Table 2.

Within- and between-run precision and accuracy for quantification of total and unbound lenalidomide in human plasma.

Nominal
Concentration
(ng/mL)
Total Lenalidomide Unbound Lenalidomide
Predicted
Concentration
(Mean ± SD)
(ng/mL)
Precision
(CV, %) a
Accuracy
(%) b
Predicted
Concentration
(Mean ± SD)
(ng/mL)
Precision
(CV, %) a
Accuracy
(%) b
Within-run accuracy and precision (n = 5 sample replicates)
5 5.29 ± 0.17 3.29 105.86 4.61 ± 0.19 4.06 92.14
15 15.20 ± 0.31 2.04 101.36 14.39 ± 0.72 5.01 95.91
300 293.75 ± 1.89 0.65 97.92 298.26 ± 5.51 1.85 99.42
800 744.01 + 12.09 1.62 93.00 776.20 ± 5.63 0.73 97.02
Between-run accuracy and precision (n = 3 runs with 5 replicates/run)
5 5.06 ± 0.39 7.65 101.10 4.92 ± 0.52 10.55 98.48
15 14.97 ± 0.48 3.22 99.80 14.09 ± 0.82 5.80 93.95
300 297.42 ± 5.04 1.70 99.14 292.15 ± 6.59 2.26 97.38
800 755.59 ± 15.45 2.05 94.45 763.95 ± 15.15 1.98 95.49

a CV (%) = (standard deviation of the predicted concentrations/nominal concentration) × 100. b Accuracy (%) = (predicted concentration/nominal concentration) × 100.