. 2024 Aug 3;27(4):825–830. doi: 10.1007/s40477-024-00941-5

Table 2.

Concordance rate between handheld ultrasound device (HHUSD) and high-end ultrasound device (HEUSD) in B-lines detection was expressed in mean ± standard deviation. Concordance rates between the two devices for the dection of pleural effusion and lung consolidations and for inferior cava vein (ICV) assessment were expressed as number and their percentages

 Ultrasound findings detected Concordance between HHUSD and HEUSD
Total enrolled population Patients with BMI > 30 kg/m2
B-lines 79.3% ± 17.7 78.9% ± 12.6
Pleural effusion 88.6% 86.5%
Lung consolidations 82.3% 79.5%
Presence/absence of ICV ectasia 88.7% 86.3%
Respiratory excursions of the ICV (> or < 50% of the diameter) 84.9% 85.8%