r) | 0.9999 | 0.9999 |
Slope ± SDa | 317.13 ± 3.17 | 24.65 ± 0.49 |
Intercept ± SDa | 31.58 ± 0.33 | 18.82 ± 0.03 |
LODb (μg/mL) | 0.03 | 0.07 |
SDa of residuals | 6.61 | 1.02 |
% RSDc | 1.24 | 1.25 |
% Error | 0.44 | 0.44 |
a Standard deviation
b Limit of detection