Skip to main content
. 2024 Oct 8;16(10):1581. doi: 10.3390/v16101581

Table 4.

Overall concordance rate between the PA method with a 16× antibody titer as the cutoff value and CLEIA (Lumipulse HTLV-I/II).

PA_SERO Total
Positive (≥16×) Negative (<16×)
CLEIA_LU Positive (COI ≥ 1.0) 73 17 90
Negative (COI < 1.0) 0 2 2
Total 73 19 92
Overall concordance rate 81.5%