Table 2.

Ground truth of in-stent stenosis.

Stent types Tube diameter Stent diameter Lumen measured Plaque measured Strut thickness True lumen stent True lumen stenosis True plaque Stenosis (%)
Chrono 2.80 2.80 0.90 1.90 0.080 2.640 0.820 1.82 68.94
Coroflex Please 2.80 2.80 1.20 1.60 0.120 2.560 1.080 1.48 57.81
Endeavor 2.80 2.80 0.90 1.90 0.091 2.618 0.809 1.81 69.10
Promus Element Plus 3.00 3.00 1.00 2.00 0.081 2.838 0.919 1.92 67.62
Radius 2.80 2.80 1.30 1.50 0.085 2.630 1.215 1.42 53.80
Tantal Coronary 2.60 2.60 0.90 1.70 0.058 2.484 0.842 1.64 66.10

All measurements given in (mm), except the relative diameter stenosis (%). The stent diameter, lumen, and plaque were measured angiographically including the stent strut; the true lumen and true plaque diameter are equal to the measured diameter minus the strut thickness.