Table 2.

Sensitivity, specificity, positive and negative predictive value as well as accuracy of nonimaging parameters to predict trapped air, with the trapped air % from spirometry-controlled computed tomography used as the gold standard and comparitor for this analysis

Parameter Sensitivity, % Specificity, % Positive Predictive Value, % Negative Predictive Value, % Accuracy, %
ppFEV1 28 83 84 27 42
ppFVC 11 83 68 23 29
ppFEF25–75% 63 75 90 37 67
ppFEF75% 61 50 81 26 58
RV/TLC% 11 100 100 26 32
LCI2.5 47 100 100 38 60
LCIShX 63 100 100 46 72
UVLU 79 100 100 60 80
5IC VTG/VC 47 100 100 25 55
5VC VTG/VC 50 100 100 37 61

5IC, 5 inspiratory capacity breaths; 5VC, 5 vital capacity; LCI, lung clearance index; ppFEF25–75%, percent predicted forced expiratory flow at 25–75% of the forced vital capacity (FVC); FEF75%, percent predicted forced expiratory flow when 75% of FVC has been exhaled; ppFEV1, percent predicted forced expiratory volume in 1 second; ppFVC, percent predicted forced vital capacity; RV, residual volume; TLC, total lung capacity; UVLU, under-ventilated lung units; VTG, volume of trapped gas.