Skip to main content
. 2024 Oct 20;16(10):1638. doi: 10.3390/v16101638

Table 4.

Influenza B: diagnostic performance of ID NOW assay.

RT-qPCR
Positive (%) Negative (%) Total (%)
ID NOW Whole population (n = 194)
Positive (%) 27 (13.9) 1 (0.5) 28 (14.4)
Negative (%) 0 (0.0) 166 (85.6) 166 (85.6)
Total (%) 27 (13.9) 167 (86.1) 194 (100)
0–3 years (n = 101)
Positive (%) 3 (3.0) 0 (0.0) 3 (3.0)
Negative (%) 0 (0.0) 98 (97.0) 98 (97.0)
Total (%) 3 (3.0) 98 (97.0) 101 (100)
>3 years (n = 93)
Positive (%) 24 (25.8) 1 (1.1) 25 (26.9)
Negative (%) 0 (0.0) 68 (73.1) 68 (73.1)
Total (%) 24 (25.8) 69 (74.2) 93 (100)
Test performances
% 95% CI
Whole population (n = 194)
PA 100 87.2–100
NA 99.4 96.7–100
PPV 96.4 79.3–99.5
NPV 100 97.8–100
LR+ 167 23.7–1178.6
LR− 0.0
Accuracy 99.5 97.2–99.9
0–3 years (n = 101)
PA 100 29.2–100
NA 100 96.3–100
PPV 100 29.2–100
NPV 100 96.3–100
LR+ N/A
LR− 0.0
Accuracy 100 96.4–100
>3 years (n = 93)
PA 100 85.8–100
NA 98.6 92.2–100
PPV 96.0 77.4–99.4
NPV 100 94.7–100
LR+ 69.0 9.9–482.9
LR− 0.00
Accuracy 98.9 94.1–99.9