Skip to main content
. 2024 Oct 18;14(45):32876–32882. doi: 10.1039/d4ra05715e

Accuracy (N = 3) and precision results – repeatability (N = 6 samples per concentration level in one day) and intermediate precision (N = 9 samples per concentration level in 3 days).

ALP concentration (μg mL−1) Accuracy Repeatability Intermediate precision
Bias (%) Migration time RSD (%) IS-corrected peak area RSD (%) Migration time RSD (%) IS-corrected peak area RSD (%)
40 3.0 0.08 1.25 1.73 4.96
50 0.9 0.17 0.67 1.00 3.93
60 2.3 0.29 1.10 0.31 5.87