Table 6.

Cutoff values and their diagnostic accuracy for each test kit for diagnosing HAM.

SERO LU LU-P HISCL UD Abbott Elecsys ESPLINE INNO-LIA
Cutoff values for blood Titer
16×
COI
1.0
COI
1.0
COI
1.0
COI
1.0
S/CO
1.0
COI
1.0
* *
Sensitivity 97.1% 100% 100% 84.1% 100% 100% 100% 100% 91.3%
Specificity 73.9% 8.7% 13.0% 91.3% 13.0% 18.2% 34.8% 30.4% 65.2%
Cutoff values equivalent to the 4× titer of the PA method Titer
COI
4.7
COI
5.4
COI
0.15
COI
3.3
S/CO
2.8
COI
2.3
n/a n/a
Sensitivity 100% 100% 98.6% 98.6% 100% 98.6% 100% n/a n/a
Specificity 30.4% 52.2% 47.8% 34.8% 43.5% 54.5% 43.5% n/a n/a
Cutoff values for maximum specificity with 100% sensitivity Titer
COI
4.7
COI
4.7
COI
0.25
COI
3.3
S/CO
2.5
COI
2.5
n/a n/a
Sensitivity 100% 100% 100% 98.6% ** 100% 100% 100% n/a n/a
Specificity 30.4% 52.2% 47.8% 39.1% 43.5% 54.5% 56.5% n/a n/a

COI, cutoff index; S/CO, signal-to-cutoff ratio; n/a, not applicable. * Based on the manufacturer’s instructions. ** Sensitivity did not reach 100% because one patient with HAM had COI = 0.