Published in final edited form as: Med Phys. 2023 Jul 10;51(2):1364–1382. doi: 10.1002/mp.16585

TABLE 1.

Angular separation between projections of simulated and patient acquired scans.

Projections Acquisition Average angular separation (°) Maximum angular separation (°) Angular RMSE
200 Ideal Angular Separation 10.00 10.00 N/A
200 Constant Velocity 9.2 s 6.82 ± 18.14 61.02 ± 25.87 19.58 ± 7.37
200 Constant Velocity 60 s 9.37 ± 9.53 31.49 ± 15.44 9.87 ± 6.14
200 Constant Velocity 120 s 9.39 ± 9.38 32.93 ± 14.46 10.03 ± 5.82
200 Constant Velocity 240 s 9.27 ± 9.78 33.97 ± 15.42 10.62 ± 6.88
200 Variable Angular Gap 20° 9.69 22.19 2.58
200 Variable Angular Gap 30° 9.57 32.43 4.93
200 Variable Angular Gap 40° 9.61 40.85 7.07
200 Static Angular Gap 20° 10.46 20.52 2.28
200 Static Angular Gap 30° 10.99 30.52 4.51
200 Static Angular Gap 40° 11.51 40.52 6.78
200 ADAPT Patient Mean 9.95 ± 0.10 14.80 ± 3.15 1.72 ± 1.17