Skip to main content
. 2024 Oct 31;14:26222. doi: 10.1038/s41598-024-75216-4

Table 2.

System suitability parameters of the proposed chromatographic method.

Parameters PAR DEX RIV Reference Value37.
Retention time (tR) 3.19 ± 0.01 7.02 ± 0.01 7.81 ± 0.01
Capacity factor ( K´) 1.13 3.68 4.21 1 < K < 10
Resolution (Rs) 2.68 6.66 Rs ≥ 1.50
Tailing factor (T) 1.01 1.04 1.02 0.80–1.20
Selectivity factor (α) - 3.26 1.14 α > 1.00
Theoretical plates number (N) 5821 53,077 66,656 > 2000