. 2024 Oct 8;16(10):1581. doi: 10.3390/v16101581

Table 3.

Overall concordance rate between the PA method with a 4× antibody titer as the cutoff value and CLEIA (Lumipulse HTLV-I/II).

PA_SERO Total
Positive (≥4×) Negative (<4×)
CLEIA_LU Positive (COI ≥ 1.0) 85 5 90
Negative (COI < 1.0) 0 2 2
Total 85 7 92
Overall concordance rate 94.6%

PA, particle agglutination; CLEIA, chemiluminescence enzyme immunoassay; COI, cutoff index.