Skip to main content
. 2024 Oct 20;16(10):1638. doi: 10.3390/v16101638

Table 3.

Influenza A: diagnostic performance of ID NOW assay.

RT-qPCR
Positive (%) Negative (%) Total (%)
ID NOW Whole population (n = 194)
Positive (%) 27 (13.9) 0 (0.0) 27 (13.9)
Negative (%) 3 (1.5) 164 (84.5) 167 (86.1)
Total (%) 30 (15.5) 164 (84.5) 194 (100)
0–3 years (n = 101)
Positive (%) 13 (12.9) 0 (0.0) 13 (12.9)
Negative (%) 3 (3.0) 85 (84.2) 88 (87.1)
Total (%) 16 (15.8) 85 (84.2) 101 (100)
>3 years (n = 93)
Positive (%) 14 (15.1) 0 (0.0) 14 (15.1)
Negative (%) 0 (0.0) 79 (84.9) 79 (84.9)
Total (%) 14 (15.1) 79 (84.9) 93 (100)
Test performances
% 95% CI
Whole population (n = 194)
PA 90.0 73.5–97.9
NA 100 97.8–100
PPV 100 87.2–100
NPV 98.2 94.9–99.3
LR+ N/A
LR- 0.1 0.03–0.29
Accuracy 98.4 95.5–99.7
0–3 years (n = 101)
PA 81.3 54.4–96.0
NA 100 95.8–100
PPV 100 75.3–100
NPV 96.6 91.1–98.7
LR+ N/A
LR− 0.2 0.1–0.5
Accuracy 97.0 91.6–99.4
>3 years (n = 93)
PA 100 76.8–100
NA 100 95.4–100
PPV 100 76.8–100
NPV 100 95.4–100
LR+ N/A
LR− 0.0
Accuracy 100 96.1–100

N/A: not available.