Skip to main content
Europace logoLink to Europace
. 2024 Oct 1;26(10):euae251. doi: 10.1093/europace/euae251

Intravascular haemolysis and acute kidney injury following atrial fibrillation ablation: a report using two different systems for pulsed field ablation

Maarten A J De Smet 1,, Clara François 2, Benjamin De Becker 3, Rene Tavernier 4, Jean-Benoît le Polain de Waroux 5, Sébastien Knecht 6, Mattias Duytschaever 7,b
PMCID: PMC11493093  PMID: 39351957

Abstract

Keywords: Atrial fibrillation, Ablation, Pulsed field, Radiofrequency, Acute kidney injury, Haemolysis


Pulmonary vein isolation (PVI) is an established cornerstone of rhythm management in patients with atrial fibrillation (AF).1,2 Before pulsed field ablation (PFA), PVI lesions were solely created by thermal radiofrequency (RF), cryo, or laser energy. Pulsed field ablation, resulting in irreversible electroporation of cell membranes to which cardiac muscle is especially sensible, is the first non-thermal energy that provides effective PVI with good durability, including non-inferiority when compared with RF.2,3

Initial reports showed an excellent clinical safety profile for PFA.3,4 More recently, two cases of severe acute kidney injury (AKI) secondary to intravascular haemolysis (IH) have been described following AF ablation with 174 and 126 PF applications.5 Also, another study reported increased serum creatinine and haemoglobinuria following a higher number of PF applications without hydration.6 In MANIFEST-17k, the largest worldwide PFA registry, 5 out of more than 17 000 patients (0.03%) were reported to have transient haemolysis-related kidney failure requiring temporary dialysis (143 ± 27 PF applications).7 All these reports only studied the pentaspline PFA catheter since its clinical approval in 2021 and agreed on the fact that some degree of haemolysis occurs in most patients and is proportionate to the number of PF applications.5,7,8 So far, no study compared novel PFA technologies on the market and the question remains whether this effect also occurs in these technologies.

Here, we assessed peri-procedural IH and kidney function following PVI-only or PVI with additional lesion sets (PVI-plus group; roof and posterior mitral isthmus lines, vein of Marshall ethanolization, and/or posterior wall isolation) using two commercially available PFA systems (Varipulse variable loop PFA catheter, Biosense Webster; Farawave pentaspline PFA catheter, Boston Scientific) or RF energy (ThermoCool SmartTouch, Biosense Webster). Variable loop PFA procedures used the CARTO 3 system (Biosense Webster Inc., Diamond Bar, CA, USA) for mapping. For the pentaspline PFA catheter, PV angiograms were performed unless contra-indicated [contrast allergy, risk of thyrotoxicosis, advanced chronic kidney disease (CKD)]. Patients provided a blood sample before ablation and a blood and urine sample 24 h post-ablation. Acute kidney injury was classified according to the Kidney Disease Improving Global Outcomes classification: serum creatinine 1.5–1.9 times baseline or ≥0.3 mg/dL increase (Stage 1), 2–2.9 times baseline (Stage 2), and 3 times baseline or increase of serum creatinine ≥ 4 mg/dL or initiation of renal replacement therapy (RRT) (Stage 3). The study was carried out according to the principles of the Declaration of Helsinki and was approved by the local ethical committee. Data are reported as mean ± standard deviation or median (IQR). Comparative statistics were performed using one-way ANOVA or Student’s t-test for normally distributed continuous variables and χ2 test for percentages or categorical data. The Mann–Whitney test was used for non-uniformly distributed data.

One hundred ninety-eight consecutive patients underwent a first ablation procedure using the variable loop PFA catheter in 54 out of 198 patients (27%), the pentaspline PFA catheter in 98 patients (49%), or RF in 46 patients (24%). Pulmonary vein isolation-only ablation was performed in 129 out of 198 patients (65%). In 69 patients (35%), PVI-plus was performed. Demographic and clinical characteristics were not significantly different between ablation modalities (Table 1). However, the PVI-plus group showed significantly more CKD, structural heart disease (SHD) and heart failure, lower LVEF, and larger left atria as compared with PVI-only. Mean RF applications were 68 ± 15 (range 38–106) for PVI-only and 88 ± 15 (range 57–118) for PVI-plus with a mean total RF ablation (RFA) time of 1336 ± 445 s and 1866 ± 565 s, respectively. Using the variable loop PFA catheter, mean applications were 46 ± 6 (range 36–64) for PVI-only and 62 ± 11 (range 50–83) for PVI-plus, whereas these were 33 ± 2 (range 32–44) and 46 ± 6 (range 32–66) applications, respectively, for the pentaspline catheter.

Table 1.

Summary table showing demographics, clinical and procedural characteristics, and haemolysis parameters following radiofrequency ablation, variable loop PFA, and pentaspline PFA

Baseline characteristics Radiofrequency Variable loop PFA Pentaspline PFA
PVI-only (n = 32) PVI-plus (n = 14) PVI-only (n = 40) PVI-plus (n = 14) PVI-only (n = 57) PVI-plus (n = 41)
Age (years) 68 ± 10 71 ± 8 66 ± 11 67 ± 9 65 ± 12 69 ± 8
Female, n (%) 14 (44) 4 (29) 12 (30) 5 (36) 24 (42) 18 (44)
SHD, n (%) 12 (38) 8 (57)* 10 (25) 6 (43)* 17 (30) 14 (34)
 CAD, n 6 4 5 1 8 8
 DCM, n 1 0 0 1 0 0
 HCM, n 1 1 1 0 2 0
 VHD, n 2 0 1 1 1 0
 Congenital, n 1 0 0 0 2 0
 TCMP, n 1 3 3 0 4 6
Heart failure, n (%) 6 (19) 9 (64)* 10 (25) 6 (43)* 12 (21) 12 (29)
 HFpEF, n 2 3 4 2 2 3
 HFmrEF, n 3 5 4 2 7 6
 HFrEF, n 1 1 2 2 3 3
LVEF (%) 61 ± 7 55 ± 11* 62 ± 5 55 ± 17* 63 ± 8 58 ± 11**
LAD (mm) 41 ± 5 45 ± 4** 39 ± 5 47 ± 8*** 39 ± 6 43 ± 6***
LAVI (mL/m²) 27 ± 10 37 ± 14* 29 ± 8 37 ± 11* 29 ± 9 45 ± 10***
CKD, n (%) 3 (9) 5 (36)** 1 (3) 5 (36)** 8 (14) 11 (27)*
Serum creatinine (mg/dL) 1.01 ± 0.34 1.03 ± 0.28 0.91 ± 0.16 0.99 ± 0.2 0.94 ± 0.4 0.99 ± 0.3
eGFR (CKD-EPI, mL/min) 73 ± 20 71 ± 18 83 ± 14 75 ± 17 80 ± 18 70 ± 18**
DM, n (%) 4 (13) 6 (43)* 3 (8) 2 (14) 7 (12) 7 (17)
AHT, n (%) 15 (47) 8 (57) 10 (25) 8 (57) 21 (37) 22 (54)
Previous stroke, n (%) 1 (3) 1 (7) 3 (8) 0 (0) 8 (14) 5 (12)
PAD, n (%) 0 (0) 0 (0) 1 (3) 0 (0) 1 (2) 3 (7)
CHA2DS2-Vasc score 2 (1–4) 3 (2–5) 2 (0–3) 3 (1–4) 2 (1–3) 3 (2–5)*
Outcomes Radiofrequency Variable loop PFA Pentaspline PFA
No AKI (n = 46) No AKI (n = 50) Stage 1 (n = 4) No AKI (n = 94) Stage 1 (n = 3) Stage 3 (n = 1)
Age (years) 69 ± 9 63 ± 13 69 ± 6 68 ± 10 72 ± 7 78
Female, n (%) 21 (46) 16 (32) 1 (25) 42 (45) 0 (0) 0 (0)
PVI-plus, n (%) 14 (30) 12 (24) 2 (50) 38 (40) 3 (100) 0 (0)
CKD, n (%) 17 (37) 6 (12) 3 (75) 27 (29) 2 (75) 1 (100)
DM, n (%) 10 (22) 3 (6) 2 (50) 13 (14) 1 (33) 1 (100)
AHT, n (%) 23 (50) 16 (32) 3 (75) 40 (43) 2 (75) 1 (100)
SHD, n (%) 20 (43) 6 (12) 2 (50) 30 (32) 2 (75) 1 (100)
Heart failure, n (%) 15 (33) 14 (28) 2 (50) 21 (22) 2 (75) 1 (100)
LVEF (%) 59 ± 9 62 ± 8 43 ± 22 61 ± 10 42 ± 17 40
CHA2DS2-Vasc score 2 (2–4) 2 (0–3) 3 (3–3) 2 (1–4) 5 (4–5) 6
RF time (s) or PF applications 1501 ± 540 50 ± 10 58 ± 16 38 ± 7 50 ± 11 32
Change in serum creatinine (mg/dL) −0.1 ± 0.11 −0.08 ± 0.07 0.38 ± 0.06 −0.07 ± 0.07 0.37 ± 0.1 0.21
Free plasma Hb (mg/L) 189 ± 181 335 ± 211 943 ± 1106 477 ± 330 756 ± 201 1600
Haptoglobin (g/L) 1.1 ± 0.49 0.58 ± 0.36 0.02 ± 0.01 0.82 ± 0.51 0.44 ± 0.16 0.2
Total bilirubin (mg/dL) 0.83 ± 0.38 1.4 ± 0.65 2.75 ± 0.9 1.07 ± 0.6 2.04 ± 1.54 3.9
LDH (U/L) 211 ± 36 215 ± 37 322 ± 75 261 ± 51 362 ± 83 403
Haemoglobinuria, n (%) 3 (7) 5 (10) 4 (100) 6 (6) 3 (100) 1 (100)

Data are means ± standard deviation or median (IQR). *, **, and *** indicate P < 0.05, P < 0.01, or P < 0.001, respectively, as compared with PVI-only. For clarity, only statistically significant P-values are indicated.

AHT, arterial hypertension; AKI, acute kidney injury; CAD, coronary artery disease; CKD, chronic kidney disease; DCM, dilated cardiomyopathy; DM, diabetes mellitus; eGFR, estimated glomerular filtration rate; Hb, haemoglobin; HCM, hypertrophic cardiomyopathy; HFmrEF, heart failure with mid-range ejection fraction; HFpEF, heart failure with preserved ejection fraction; HFrEF, heart failure with reduced ejection fraction; LDH, lactate dehydrogenase; LAD, left atrial diameter; LAVI, left atrial volume index; LVEF, left ventricular ejection fraction; PAD, peripheral artery disease; PVI, pulmonary vein isolation; SHD, structural heart disease; TCMP, tachycardiomyopathy; VHD, vascular heart disease.

Regardless of ablation modality, total haemoglobin was significantly (P < 0.0001) reduced post-ablation without significant differences in the PVI-only and PVI-plus groups (Figure 1A). No patient required transfusion. However, haptoglobin was significantly decreased after PFA (P < 0.0001) and not following RFA. Twenty-four hours after ablation with either PF modality, this was most pronounced for PVI-plus as compared with PVI-only (P < 0.0001). Additionally, reticulocyte count, free plasma haemoglobin, total bilirubin, and lactate dehydrogenase significantly increased post-ablation for the variable loop and pentaspline PFA catheters (P < 0.0001), but not for RF. This was more so in the PVI-plus group as compared with the PVI-only group (P-values ranging from 0.0242 to 0.0031 for variable loop and all P < 0.0001 for pentaspline PFA). Twenty-four hours after PFA, a significant proportion of patients showed haemoglobinuria (3%, 15%, and 11% following RF, variable loop, or pentaspline PVI-only ablation, respectively, and 7%, 29%, and 33% for PVI-plus ablation) (Figure 1A). No significant differences in IH were observed between the two PFA catheters.

Figure 1.

Figure 1

(A) Pre- and 24 h post-ablation blood sample and urine analysis for RFA (n = 46 patients), variable loop PFA (n = 54 patients), and pentaspline PFA (n = 98 patients). For each graph: RFA is indicated on the left (black), variable loop PFA in the middle (red) and pentaspline PFA on the right (blue). PVI-only is indicated more transparently, while PVI-plus is indicated in full color. For clarity, only statistically significant P-values are shown on the graphs. (B) Parts of whole graphs indicating patients with and without AKI following RFA, variable loop PFA, and pentaspline PFA. AKI, acute kidney injury; eGFR, estimated glomerular filtration rate; LDH, lactate dehydrogenase; PFA, pulsed field ablation; RFA, radiofrequency ablation.

Globally, serum creatinine decreased and estimated glomerular filtration rate (eGFR) improved post-ablation in all groups (P < 0.0001). However, using PFA, eight patients developed AKI (four Stage 1 following variable loop PFA, three Stage 1, and one Stage 3 following pentaspline PFA; Figure 1B). No AKI was observed following RFA (P = 0.04 and 0.007 vs. variable loop and pentaspline PFA, respectively). No significant differences in renal outcomes were observed between the variable loop and pentaspline PFA catheter (P = 0.26). Acute kidney injury occurred more in older patients, with known CKD, diabetes mellitus, and arterial hypertension, with SHD and more depressed LVEF and with a higher number of PF applications and more pronounced haemolysis (Table 1). Stage 3 AKI occurred in one patient with known advanced CKD (baseline eGFR 14 mL/min) following PVI-only (32 PF applications, no pulmonary vein angiograms) pentaspline PFA. Following ablation, he developed progressive renal injury (serum creatinine from 3.58 to 3.79 mg/dL) with hyperkalaemia and was initiated on RRT. All patients recovered back to baseline kidney function, except for the patient with Stage 3 AKI.

Our study confirms (i) that peri-procedural haemolysis occurs in a significant proportion of patients following PFA, (ii) that it is more pronounced when PVI is associated with additional lesion sets, and (iii) that it confers a risk of AKI.5,7,8 We add that this effect occurs regardless of the type of PFA catheter used. This effect was not observed following RFA and contrasts a recent publication by Osmancik et al.9 where borderline changes in haemolysis parameters were also observed following RFA. In the same publication, 1 out of 47 patients treated with PFA showed AKI Stage 1 (2.13%), which is in line with our reported 5.26% (8 out of 152 PFA cases). We add that AKI patients are generally older, more likely received PVI-plus ablation, and have CKD, heart failure, diabetes mellitus, and arterial hypertension. Our findings corroborate that prognosis is good, except for advanced CKD. However, where previous studies emphasized a larger number of PF applications necessary for renal injury, our data suggest that even a standard number of applications may provide deleterious effects. As such, additional factors such as tissue contact and the possible effect of PV angiograms on post-procedural kidney function may be essential in these patients.10

In conclusion, PFA-associated renal injury exists with both the pentaspline and variable loop PFA catheter. In the setting of advanced CKD, progression of kidney failure may occur, also with a limited number of PF applications.

Contributor Information

Maarten A J De Smet, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Clara François, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Benjamin De Becker, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Rene Tavernier, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Jean-Benoît le Polain de Waroux, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Sébastien Knecht, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Mattias Duytschaever, Department of Cardiology, AZ Sint-Jan Hospital Bruges, Ruddershove 10, B-8000 Brugge, Belgium.

Funding

None declared.

Data availability

The data and analysis methods that support the findings of this study are available from the corresponding author upon reasonable request.

References

  • 1. Hindricks  G, Potpara  T, Dagres  N, Arbelo  E, Bax  JJ, Blomström-Lundqvist  C  et al.  2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS) the task force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J  2021;42:373–498. [DOI] [PubMed] [Google Scholar]
  • 2. Tzeis  S, Gerstenfeld  EP, Kalman  J, Saad  E, Shamloo  AS, Andrade  JG  et al.  2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation. Europace  2024;26:euae043. [DOI] [PubMed] [Google Scholar]
  • 3. Grimaldi  M, Di Monaco  A, Gomez  T, Berman  D, Datta  K, Sharma  T  et al.  Time course of irreversible electroporation lesion development through short- and long-term follow-up in pulsed-field ablation-treated hearts. Circ Arrhythmia Electrophysiol  2022;15:e010661. [DOI] [PubMed] [Google Scholar]
  • 4. Reddy  VY, Gerstenfeld  EP, Natale  A, Whang  W, Cuoco  FA, Patel  C  et al.  Pulsed field or conventional thermal ablation for paroxysmal atrial fibrillation. N Engl J Med  2023;389:1660–71. [DOI] [PubMed] [Google Scholar]
  • 5. Venier  S, Vaxelaire  N, Jacon  P, Carabelli  A, Desbiolles  A, Garban  F  et al.  Severe acute kidney injury related to haemolysis after pulsed field ablation for atrial fibrillation. Europace  2023;26:euad371. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Mohanty  S, Casella  M, Compagnucci  P, Torlapati  PG, Della Rocca  DG, La Fazia  VM  et al.  Acute kidney injury resulting from hemoglobinuria after pulsed-field ablation in atrial fibrillation: is it preventable?  JACC Clin Electrophysiol  2024;10:709–15. [DOI] [PubMed] [Google Scholar]
  • 7. Ekanem  E, Neuzil  P, Reichlin  T, Kautzner  J, van der Voort  P, Jais  P  et al.  Safety of pulsed field ablation in more than 17,000 patients with atrial fibrillation in the MANIFEST-17 K study. Nat Med  2024;30:2020–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Ekanem  E, Reddy  VY, Schmidt  B, Reichlin  T, Neven  K, Metzner  A  et al.  Multi-national survey on the methods, efficacy, and safety on the post-approval clinical use of pulsed field ablation (MANIFEST-PF). Europace  2022;24:1256–66. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Osmancik  P, Bacova  B, Herman  D, Hozman  M, Fiserova  I, Hassouna  S  et al.  Periprocedural intravascular hemolysis during atrial fibrillation ablation: a comparison of pulsed field with radiofrequency ablation. JACC Clin Electrophysiol  2024;10:1660–71. [DOI] [PubMed] [Google Scholar]
  • 10. Nies  M, Koruth  JS, Mlček  M, Watanabe  K, Tibenská  VC, Královec  Š  et al.  Hemolysis after pulsed field ablation: impact of lesion number and catheter-tissue contact. Circ Arrhythm Electrophysiol  2024;17:e012765. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data and analysis methods that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Europace are provided here courtesy of Oxford University Press

RESOURCES