Abstract
Study Objectives:
From 2019–2023, the United States Food and Drug Administration has cleared 9 novel obstructive sleep apnea-detecting wearables for home sleep apnea testing, with many now commercially available for sleep clinicians to integrate into their clinical practices. To help clinicians comprehend these devices and their functionalities, we meticulously reviewed their operating mechanisms, sensors, algorithms, data output, and related performance evaluation literature.
Methods:
We collected information from PubMed, United States Food and Drug Administration clearance documents, ClinicalTrials.gov, and web sources, with direct industry input whenever feasible.
Results:
In this “device-centered” review, we broadly categorized these wearables into 2 main groups: those that primarily harness photoplethysmography data and those that do not. The former include the peripheral arterial tonometry-based devices. The latter was further broken down into 2 key subgroups: acoustic-based and respiratory effort-based devices. We provided a performance evaluation literature review and objectively compared device-derived metrics and specifications pertinent to sleep clinicians. Detailed demographics of study populations, exclusion criteria, and pivotal statistical analyses of the key validation studies are summarized.
Conclusions:
In the foreseeable future, these novel obstructive sleep apnea-detecting wearables may emerge as primary diagnostic tools for patients at risk for moderate-to-severe obstructive sleep apnea without significant comorbidities. While more devices are anticipated to join this category, there remains a critical need for cross-device comparison studies as well as independent performance evaluation and outcome research in diverse populations. Now is the moment for sleep clinicians to immerse themselves in understanding these emerging tools to ensure our patient-centered care is improved through the appropriate implementation and utilization of these novel sleep technologies.
Citation:
Chiang AA, Jerkins E, Holfinger S, et al. OSA diagnosis goes wearable: are the latest devices ready to shine? J Clin Sleep Med. 2024;20(11):1823–1838.
Keywords: wearable, sleep technology, obstructive sleep apnea (OSA), home sleep apnea testing (HSAT), apnea-hypopnea index (AHI), photoplethysmography (PPG), peripheral arterial tonometry (PAT), acoustic, accelerometry, artificial intelligence (AI)
BRIEF SUMMARY
Current Knowledge/Study Rationale: The landscape of sleep medicine is undergoing swift transformations, driven by the advancement of sleep technologies, with wearable technologies at the forefront. From 2019–2023, the United States Food and Drug Administration cleared 9 novel obstructive sleep apnea-detecting wearables for home sleep apnea testing, further solidifying their role in the field.
Study Impact: In this review, we present comprehensive details of these wearables pertinent to sleep clinicians, including key validation study results, considerations when appraising performance evaluation studies, and critical insights into the clinical implementation of novel technologies. This review offers sleep clinicians vital discernment to ensure they grasp the nuances of these emerging tools, navigate their complexities, and establish readiness for the ongoing evolution of sleep technologies.
INTRODUCTION
Sleep is increasingly recognized as a multifaceted and integral component of overall health.1–3 Notably, obstructive sleep apnea (OSA), one of the most prevalent sleep disorders, represents a significant challenge to both individual and global health, given its established associations with cardiovascular and metabolic comorbidities.4,5 In 2016, the estimated prevalence of OSA in the United States adult population was 12%, affecting 29.4 million individuals, with nearly 80% of cases remaining undiagnosed.6–8 There are also considerable costs associated with undiagnosed OSA, including indirect expenses, estimated to reach $149.6 billion.6,7 Further, recent longitudinal data indicates a rise in OSA prevalence over the past 2 decades.9,10 As OSA awareness and prevalence continue to increase, the demand for efficient, affordable, precise, and validated diagnostic sleep medicine services will likewise rise.11–13 With the growing focus on sleep health and the pursuit of a “good night’s sleep” by millions, the market for commercial and medical-grade wearables for sleep monitoring is expanding rapidly.14–20
The current diagnosis of OSA primarily relies on traditional in-lab overnight polysomnography (PSG) or home sleep apnea testing (HSAT).21 Both diagnostic methods for OSA present their own set of challenges and both approaches offer a 1-night OSA assessment compared to novel wearables that offer the advantage of multinight assessment.21–22 From 2019–2023, a total of 9 novel wearables were cleared by the United States Food and Drug Administration (FDA) for home diagnosis of adult OSA.23–31 Table 1 lists the names and FDA clearance dates of these 9 devices.
Table 1.
| Categories | Device (Date of FDA Clearance) | Sensors, Sensing Location, and Operating Mechanisms | SCOPER33 |
|---|---|---|---|
| PPG Signals | |||
| PPG/PAT-based | NightOwl (3/6/2020) | Fingertip PAT (no probe) with accelerometry; machine learning/deep learning-based algorithms for respiratory event detection and sleep staging. | S3C2O1xP0E0R0 |
| ANNE Sleep* (7/11/2022) | Fingertip PAT (no probe), chest sensor with ECG, and accelerometry; machine learning for sleep staging and manual scoring for respiratory event identification. | S3C2O1xP2E4R0 | |
| Somfit# (11/30/2023) | Forehead PAT (no probe) with frontal EEG/EOG × 3, accelerometry, and microphone; deep learning for sleep staging and rule-based algorithm for respiratory event detection. | S2C2O2P0E0R0 | |
| Belun Ring‡ (2/23/2023) | Proximal phalanx PPG with accelerometry; deep learning neural networks for both sleep staging and respiratory event detection. | S3C4O2P0E0R0† | |
| SleepImage SaMD (8/14/2019) | Proximal phalanx (ring) or fingertip PPG with accelerometry; rule-based algorithms leveraging cardiopulmonary coupling for respiratory event detection and unconventional sleep stage classification. | S3C4O2P0E0R0† (ring) and S3C4O1xP0E0R0 (fingertip) | |
| Non-PPG Signals | |||
| Acoustic-based | AcuPebble Ox100 (2/3/2023) | Proprietary piezoelectric microphone at the front of the neck capturing respiratory, cardiac, and movement sounds; fingertip pulse oximeter (Ox100); rule-based algorithms for respiratory event detection; no sleep staging. | S0C4O1xP0E0R5 (Ox100) |
| BresoDX1 (12/21/2022) | Suprasternal notch microphone capturing tracheal sound with accelerometry sensing trachea-sternal motion, and fingertip pulse oximeter; deep neural networks for respiratory event detection; no sleep staging. | S3C4O1xP2E0R5 (with pulse oximeter added) | |
| Effort-based | Sunrise2 (12/22/2022) | Chin accelerometry and gyroscope measuring mandibular movement; incorporated thermistor and pulse oximeter in the chin sensor. Machine learning algorithms for sleep staging and respiratory event detection. | S3C4O2P2E4R3 |
| Wesper Lab (12/21/2021) | Chest and abdomen RIP, accelerometry, embedded recorder, and fingertip pulse oximeter; no AI algorithms described in the published articles or the FDA 510(k) premarket notifications. | S3C4O1xP2E4R5 (with pulse oximeter added) | |
Auto-scoring of respiratory events is reportedly underway. #No published validation study as of this writing. ‡Belun Cor body sensor that detects respiratory effort and body position is currently under investigation. †It remains unclear whether novel pulse oximetry at the proximal phalanx should be categorized at O1/O1x or O2. AI = artificial intelligence, ECG = electrocardiogram, EEG = electroencephalogram, EOG = electrooculogram, FDA = United States Food and Drug Administration, OSA = obstructive sleep apnea, PAT = peripheral arterial tonometry, PPG = photoplethysmography, RIP = respiratory inductance plethysmography, SaMD = software as a medical device, SCOPER = sleep, cardiovascular, oximetry, position, effort, and respiratory.
The rapid emergence of OSA-detecting wearable platforms brings forth both opportunities and challenges to the sleep science community.15–17 Newer OSA diagnostic technologies, aimed to overcome the limitations of PSG and HSAT, utilize innovative sensing structures, advanced data analysis methods, and less intrusive monitoring approaches.15–17 Additionally, they hold the potential for offering continuous and cost-effective data monitoring, generating longitudinal data that could provide insights into individual health and advance the understanding of global sleep health.2,15–19 However, as with any new technology introduced, there are caveats to consider. Many use artificial intelligence (AI)/machine learning (ML)/deep learning (DL) approaches that perform tasks by learning from an underlying distribution of data. As data sourced from diverse age groups, racial/ethnic backgrounds, genders, and healthy vs clinical populations may evolve, a well-trained and described model must be sustainable in real-world scenarios.32
In this review, we primarily focus on the FDA-cleared wearables to provide insights and literature review. This “device-centered” review encompasses details, including FDA status, mechanisms; sensors; sleep, cardiovascular, oximetry, position, effort, and respiratory parameters categorization; algorithms; data output; performance literature review; and comments on the strengths and limitations of each device.33 Notably, while the respiratory event index might be a preferred term for HSAT, we opted to utilize apnea-hypopnea index (AHI) and not respiratory event index in this manuscript due to the prevailing use of AHI in both the wearable research publications and marketing materials/websites of commercialized devices.33 Also, most novel wearables, although without electroencephalogram (EEG), do offer total sleep time (TST), sleep staging, and AHI at 3% desaturation (AHI3%) through various sensors and proprietary algorithms. To simplify understanding and avoid confusion by renaming each device’s specific output, these surrogate outputs will be referred to as the familiar TST, sleep stages, and AHI.
Additionally, it is crucial to note that the hardware and software of these devices are updated frequently, and the information provided reflects the best available knowledge at the time of manuscript submission, gathered from peer-reviewed publications, FDA documents, ClinicalTrials.gov, and web sources, with direct industry inputs whenever feasible.34–36 Lastly, it is worth noting that although several authors have past or present involvement with the American Academy of Sleep Medicine (AASM) Emerging Technology Committee, the viewpoints expressed herein do not necessarily reflect those of the AASM.
FDA clearance of Class II OSA-detecting wearables and software as a medical device (SaMDs)
All the wearables discussed in this paper have attained Class II FDA clearance. Most underwent 510(k) FDA clearance, which mandates that the device be equivalent to a “predicate” technology already classified as Class II.37 The only exception is Sunrise, which was granted Class II through the de novo pathway, as there was no similar predicate available for comparison.38 Class II technologies necessitate a prescription and are considered medical-grade. Notably, there exists a distinct FDA category for software algorithms used for medical purposes, known as SaMD.39 SleepImage is an example.
Classification of novel OSA-detecting wearables
In conventional sleep testing, the predominant methodologies involve thermistor and pressure transducer airflow measurements, with a notable exception of WatchPAT (ZOLL-Itamar, Atlanta, Georgia).21,33 Conversely, the novel OSA-detecting wearables do not directly gauge airflow. Instead, they utilize novel sensors and proprietary data processing algorithms to derive AHI surrogates. These emerging wearable systems can be broadly categorized based on their operating mechanisms and sensing modalities into 2 main groups: those that primarily harness photoplethysmography (PPG) data and those that do not.40 The non-PPG wearables can be further refined into 2 key subgroups: acoustic-based and respiratory effort-based devices. Table 1 shows our categorization of these 9 OSA-detecting wearables, with sleep, cardiovascular, oximetry, position, effort, and respiratory parameters included.33
PPG-based wearables
First conceptualized in the 1930s, PPG has found versatile applications in pulse oximetry and is now widely incorporated into consumer wearable sleep trackers.16–19,40,41 By employing PPG, a noninvasive optical technique, these wearables can monitor variations in pulsatile blood flow volume within the microvascular circulation.40–45 These applications encompass pulse rate (PR) and oxygen saturation (SpO2) measurement, alongside the derivation of sleep stages.40 Classic examples include Belun Ring and SleepImage SaMD.23,30
WatchPAT, a widely used PPG-based HSAT, leverages a pneumo-optical peripheral arterial tonometry (PAT) probe positioned at the distal digit to record the pulsatile volume fluctuations occurring in the arteries of peripheral vascular beds.46 WatchPAT captures vasoconstrictive responses triggered by bursts of sympathetic activity arising from respiratory events. WatchPAT achieves this by maintaining a consistent subdiastolic pressure via a PAT probe, optimizing the dynamic range of the PPG signal.46,47 This strategy unloads vascular wall tension, prevents venous pooling to mitigate the influence of confounding factors like veno-arteriolar reflex vasoconstriction, and minimizes artifacts stemming from respiration and movement, enhancing measurement accuracy.46,47 Alternatively, NightOwl, a novel FDA-cleared PAT-based wearable, derives PAT signals from PPG utilizing a miniaturized sensor accompanied by proprietary software algorithms for signal processing to counteract the venous pooling effect.24,48 Recently, the FDA also categorized ANNE Sleep and Somfit as PAT devices.28,31
For this review, PPG and PAT-based devices are grouped into a single discussion as the line between PPG and PAT has become increasingly blurred. Before NightOwl received FDA clearance as a PAT device, capturing PAT signals necessitated a PAT probe.46–48 Presently, ambiguity exists surrounding the FDA’s definition and criteria for PAT devices as what qualifies a PPG-based device as a PAT device is unclear. In the recent FDA Somfit 510(k) notification, it was acknowledged that “There is no standardized methodology for calculation of PAT and numerical values of the PAT signal are expected to differ based on electronic characteristics of the acquisition hardware, and other described factors.”31 Moreover, the specific methods employed by individual PAT devices to acquire or derive PAT signals remain proprietary, further complicating the differentiation between PPG- and PAT-based devices.
Acoustic-based wearables
AcuPebble and BresoDX1 employ acoustic sensors (microphones) positioned over the anterior neck to capture signals related to respiratory, cardiac, and movement sounds.25,29 Additionally, BesoDX1 integrated an accelerometer to detect trachea-sternal motion to improve AHI accuracy.
Respiratory effort-based wearables
Although Sunrise leverages ML algorithms to gauge mandibular movements (MMs), offering insights into the respiratory drive, Wesper Lab acquires respiratory inductance plethysmography (RIP) signals from patches affixed to the chest and abdomen to derive flow.26,27
OSA-detecting wearables vs flow-based type III HSATs
Conventional flow-based type III HSAT devices are considered acceptable for diagnosing adult OSA in those with a high pretest probability without significant comorbidities.21,49 As the TST is often unattainable, the AHI (or more precisely, respiratory event index) calculation typically employs total recording time (TRT), leading to an underestimation of OSA severity.21,22 Conversely, many OSA-detecting wearables estimate TST utilizing accelerometry, potentially offering more accurate AHI measurement. It is noteworthy that these wearables may overestimate or underestimate AHI depending on the AHI criteria (3% or 4%) and AHI thresholds used (5, 15, or 30 events/h). Different generations of algorithms may also exhibit various performance characteristics. Some may overestimate AHI within the low AHI range (0–15 events/h) and underestimate AHI within the high AHI range (> 30 events/h, particularly).50–52 The overestimation of AHI in the low AHI range may be attributed to the inclusion of respiratory effort-related arousals or autonomic arousals from other causes. Conversely, the underestimation in the high AHI range may stem from missed respiratory events during consecutive occurrences over a short timeframe or the exclusion of motion-related poor PPG signals in severe OSA.51,52 Further details regarding the disparities between type III HSATs, PPG-based wearables, and non-PPG-based wearables are summarized in Table 2.
Table 2.
Conventional type III flow-based HSAT system vs PPG-based wearables vs non-PPG wearables.
| Conventional Type III Flow-Based HSATs | PPG-Based Wearables | Non-PPG-Based Wearables | |
|---|---|---|---|
| Measurement sites | Nasal flow, effort, and fingertip pulse oximetry | Varies (fingertip, proximal phalanx, forehead) | Varies (anterior neck, chest/abdomen. mandible) |
| Pulse oximetry | Yes | Yes | Varies |
| Accelerometry | No (majority) | Yes | Most |
| TST and sleep stages | No (most) | Yes | Varies |
| CSA detection | Yes | Varies | Most.* Mechanisms vary. |
| Arousal detection | No | Yes (autonomic arousals) | Varies |
| Device over- or underestimate OSA | Generally underestimate AHI if TRT used for AHI calculation. | Varies. Auto-scoring may over- or underestimate AHI depending on the AHI criteria (3% or 4%) and AHI cutoffs employed.† | Varies. Auto-scoring may over- or underestimate AHI. |
| Body position | Varies | Varies. Needs body sensor | Most |
| Raw data/channel accessibility | Yes | Yes | Most |
| Respiratory event scoring method | Manual (may have the capability of auto-scoring). | Varies. Mostly auto-scoring with the capability of manual overscoring. | Mostly auto-scoring. Some may allow manual overscoring. |
| Multiple-night | Varies. Mostly single night. | Yes | Most |
| Disposable device | No, reusable only | Varies | Varies |
Many lack validation studies. †Different generations of algorithms may exhibit various performance characteristics. AHI = apnea-hypopnea index, CSA = central sleep apnea, EEG = electroencephalometry, HSAT = home sleep apnea testing, OSA = obstructive sleep apnea, PAT = peripheral arterial tonometry, PPG = photoplethysmography, TRT = total recording time, TST = total sleep time.
PPG-based wearables vs pulse oximeters
While some may perceive PPG-based OSA-detecting wearables as mere fancier pulse oximeters, they actually offer a broad array of features and functionalities, surpassing traditional pulse oximeters. These wearables employ either rule-based or AI/ML/DL algorithms to derive metrics such as AHI, sleep-wake states, TST, sleep onset latency, sleep efficiency (SE), wake after sleep onset (WASO), and sleep stages. Furthermore, many wearables can be worn over multiple nights and provide access to raw data as well as manual scoring of respiratory events.
REVIEW OF FDA-CLEARED OSA-DETECTING WEARABLES
Wearables using PPG signals
Table 3 outlines the key distinctions among the FDA-cleared PPG-based wearables.
Table 3.
Comparison among FDA-cleared novel PPG-based OSA-detecting wearables.
| NightOwl | ANNE Sleep | Somfit | Belun Ring | SleepImage SaMD | |
|---|---|---|---|---|---|
| FDA clearance | II, 510(k) | II, 510(k) | II, 510(k) | II, 510(k) | II, 510(k) |
| Intended use | OSA diagnosis | Moderate to severe OSA | OSA diagnosis | Moderate to severe OSA | OSA diagnosis |
| Pediatric use | No | No | No | No | Yes (age > 2) |
| Measuring site | Fingertip | Fingertip, chest | Forehead | Proximal phalanx | Proximal phalanx or fingertip |
| Sensors | PAT, Acc | PAT, Acc | PAT, EEG, Acc, microphone | PPG, Acc | PPG, Acc |
| AI/ML/DL algorithms | Yes, ML/DL (AHI, sleep staging) | Yes, DL (sleep-wake) | Yes, DL (sleep staging) | Yes, DL (sleep staging, AHI) | No, rule-based |
| Snoring | No | Yes | Yes | No | No |
| OSA detection | AHI3%, AHI4% (auto-scoring) | AHI4% (manual scoring) | AHI3%, AHI4% (auto-scoring), no published data | AHI4% (auto-scoring)* | AHI3%, AHI4% (auto-scoring) |
| CSA detection | Yes† | Yes, but no data published | No | No* | Yes |
| TST and sleep stages | Yes; limited published data | Yes; no published data | Yes; white paper only | Yes, published data | Yes, sleep staging unconventional |
| PRV data | No | No | No | Yes | No |
| Body sensor | No | Suprasternal | Forehead sensor | No* | No |
| Body position | No | Yes | Yes | No* | No |
| Raw data accessibility | Yes | Yes | Yes | Yes | Yes |
| Manual editing | Yes | Yes | Yes | Yes | Yes |
| Multiple-night | Yes | Yes | Yes | Yes | Yes |
| Disposable or reusable | Both (Disposable: NightOwl Mini) | Reusable | Reusable | Reusable | Reusable |
| Must pair with a smartphone | Yes | Utilizing a “smart charger” | Yes | With or without a smartphone | Yes |
Belun ring AHI3%, Belun CSA detection, and body position are currently under investigation. †CSA flagged, Europe only; not in the United States. Acc = accelerometry, AHI = apnea-hypopnea index, AI = artificial intelligence, CSA = central sleep apnea, CPC = cardiopulmonary coupling, DL = deep learning, EEG = electroencephalography, FDA = United States Food and Drug Administration, ML = machine learning, OSA = obstructive sleep apnea, PAT = peripheral arterial tonometry, PPG = photoplethysmography, PRV = pulse rate variability, SaMD = software as a medical device, TST = total sleep time.
NightOwl
Introduction: The NightOwl (Ectosense, Leuven, Belgium; now under ResMed) is an FDA-cleared Class II PAT wearable platform for the diagnosis of adult OSA.24,53 It notably stands as the pioneering wearable to utilize software algorithms for the derivation of PAT from PPG signals without using a PAT probe.48
Sensors and mechanisms: The NightOwl platform incorporates PPG and accelerometry sensors. This compact sensor, comparable in size to a fingertip, can be easily self-applied to a fingertip.51
Algorithms: The NightOwl system utilizes proprietary ML/DL-based algorithms. For detecting respiratory events, it employs an ensemble-of-trees classifier that analyzes changes in PAT, PR, and SpO2. For sleep staging, a neural network is used for sleep-wake classification, while an ensemble-of-trees classifier distinguishes between rapid eye movement (REM) and non-REM (NREM) sleep (personal communication, Frederik Massie).
Data output: The current NightOwl platform presents raw data from PAT, SpO2, PR, and motion (actigraphy), as well as analyzed output including TST, SE, WASO, sleep stages (REM, NREM, wake), AHI3% and AHI at 4% desaturation (AHI4%), oxygen desaturation index (ODI) at 3% and ODI at 4% (ODI4%), SpO2, and PR data.24,48
Performance literature review: Five distinct studies explored the validity and diagnostic accuracy of the NightOwl platform (Table S1 in the supplemental material). The first validation effort, spearheaded by Massie et al involved a comparative analysis between NightOwl and concurrent in-lab PSG, utilizing the AASM Rule 1A (3% rule). This investigation revealed a substantial agreement in OSA severity classification.51 Subsequently, Van Pee et al conducted a multicenter validation study involving 3 centers in Miami, Florida, and 1 in Belgium. Both AASM Rule 1A and Rule 1B (4% rule) were employed, with NightOwl achieving a moderate-to-substantial agreement with PSG-AHI. Notably, this study introduced an innovative concept of “near-boundary double-labeling” to address instances where AHI values approach category thresholds.48 More recently, Lyne et al conducted an independent evaluation of the NightOwl, assessing the performance of both the disposable NightOwl-Mini and the reusable NightOwl using AASM Rule 1A. Their analysis indicated a moderate agreement in the OSA severity categorization when compared to in-lab PSG.54 In addition, Massie et al explored REM OSA phenotyping based on NightOwl data, reporting a low sensitivity with a high specificity.55 The same research group also proposed a method to detect central events using NightOwl PPG data and did so without the need for any additional body sensor.56 While seemingly robust sensitivity and specificity were reported in CSA detection, the results lacked external validation from an independent, separate dataset and cannot be considered conclusive.
Comments: The NightOwl system offers both AHI3% and AHI4% with overall consistent accuracy and provides users with a choice between disposable (NightOwl Mini) and reusable devices.48,54 Its performance in detecting central events without a body sensor requires further validation.56 The NightOwl was initially launched by Ectosense in June 2020 but has not been formally relaunched since its acquisition by ResMed in October 2021.
ANNE sleep
Introduction: The ANNE Sleep system (Sibel Inc., Niles, Illinois) is a flexible dual-sensor wearable system that has earned clearance from the FDA as a Class II PAT HSAT device.57 This clearance granted ANNE Sleep as a tool for diagnosing moderate-to-severe OSA in adults.28
Sensors and mechanisms: The ANNE Sleep system is composed of 2 integral units: a chest unit and a limb unit. Nestled at the suprasternal notch, the chest unit houses a single-lead electrocardiography (ECG), a 3-axis accelerometer, and a temperature sensor.58 This amalgamation of sensors enables monitoring of sleep parameters, including snoring, heart rate, heart sounds, respiratory rate, chest wall movement, body position, and skin temperature. The limb unit on the fingertip employs PAT, SpO2, and peripheral skin temperature.
Algorithms: The ANNE Sleep’s sleep-wake classifier is based on a proprietary AI-enabled algorithm that integrates recurrent neural network and convolutional neural network. While the respiratory event scoring is a manual process, an exploration of the clinical viability of automated ANNE Sleep respiratory event scoring is allegedly underway.58
Data output: The ANNE Sleep platform offers a comprehensive array of channels from PAT, SpO2, PR, heart rate, snoring, body position, and respiratory effort, along with output including TST, SpO2 data, PR data, and the manually scored AHI.28,58 The measurement of pulse transit time from ECG and PPG is a noteworthy inclusion, which can potentially aid in distinguishing between obstructive and central events.59
Performance literature review: The validation literature on ANNE Sleep is limited, with just a single published study to date (Table S2 in the supplemental material). Davies et al compared ANNE Sleep to concurrent in-lab PSG in a multicenter study and demonstrated impressive sensitivity and specificity at PSG-AHI cutoff of 15 events/h using manually scored AHI4%.58 AHI3% data were not reported.
Comments: Presently, ANNE Sleep stands as the sole PPG-based wearable system incorporating a body sensor, which provides many important physiological metrics including ECG and pulse transit time. While the accessibility of respiratory effort and the derived pulse transit time can assist in the distinction between central and obstructive events, its efficacy remains unconfirmed. Furthermore, while employing DL for sleep-wake determination, the accuracy in classifying sleep-wake states or sleep stages has yet to be established in adult patients with OSA.
Somfit
Introduction: The Somfit (Compumedics Limited, Abbotsford, Australia) represents the latest PPG-based wearable addition to adult home OSA diagnosis.31 The FDA clearance enables the device to be utilized for OSA diagnosis, sleep staging (stages N1, N2, N3, REM, wake), and snoring level assessment.31
Sensors and mechanisms: Somfit captures data from the forehead utilizing 3 frontal electrodes, a triaxial accelerometer, a microphone, and PPG.31 Its forehead reflectance oximetry is similar to SomnaPatch (Respironics, Monroeville, Pennsylvania).31 It also leverages frontal Fp1 and Fp2 EEG/electrooculogram to determine sleep stages.31 Similar to NightOwl, Somfit does not employ a PAT probe.
Algorithms: The Somfit sleep staging algorithm, utilizing 3 channels (differential frontal EEG and approximations of left and right electrooculograms), is a proprietary DL algorithm based on a fully convolutional multilayer neural network. The automated PAT respiratory events scoring is rule-based (per communication with Compumedics).
Data output: The Somfit platform presents channels from EEG, PAT, SpO2, PR, snoring, and motion, alongside respiratory and sleep metrics such as AHI, ODI, snoring level, and sleep stages.31
Performance literature review: While a recent cross-device study evaluated the validity of Somfit in assessing sleep-wake state in young healthy individuals, Somfit currently has no peer-reviewed publication concerning OSA diagnosis to our knowledge.60 According to the FDA 510(k), a clinical study was conducted comparing Somfit with WatchPAT-300 against PSG (Table S3 in the supplemental material).31 The reported difference between Somfit-AHI and WatchPAT-AHI was small, indicating noninferiority The Somfit sleep staging performance appears to be more accurate than WatchPAT-300, likely due to the use of EEG.31 No other statistics measures concerning OSA diagnosis performance including accuracy, sensitivity, specificity, or confusion matrix were included in the 510(k) document.31
Comments: Somfit stands out as the only FDA-cleared OSA-detecting wearable that detects EEG and is worn on the forehead. Somfit lacks respiratory effort monitoring and does not distinguish between obstructive and central apnea.31 Presently, very limited information is available for Somfit, necessitating future reassessment.
Belun Ring
Introduction: The Belun Ring (also known as Belun Sleep System BLS-100, Belun Technology Company Limited, Hong Kong) was recently cleared by the FDA as a Class II HSAT.30,61 This clearance allows the device to be employed for diagnosing moderate-to-severe OSA and sleep stage determination.30
Sensors and mechanisms: The Belun Ring, in its current iteration, is equipped with PPG and accelerometry. This ring-shaped wearable sensor, placed on the proximal phalanx of the nondominant hand index finger, captures PPG signals from the phalangeal radialis indicis artery.
Algorithms: The Belun Ring is complemented by its second-generation (Gen2) software incorporating 2 proprietary DL neural network algorithms, designed to facilitate respiratory events detection and sleep staging.62 The respiratory event detection algorithm was developed using a convolutional neural network with enhanced representational power to estimate AHI based on SpO2 and PR input. The model was trained using 2 independent datasets including the Sleep Heart Health Study open-source dataset.52,62 The sleep staging algorithm utilizes SpO2, PR, and accelerometry signals to learn features and patterns. Its structure comprises convolutional neural network for feature extraction and Transformer-based recurrent neural network for comprehensive classification.62
Data output: The current Belun platform display presents channels including PPG, SpO2, PR, motion, and derived sleep stages. The data output encompasses TRT, TST, SE, WASO, sleep stages (wake, REM, and NREM), AHI4%, ODI4%, SpO2, respiratory rate, PR, and commonly used pulse rate variability metrics in both time and frequency domains.30,61,62
Performance literature review: The performance of the Belun Ring has been evaluated in 4 studies: 2 utilizing generation-1 and 2 employing Gen2 algorithms (Table S4 in the supplemental material).52,62–64 While the generation-1 algorithms tended to overestimate AHI4% at the cutoff of 5 events/h, significant improvements were observed with Gen2 algorithms.52,62 Additionally, the Gen2 algorithms displayed satisfactory accuracy in identifying sleep stages.62 A recent independent study from Singapore by Ou et al explored the efficacy of the Gen2 algorithms in identifying OSA in patients with hypertension and at least 1 cardiovascular risk factor, irrespective of sleepiness. This study confirmed the device’s robust performance in detecting OSA at AHI4% cutoffs of 15 and 30 events/h.64
Comments: The Belun Ring stands out as the sole fully DL-powered OSA-detecting wearable, offering AHI4%, 3-stage sleep staging, and comprehensive pulse rate variability metrics.62,65 The Gen2 Belun Ring exhibits solid performance in identifying moderate-to-severe OSA using AHI4% across different races/ethnicities and in populations with high body mass index.62,64 Notably, the present version does not provide AHI3% and central event detection. Currently under validation investigation is the Belun generation-3 algorithms with a novel Belun Cor subxiphoid sensor, capable of capturing body position and respiratory effort, detecting central events, and offering both AHI3% and AHI4%.66
SleepImage SaMD
Introduction: The SleepImage SaMD (MyCardio, Denver, Colorado) is on its third iteration with the most recent 510(k) clearance for the diagnosis of sleep-disordered breathing in adults, adolescents, and children, utilizing PPG/pulse oximetry as an input signal.23
Sensors and mechanisms: The SleepImage SaMD utilizes either ECG or PPG signals for cardiopulmonary coupling (CPC) spectral analysis, calculating interactions between pulse rate variability and tidal volume variability to generate metrics for sleep state assessment and sleep-disordered breathing diagnosis.67,68 Either signal can be analyzed for synchronization using the oscillation amplitude at a given frequency (Cross-Spectral Power) and phase relationship (Coherence). CPC is defined as the [Cross-Spectral Power]2 × [Coherence].67,68 SleepImage AHI and respiratory disturbance index (RDI) are calculated using changes in PPG, CPC, and SpO2. Three-dimensional spectral analysis is used to discriminate between obstructive and central events.69 For sleep staging, SleepImage SaMD uses nontraditional scoring based on the CPC coupling frequency. Stable sleep occurs in high-frequency coupling ranges between 0.1 and 0.5 Hz and is associated with stage N3 sleep, with some of stage N2. Unstable sleep is defined by low-frequency coupling ranges 0.01–0.1 Hz and is representative of stage N1, with some of stage N2. Very low-frequency coupling occurs during wake and stage REM sleep.67,68 Currently, SleepImage SaMD couples with 2 types of sensors, with the original ring-type device positioned in the proximal phalanx and a new fingertip sensor. Both utilize PPG optical sensors with an accelerometer.
Algorithms: SleepImage SaMD operates using proprietary ruled-based algorithms.
Data output: The SleepImage platform offers channels including PPG, SpO2, PR, pulse rate variability, motion, and derived respiration, alongside metrics including AHI3%, AHI4%, RDI, central AHI, ODI, apnea duration, SpO2 data, PR data, SE, sleep onset latency, TST, WASO, sleep quality index, and time in various CPC sleep stages.23,68
Performance literature review: While many publications have studied sleep-disordered breathing using CPC, relatively few studies have rigorously evaluated the performance of SleepImage SaMD using PPG/pulse oximetry signals. Three recent studies assessed the SleepImage SaMD with ECG or PPG, as detailed in Table S5 in the supplemental material.70–72 Hilmisson et al performed a secondary analysis of full-night PSG files from the Childhood Adenenotonsillectomy Trial.70,73 The authors compared SleepImage-AHI3% derived from the PSG PPG to the manually scored PSG-AHI3% (hypopnea defined as > 50% drop in airflow with a 3% drop in SpO2 or arousal) and demonstrated a high level of agreement.70 Al Ashry et al compared SleepImage SaMD to PSG using SleepImage-AHI3% derived from PSG ECG and SpO2 signals from the Apnea Positive Pressure Long-Term Efficacy Study and showed a strong correlation between the SleepImage-AHI3% and the PSG-AHI3%.71,74 It is worth noting that (1) the PSG AHI3% criteria used in both studies differ from the current AASM 3% rule, and (2) the SleepImage AHI in these 2 studies were calculated using the ECG or PPG signals from the PSG (not from a separate wearable PPG sensor). Most recently, Lu et al conducted the first performance evaluation study using the SleepImage ring.72 Due to the limited sample size, drawing definitive conclusions regarding its accuracy is challenging, necessitating further validation.
Comments: SleepImage SaMD stands out as the sole CPC-based OSA-detecting wearables, cleared by the FDA for diagnosing OSA across a broad age range (age ≥ 2).23,68 A minor inconvenience is that SleepImage generates nonconventional sleep stages, and a comprehensive grasp of its unique CPC sleep staging is imperative for clinicians.68 Additionally, the referenced large-scale secondary studies derived SleepImage AHI by utilizing ECG or PPG/pulse oximetry data from the concurrent PSG.70,71 Hence, caution should be exercised in extrapolating these results to a separate SleepImage ring or fingertip sensor. Lastly, while CSA detection is feasible through ECG-based CPC 3-dimensional spectral analysis, its accuracy using a PPG-based device necessitates additional validation.69
Devices using non-PPG signals: acoustic-based wearables
Table 4 delineates the primary differences among the FDA-cleared non-PPG-based devices.
Table 4.
Comparison among FDA-cleared non-PPG-based OSA-detecting wearables.
| AcuPebble Ox100 | BresoDX1 | Sunrise2 | Wesper Lab | |
|---|---|---|---|---|
| FDA clearance | II, 510(k) | II, 510(k) | II (de novo pathway) | II, 510(k) |
| Intended use | OSA diagnosis | Moderate to severe OSA | OSA diagnosis | OSA diagnosis |
| Pediatric use | No | No | No (Europe, ≥ age 3) | No |
| Measuring site | Anterior Neck over trachea | Suprasternal notch over trachea | Chin | Chest, abdomen |
| Sensors | Microphone, paired fingertip pulse ox | Microphone, Acc, paired fingertip pulse ox | Acc, gyroscope, chin pulse ox, thermistor | RIP, Acc, recorder, fingertip pulse ox |
| AI/ML/DL algorithms | No, rule-based | Yes, DL (AHI) | ML (AHI, sleep staging) | Unclear |
| Snoring | Yes | No snoring level | No | Yes |
| Airflow | Acoustic flow | Acoustic flow | Thermistor flow | RIP flow |
| OSA detection | AHI3%, AHI4%* | AHI3%† | RDI‡, AHI3%*, AHI4%* | AHI3%†, AHI4%* |
| CSA detection | Yes (limited data) | No | Yes* | Yes* |
| TST and sleep stages | TST yes*; no sleep staging | No | Yes (light, deep, REM, wake) | Yes* |
| Body position | Only in Europe§ | Yes | Head position | Yes |
| Raw data accessibility | Channels only, not raw acoustic data | Lack of information | Yes | Yes |
| Manual editing | Yes | Lack of information | Yes | Yes |
| Multiple-night | Yes | Lack of information | Yes (up to 3 nights) | Yes |
| Disposable or reusable | Reusable | Lack of information | Disposable only | Reusable |
| Must pair with a smartphone | Yes | No, using a hub | Yes | Yes |
Lack of validation study. †AHI calculation based on TRT, not TST. ‡Validation studies using “optimal Sunrise-RDI cutoffs” and the cutoff points were assessed at the highest value of the Youden index (sensitivity plus specificity minus 1). §The upcoming Gen 2 AcuPebble sensor is reportedly set to include accelerometry and body position information. Acc = accelerometry, AHI = apnea-hypopnea index, AI = artificial intelligence, CSA = central sleep apnea, DL = deep learning, FDA = United States Food and Drug Administration, ML = machine learning, OSA = obstructive sleep apnea, ox = oximeter, PPG = photoplethysmography, REM = rapid eye movement, RIP = respiratory inductance plethysmography, TST = total sleep time.
AcuPebble
Introduction: The AcuPebble (Acurable, London, United Kingdom) offers multiple models for adult OSA assessment.75 Using the Drowzle smartphone app as a predicate, the AcuPebble SA100 first received FDA clearance in 2021 for OSA home screening.25 Subsequently, citing SA100 as a predicate, the upgraded Ox100, equipped with a fingertip pulse oximeter, was cleared by the FDA for adult OSA diagnosis.76
Sensors and mechanisms: The AcuPebble SA100 and Ox100 are equipped with proprietary piezoelectric microphones that gather sounds related to breathing, cardiac rhythms, and movements.25,76 Positioned at the base of the anterior neck, above the suprasternal notch, the sensor captures acoustic signals. Despite the lack of pulse oximetry in SA100, AcuPebble investigators asserted that “relative reductions in SpO2 can be obtained from the acoustic signal processing features derived from surrogate physiological signals, as opposed to directly from SpO2 levels.”77 Notably, while AcuPebble lacks an accelerometer, its algorithms reportedly can infer activity using acoustic signals.
Algorithms: The AcuPebble system operates using proprietary rule-based algorithms. While Ox100 offers PPG and SpO2 data, these data are currently not integrated into the Ox100 algorithms. Both Ox100 and SA100 utilize identical algorithms to derive AHI.
Data output: The SA100 platform allows access to various channels, encompassing acoustic-derived flow, respiratory inhalation and exhalation phases, snoring, PR (from sounds), cardiac sounds, and movements.25,77 While both SA100 and Ox100 provide AHI3%, AHI4%, ODI at 3%, and ODI4%, SA100 does not present SpO2 data as in Ox100. AcuPebble devices reportedly offer TST but have no sleep staging (per communication with Acurable).
Performance literature review: Three studies related to AcuPebble have been published thus far.77–79 Rodriguez-Villegas et al conducted a small preliminary pilot study exploring the accuracy of a prototype device in detecting apnea and hypopnea events using type III portable testing as a reference.78 Devani et al showcased impressive SA100 AHI3% and AHI4% accuracy for the diagnosis of moderate-to-severe OSA, which was referenced in the FDA 510(k) documents of both SA100 and Ox100.77 It is worth noting that in this study: (1) AcuPebble was compared to portable testing (not PSG), reflecting only agreement with the reference HSAT; (2) the AHI3% criteria did not adhere to the AASM 3% rule as the reference HSAT lacks EEG; (3) AcuPebble AHI was estimated from acoustic-derived airflow, acoustic-based TST, and “acoustic-derived relative O2 desaturation” without pulse oximetry; (4) although the accuracy in distinguishing between central and obstructive events was reportedly good, the data were collected from only 10 patients.77 More recently, Sanchez Gomez et al compared AcuPebble SA100 to concurrent in-lab PSG using the AASM 3% rule, demonstrating good AHI3% performance at PSG-AHI cutoff of 15 (Table S6 in the supplemental material).79 No data regarding AHI4% or AHI3% at cutoffs 5 and 30 events/h were presented. Currently, there are no Ox100 validating publications.
Comments: The AcuPebble is noteworthy as the first FDA-cleared acoustic-based OSA-detecting wearable. It offers AHI3% and AHI4%, along with signals visualization and annotation features. While the utilization of acoustic signals for assessing relative oxygen desaturation is innovative, the underlying mechanisms require further elucidation and independent validation. Furthermore, AcuPebble lacks published data on sleep-wake metrics and offers no sleep stages. The evaluation of CSA detection was conducted on small samples, raising questions about its generalizability.77,79 Lastly, the failure rate in older adults and individuals with obesity remains unclear, as loose neck skin may cause the sensor to swing during testing, leading to inaccuracies.77,79 Hence, careful patient selection is essential.
BresoDX1
Introduction: The BresoDX1 (BresoTEC, Toronto, Canada), an acoustic-based wearable, recently received clearance from the FDA as a Class II HSAT device for the diagnosis of moderate-to-severe OSA in adult patients.29,80
Sensors and mechanisms: The use of sound energy alone to estimate airflow presents challenges, as snoring can interfere with tracheal sound signals, and the airflow amplitude exhibits opposite polarity as opposed to the sound energy with a single positive polarity.81 Beyond capturing tracheal sound, the Breso DX1 integrated tracheal movements in the craniocaudal (x-axis) and anterior-posterior directions (z-axis), which are less affected by snoring and share similar polarity with airflow signals.81 The BresoDX1 is comprised of a BresoSensor (microphone and accelerometer), BresoHub (data hub), and a fingertip pulse oximeter. The BresoSensor is placed at the suprasternal notch to capture tracheal sounds and trachea-sternal motion data. These data are then fed into proprietary deep neural network (DNN) algorithms to derive airflow and respiratory effort.29,81,82
Algorithms: The latest BresoDX1 model includes 2 DNN algorithms with DNN1 for identifying spontaneous (short-term) changes in feature time-series and DNN2 for detecting long-term changes, designed to take into account circulation delay between SpO2 detected at the finger and tracheal signals.82 A comprehensive explanation of the 2 DNNs and their optimization process is detailed in the supplemental document accompanying the 2023 Nature and Science of Sleep article.82
Data output: Based on the 510(k), the BresoDX1 platform offers channels for airflow, respiratory effort, tracheal breath sounds, trachea-sternal movement, PR, SpO2, and body position.29
Performance literature review: Several earlier studies have explored the performance of the BresoDX series.83–85 More recently, 4 studies were published utilizing the BresoDX1 patch.81,82,86,87 Saha et al compared the AHI3% (with TRT used for AHI calculation) derived from the BresoDX1 algorithm with PSG-AHI3%, demonstrating good accuracy at AHI cutoffs of 15 and 30 events/h.86 Subsequently, Montazeri Ghahjaverestan et al assessed 2 algorithms aimed at estimating respiratory motion and airflow from tracheal signals, noting that trachea-sternal motion offers a more accurate estimate of airflow than tracheal sounds.81 Montazeri Ghahjaverestan et al also conducted a pilot study exploring the feasibility of estimating OSA severity by analyzing the estimated tidal volume and snoring sounds extracted from tracheal signals.87 This study showed that the tidal volume and snoring sounds extracted from tracheal signals could be used to estimate AHI. Most recently, Montazeri Ghahjaverestan et al integrated pulse oximetry into the BresoDX1 and compared BresoDX1-derived AHI3% (using TRT, not TST) to concurrent PSG-AHI3%.82 At PSG-AHI3% cutoffs of 15 and 30 events/h, BresoDX1 with pulse oximetry demonstrated satisfactory sensitivity and specificity. Notably, no AHI4% data were presented (Table S7 in the supplemental material).
Comments: The BresoDX1 offers a dual emphasis on both tracheal acoustic signals and trachea-sternal motion. However, it lacks TST, sleep stages, AHI3% using TST, and AHI4% and does not distinguish between central and obstructive events. Significantly, the BresoTEC website lacks information pertaining to the BresoDX1. Currently, it appears that the BresoDX1 is not commercially available.
Devices using non-PPG signals: effort-based wearables
Sunrise
Introduction: In 2022, the FDA granted Class II HSAT device clearance to Sunrise (Sunrise, Strombeek-Bever, Belgium) via the de novo pathway for adult OSA diagnosis.27,88 Using the generation-1 Sunrise wearable as a predicate, Sunrise subsequently submitted the upgraded model, known as Sunrise2, and obtained FDA clearance through the 510(k) pathway.89
Sensors and mechanisms: The generation-1 Sunrise featured an accelerometer and gyroscope, positioned on the patient’s chin in the mentolabial sulcus. This configuration enabled the capture of MM, offering insights into the respiratory drive and fluctuations in upper airway resistances during respiratory events.90 A study by Pepin et al examined the correlation between esophageal pressure and Sunrise MM signals, revealing a high level of synchronization with concurrent esophageal pressure signals, indicating that Sunrise MM served as a reliable noninvasive surrogate of esophageal pressure in patients with suspected OSA.90 Malhotra et al comprehensively reviewed and elucidated the physiology of MM during respiratory events in sleep.91 Notably, Sunrise2 integrated thermistor flow and PPG with SpO2 data into the chin sensor without adding a second sensor.
Algorithms: The Sunrise wearable utilizes ML algorithms for automated analysis of respiratory events and sleep stages.92,93 A multilabel classification algorithm has been developed for the detection of respiratory events, which processes MM raw signals and identifies 4 specifically defined target labels: wake, quiet sleep, respiratory effort, and arousals.92 Moreover, using the Extreme Gradient Boosting (XGBoost) classifier as the core algorithm, the Sunrise researchers developed and trained an ML sleep staging algorithm.93 The algorithm automatically classifies MM patterns into 1 of the 4 sleep stages: wake, light NREM (stages N1 + N2), deep NREM (stage N3), or REM.
Data output: The Sunrise2 biosignal channels include thermistor flow, SpO2, mandibular jaw movements by gyroscope, mandibular jaw movements by accelerometry, PR, position, and derived sleep stages.27 Various metrics are also provided, including AHI3%, AHI4%, OAHI, RDI, obstructive RDI (ORDI), respiratory effort-related arousal index, central AHI, supine AHI, nonsupine AHI, REM AHI, NREM AHI, ODI at 3%, ODI4%, arousal index, TST, sleep onset latency, REM latency, SE, WASO, and 4-stage with wake/light NREM/deep NREM/REM.
Performance literature review: The efficacy of MM in predicting respiratory events has been explored in several studies (Table S8 in the supplemental material).92–96 Martinot et al initially compared MM analysis, utilizing a different device, with simultaneous PSG in adults suspected of OSA. By employing “best MM-RDI cutoffs” of 5.9 events/h, 13.5 events/h, and 32.5 events/h to detect OSA at PSG cutoffs at 5 events/h, 15 events/h, and 30 events/h, this device achieved high sensitivity and specificity.94 Similarly, Pepin et al assessed the performance of Sunrise in a large sample, demonstrating comparable sensitivity and specificity using “optimal MM-RDI cutoffs” of 7.63 events/h and 12.65 events/h to detect OSA at PSG cutoffs at 5 events/h and 15 events/h. Notably, Sunrise consistently underestimated RDI when PSG-RDI exceeded 40 events/h, which may be attributed to its failure in detecting hypopnea events with isolated 3% desaturation without arousals.92 More recently, Kelly et al conducted a study and compared Sunrise ORDI to in-home PSG ORDI using the NOX A1 system, finding optimal Sunrise ORDI cutoffs of 9.53 events/h, 12.65 events/h, and 24.81 events/h for OSA detection at PSG-ORDI cutoffs at 5 events/h, 15 events/h, and 30 events/h.95 Additionally, Martinot et al examined the performance of Sunrise in the pediatric population, indicating that MM automated analysis holds promise for diagnosing moderate‐to‐severe pediatric OSA.96 Lastly, Le-Dong et al analyzed the Sunrise MM signals, trained a ML algorithm to provide sleep staging, and showed good agreement between MM ML analysis and PSG manual scoring in identifying sleep stages.93
Comments: Sunrise distinguishes itself with its unique capability to offer comprehensive respiratory physiological metrics using a single-point contact sensor positioned on the chin. It is crucial to note that Sunrise2, despite enhancements with thermistor flow and PPG, still relies on the original Sunrise algorithms, extracting data solely from accelerometry and gyroscope. While the additional signals captured in Sunrise2 are not incorporated for calculating respiratory metrics, they offer valuable channels for clinicians to manually overscore.
Also noteworthy is that Sunrise studies utilized Sunrise-RDI in identifying OSA, with no investigations conducted using Sunrise-AHI3% and Sunrise-AHI4%.90,92,95,96 Moreover, unlike standard device validation research, Sunrise studies present device performance using “optimal Sunrise-RDI cutoffs” rather than standard Sunrise-AHI or RDI cutoffs of 5, 15, and 30 events/h. This approach poses a challenge as the “optimal AHI/RDI cutoffs” could potentially vary across different patient populations, as demonstrated in adult participant studies where the Sunrise-RDI cutoff for detecting OSA at the PSG-RDI cutoff of 5 events/h varied.92,95 These inconsistencies raise 2 concerns. Firstly, what would be the true accuracy, sensitivity, and specificity if Sunrise-AHI or RDI cutoffs of 5, 15, and 30 events/h were employed? Secondly, which Sunrise-AHI/RDI cutoffs should clinicians adopt to assess OSA severity if standard cutoffs are not applied?
Furthermore, for men with beards, shaving hair at the mentolabial sulcus is likely required to securely position the sensor for testing. Also, it remains unclear whether trismus or other jaw movement abnormalities would impact its accuracy in diagnosing OSA.
Wesper lab
Introduction: The Wesper Lab (Wesper, New York, New York) boasts its cutting-edge design using 2 flexible patches affixed to the chest and abdomen for adult OSA diagnosis.97 A single-use Wesper Lab version initially obtained FDA clearance in 2021, followed by a multiple-use iteration, with a charging element added to the patches, which received FDA clearance in 2023.26,98 Although WesperO2, a finger pulse oximeter, secured FDA clearance in 2022, Wesper has opted to pair Wesper Lab patches with the CheckMe O2 Max (Shenzhen Viatom Technology Co., Ltd., Shenzhen, China) in its current offering.99
Sensors and mechanisms: The Wesper Lab patches measure chest and abdominal respiratory effort and RIP-derived airflow.100 Wesper Lab also detects body position and snoring using an accelerometer and recorder embedded within the patch. Its pulse oximeter provides PR and SpO2 data.
Algorithms: No AI/ML/DL algorithms were described in the 2 published articles and the 2 FDA 510(k) premarket notifications.26,98,100,101 Similar to AcuPebble and Sunrise, while Wesper Lab can offer PPG and SpO2, these metrics are not incorporated into its current algorithms.
Data output: Wesper Lab biosignal display includes RIP-derived airflow, pressure, SpO2, PR, body position, chest effort, abdominal effort, and derived sleep stages.26 Additional metrics encompass AHI3%, AHI4%, CAI, ODI at 3%, ODI4%, TST, sleep onset latency, SE, light NREM time, deep NREM time, and REM time. Wesper-TST is available either through Wesper’s own TST algorithm (no published validation thus far) or through the partnership with EnsoSleep (Ensodata, Madison, Winisconsin).102
Performance literature review: Only 2 studies on the Wesper Lab have been published (Table S9 in the supplemental material).100,101 In the study by Zandieh et al a small proof-of-concept study was conducted with 29 enrolled participants, of which only 19 provided usable data (none with severe OSA and only 3 without OSA). This pilot study demonstrated that Wesper Lab is capable of deriving airflow without nasal cannula pressure transducer or RIP belts. Importantly, its strong correlation between PSG-AHI and Wesper-AHI may be at least partially attributed to the utilization of “borrowed” PSG TST, SpO2, and body position data in the Wesper-AHI calculation.100 Raphelson et al subsequently compared Wesper Lab to PSG in 45 patients. Notably, TRT was utilized for AHI calculation, not TST. Central apneas were infrequent, and no definitive conclusion could be drawn on its utility in CSA detection.101
Comments: The Wesper Lab stands as the exclusive OSA-detecting wearable utilizing RIP. The published studies thus far have involved small cohorts with notable limitations including the absence of TST in the AHI3% calculation.101 While Wesper Lab offers both auto-scored AHI3% and AHI4%, there is a lack of published data on AHI4%. Further investigation is warranted to evaluate the accuracy of Wesper-TST and sleep staging. Additionally, its accuracy of CSA detection requires scrutiny and it is important to note both published studies reported a relatively high technical failure rate. Consequently, more research is needed to ascertain the efficacy of the Wesper Lab.
DISCUSSION
This “device-centered” review scrutinizes 9 innovative wearables designed to diagnose OSA. We examine their inner workings, exploring their operating principles, sensors, algorithms, data outputs, and pertinent performance evaluation studies. Detailed breakdowns of study demographics, exclusion criteria, and key statistical analyses are encapsulated in a supplemental document, providing a comprehensive picture of each device’s capabilities. While many of these devices hold immense promise for transforming OSA detection, further performance evaluation research is crucial to solidify their reliability and accuracy across diverse settings. Our review highlights the strengths and limitations of each wearable, underscoring the advancements they represent as well as the challenges that remain. By offering a detailed examination of these cutting-edge technologies, we hope to provide valuable insights for researchers and clinicians interested in the future of sleep apnea diagnosis and management. In this section, we will discuss a few cautions, how to implement these novel sleep technologies, and future trends.
Cautionary notes
First, it is imperative to acknowledge the uncertainty surrounding the reliability of these wearables in patients with significant cardiopulmonary or neuromuscular disorders, including atrial fibrillation, pacemakers, or congestive heart failure. These devices are best reserved for patients not falling under the exclusion criteria outlined in the validation studies. Caution must be exercised when extending the applicability to patients with excluded conditions or conditions affecting PPG (eg, skin tone) when PPG is used. It is also worth noting that most studies deployed wearables in the sleep lab and the performance at home under real-life scenarios merits further exploration.
Also vital is to comprehend that satisfactory performance demonstrated with AHI3% does not guarantee acceptable performance with AHI4%, and vice versa. Furthermore, FDA clearance for OSA diagnosis of a device does not necessarily indicate a thorough examination of performance across both AHI3% and AHI4%. Clinicians should always consult peer-reviewed literature to confirm the accuracy of essential metrics of a particular device before implementation.
Furthermore, clinicians should refrain from accepting research statistical analysis at face value. Clinicians must be vigilant in recognizing common pitfalls when interpreting performance assessment studies. Several research deficiencies, as outlined in Table 5, were observed during our review. Particular attention should be given to studies using “borrowed” PSG data for device-AHI calculation or relying on “optimal device-AHI cutoffs,” as these methodologies may inaccurately portray device performance.70,71,92,95,96,99 Additionally, when interpreting study results involving any AI/ML/DL algorithms, it is critical to ensure that the reported statistical analyses stem from an independent external validation dataset.32
Table 5.
Common sleep technology performance evaluation research pitfalls.
| Pitfalls | Description |
|---|---|
| Reference test | Comparing novel devices against type III HSAT, not gold standard PSG |
| Sample size | Small sample size |
| Race/ethnicity | Not including minorities or describing race/ethnicity |
| Eligibility criteria | Lack of clarity on exclusion criteria |
| Case selection | Not stating if cases are consecutive; lacks consort flow diagram |
| OSA severity distribution | Not reporting OSA severity distribution or biased OSA severity distribution with no or too few cases in specific OSA severity categories |
| AHI rules used in the study | Not using standard AASM 3% (1A) or 4% rules (1B) or not reporting the hypopnea criteria used |
| Using PSG data for wearable device AHI calculation | Calculating AHI by “borrowing” PSG data, such as utilizing PSG PPG data, PSG SpO2 data, PSG TST, or PSG position information |
| Using “optimal AHI cutoffs” | Not comparing performance with PSG using device-AHI cutoffs of 5, 15, and 30 but using device “optimal AHI cutoffs” |
| Statistics reporting | Only reporting correlation or limited/selected statistical analysis; not reporting standard measures such as accuracy, sensitivity, specificity, PPV, NPV, or essential data on agreement/concordance, Bland-Altman plots, Cohen’s kappa, and confusion matrix |
| AI dataset details | Inadequate description of the AI technology; lacking clarification regarding the training, internal validation, and external testing datasets |
AASM = American Academy of Sleep Medicine, AHI = apnea-hypopnea index, AI = artificial intelligence, DL = deep learning, EEG = electroencephalography, HSAT = home sleep apnea testing, ML = machine learning, NPV = negative predictive value, OSA = obstructive sleep apnea, PPG = photoplethysmography, PPV = positive predictive value, PSG = polysomnography, SpO2 = oxygen saturation, TST = total sleep time.
It is also crucial to emphasize that our understanding of performance evaluation studies must go beyond correlation coefficients, which primarily assess the strength of a linear relationship rather than equivalence between tests.103 Relying solely on the correlation coefficients overlooks variability in accuracy. Massie et al recently pointed out a notable discrepancy between reported diagnostic accuracy and correlation coefficient in WatchPAT validation studies.103 Lastly, clinicians should acquaint themselves with Band-Altman plots and confusion matrices, which are pivotal for interpreting sleep technology validation research.104,105
A framework with a structured article format for performance evaluation studies has recently been proposed, aiming to standardize the operationalization of key outcome metrics, thereby preventing misinterpretation of study outcomes and facilitating meaningful comparisons across sleep technologies.106
Implementation of novel wearables
Amid the wave of novel wearables, various pragmatic aspects require scrutiny before implementing any novel devices. Clinicians not only require confidence in the data outputs but also the capability to access and edit raw data. These novel wearables must seamlessly interface with existing middleware, scoring platforms, and practice workflows. Presented in the supplement are various key considerations beyond the medical aspects, accompanied by a sample worksheet developed for clinicians (Figure S1 in the supplemental material).
Financial considerations
The first consideration is to examine local payor policies to ensure the test aligns with reimbursement criteria. These typically adhere to AASM guidelines, with potential exceptions. Some policies impose restrictions on sensors and signals, as demonstrated by the Centers for Medicare & Medicaid Services policy update issued on March 16, 2023, which mandates that “all channels must be physical and not derived, calculated, virtualized, or created by software” and requires all data, including auto-scored data, be manually overscored.107 Comprehending such guidelines and evaluating signals from diverse devices can be challenging, often necessitating thorough investigation.
Moreover, financial consideration should encompass the costs associated with the device itself, any additional service fees, and labor costs incurred from deployment, staff training, tracking, prior data erasure, device recharging (which may extend turn-around time), scoring, editing, interpretation, and potential device loss. Furthermore, clinicians must assess whether the payor will cover therapy based on the test results. For positive airway pressure therapy, the test performed must meet established standards and be recognized by the payor for coverage. Given the low positive airway pressure adherence challenges, additional costs may further reduce patient willingness to undergo treatment.
Data considerations
An improvement in diagnostic technology may not translate well into clinical practice without proper data accessibility and management. When assessing the scoring environment, attention should be brought to the user experience. Ideally, editing should be intuitive, efficient, and include necessary user training and product support. Many companies have chosen to house the data in their own proprietary environment, which may be a limitation if exporting data is necessary or preferable. There also should be established processes to transfer data from the scoring technologist to the interpreting physician and potentially back to the scoring technologist if a rescore is requested. If multiple HSATs are utilized within the same department, the scoring workflows should ideally be similar to avoid confusion, prevent data loss, and maximize efficiency. Moreover, there must be clear protocols with explicitness about data security and transparency in data ownership.
Operational considerations
Operationalization entails scrutinizing the workflow to ensure seamless integration of new devices. If the practice currently utilizes middleware, an important consideration will include integration with existing middleware or the electronic health record to avoid duplicative work. It may be helpful to follow the device from deployment to collection. The patient will need to receive education on the device, either in person or over the phone. This can be supplemented by instructional videos. Some sleep facilities will provide technical support to patients in the middle of the night. If this is a service already provided in the existing workflow, then the clinical support team will need to be well-versed and familiar with the new device. Once the patient has completed the test, the device will either be disposed of or returned. Tracking the return will need to be operationalized, which can be straightforward for single-component devices but more complex for those with multiple components. If there are additional sensors, there should be an understanding of what happens if one sensor is lost, eg, whether the entire wearable system requires replacement or if the missing sensor can be individually replaced. Many novel wearables require the use of a smartphone or a tablet. A practice may decide to deploy these devices only to patients who already have the necessary technology. Care should be taken to avoid creating disparity. If a practice is located in an area where many patients do not possess these technologies, a consideration may be to offer more than 1 type of HSAT system to accommodate patient needs. As technology advances, it is likely that sensors and/or software will be updated or modified and an algorithm for communicating these changes should be present to ensure smooth transitions and maintain effective operations.
Strategic considerations
For any practice, investing in novel technologies represents significant financial and workflow commitments. Clinicians should meticulously evaluate the strength, credibility, and financial robustness of the companies behind these innovative technologies, along with their long-term sustainability in a competitive market. Additionally, clinicians should assess how these novel technologies align with their overarching diagnostic strategy for sleep disorders and whether they offer tangible benefits over existing approaches.
Future trends
Novel OSA-detecting wearables offer miniaturized, user-friendly, and often disposable sensors, coupled with advanced data processing, evolving AI/ML/DL algorithms, and multinight assessment capability. Looking ahead, in-lab PSG will remain the mainstay testing method when attended testing is required for the diagnosis of suspected OSA in patients with comorbidities, titration protocols, and evaluating complex sleep disorders. Ironically, similar to HSATs disrupting previous PSG-centered workflows, these novel wearables may emerge as primary diagnostic tools for patients at risk for moderate-to-severe OSA without comorbidities. For this transition to occur, several factors, including FDA clearance, Centers for Medicare & Medicaid Services/insurance testing and treatment requirements, as well as reimbursement policies, need to align with the rapidly evolving sleep technology.
Particularly noteworthy are the PPG-based wearables, as they not only can provide vital diagnostic metrics including pulse oximetry and sleep staging, but also hold the potential to explore cardiovascular outcome-predicting biomarkers like heart rate variability, hypoxic burden, delta heart rate, and pulse wave amplitude drop index.108–112 In contrast, earlier iterations of non-PPG-based wearables lack pulse oximetry, which is crucial for accurate AHI calculation. While their newer versions now display pulse oximetry data, most have not integrated the PPG/SpO2 data into their algorithms for AHI estimation. Nevertheless, the inclusion of pulse oximetry has set the stage for incorporating PPG/SpO2 data into future algorithm updates.
While this paper provides an overview of the novel OSA-detecting wearables, we did not explore advances in consumer-grade wearables, nearables, or smartphone applications (apps). These devices/algorithms/apps may ascend to medical-grade rank once they demonstrate satisfactory performance and secure FDA clearance. Some smartphone apps, such as Drowzle (Resonea, Scottsdale, Arizona), SleepCheck Rx (ResApp Health, Brisbane, Australia), and devices like AcuPebble SA100, have already received FDA clearance for “OSA prescreening,” an ambiguous term used primarily by the FDA in 510(k) documents indicating home screening.25,113,114 The appropriate utilization of these tools warrants further investigation as they may incur additional costs and may not qualify patients for therapeutic devices. At present, individuals experiencing OSA symptoms are best served by seeking medical attention and undergoing FDA-cleared diagnostic testing.
Lastly, in February 2024, the FDA classified the novel Samsung SmartWatch Sleep Apnea Feature over-the-counter mobile app (Samsung Electronics Company, Mountain View, California) into Class II through the de novo pathway for assessing the risk of moderate-to-severe adult OSA.115 While this clearance may catalyze a surge of similar offerings from other smartwatch manufacturers, its impact on OSA diagnosis is likely limited since this is labeled as an OSA “risk assessment” tool and not as a diagnostic tool. This milestone, nonetheless, signifies that the sleep medicine landscape is being swiftly reshaped by the sleep technology evolution.
Summary
While the optimal moment for the widespread adoption of OSA-detecting wearables may not have yet arrived, their role is undeniably set to rapidly expand in the foreseeable future. In light of this, the AASM recently launched a Focused Update Task Force on OSA diagnostic testing, and a new clinical practice guideline is expected to be unveiled in 2025.
In the interim, sleep clinicians should embrace these emerging tools, gain insights into the capabilities and constraints of these devices, and scrutinize the available data. Independent, multicenter performance evaluation studies, cross-device validation trials, and outcome research across diverse populations are warranted to instill clinician confidence that product claims are substantiated by robust clinical evidence. Also significantly, there is an explicit absence of device validation research in the pediatric population, highlighting the necessity to address this gap.
The call for continuous research to authenticate the efficacy of these devices is loud and clear, underscoring the notion that validation is not a one-off event but a continuous process. This path, albeit challenging, promises to yield more diagnostic alternatives, enhance and personalize care, heighten patient satisfaction, predict outcomes, advance population health, and hopefully, reduce economic expenditures associated with caring for patients with OSA in the future.
DISCLOSURE STATEMENT
All authors have reviewed and approved the manuscript. Dr. Ambrose Chiang has received research grants from Belun Technology Company Limited for conducting validation trials. Dr. Seema Khosla is a consultant for BrainTrain 2020, Jazz, and Eisai. The other authors report no conflicts of interest.
Supplemental Materials
ABBREVIATIONS
- AASM
American Academy of Sleep Medicine
- AHI
apnea-hypopnea index
- AHI3%
apnea-hypopnea index at 3% desaturation
- AHI4%
apnea-hypopnea index at 4% desaturation
- AI
artificial intelligence
- CPC
cardiopulmonary coupling
- DL
deep learning
- DNN
deep neural network
- ECG
electrocardiography
- EEG
electroencephalogram
- FDA
United States Food and Drug Administration
- Gen2
second-generation
- HSAT
home sleep apnea testing
- ML
machine learning
- MM
mandibular movement
- NREM
non-rapid eye movement
- ODI
oxygen desaturation index
- ODI4%
oxygen desaturation index at 4%
- ORDI
obstructive respiratory disturbance index
- OSA
obstructive sleep apnea
- PAT
peripheral arterial tonometry
- PPG
photoplethysmography
- PR
pulse rate
- PSG
polysomnography
- RDI
respiratory disturbance index
- REM
rapid eye movement
- RIP
respiratory inductance plethysmography
- SaMD
software as a medical device
- SE
sleep efficiency
- SpO2
oxygen saturation
- TRT
total recording time
- TST
total sleep time
- WASO
wake after sleep onset
REFERENCES
- 1. Grandner MA . Sleep, health, and society . Sleep Med Clin. 2022. ; 17 ( 2 ): 117 – 139 . [DOI] [PubMed] [Google Scholar]
- 2. Perez-Pozuelo I, Zhai B, Palotti J, et al . The future of sleep health: a data-driven revolution in sleep science and medicine . NPJ Digit Med. 2020. ; 3 : 42 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Chung J, Goodman M, Huang T, Bertisch S, Redline S . Multidimensional sleep health in a diverse, aging adult cohort: concepts, advances, and implications for research and intervention . Sleep Health. 2021. ; 7 ( 6 ): 699 – 707 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Yeghiazarians Y, Jneid H, Tietjens JR, et al . Obstructive sleep apnea and cardiovascular disease: a scientific statement from the American Heart Association . Published correction appears in Circulation. 2022;145(12):e775. Circulation. 2021. ; 144 ( 3 ): e56 – e67 . [DOI] [PubMed] [Google Scholar]
- 5. Gleeson M, McNicholas WT . Bidirectional relationships of comorbidity with obstructive sleep apnoea . Eur Respir Rev. 2022. ; 31 ( 164 ): 210256 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Frost & Sullivan . Hidden health crisis costing America billions. Underdiagnosing and undertreating obstructive sleep apnea draining healthcare system . American Academy of Sleep Medicine; . http://www.aasmnet.org/sleep-apnea-economic-impact.aspx . Published 2016. . Accessed March 8, 2024. [Google Scholar]
- 7. Watson NF . Health care savings: the economic value of diagnostic and therapeutic care for obstructive sleep apnea . J Clin Sleep Med. 2016. ; 12 ( 8 ): 1075 – 1077 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Kapur V, Strohl KP, Redline S, Iber C, O’Connor G, Nieto J . Underdiagnosis of sleep apnea syndrome in U.S. communities . Sleep Breath. 2002. ; 6 ( 2 ): 49 – 54 . [DOI] [PubMed] [Google Scholar]
- 9. Young T, Palta M, Dempsey J, Peppard PE, Nieto FJ, Hla KM . Burden of sleep apnea: rationale, design, and major findings of the Wisconsin Sleep Cohort study . WMJ. 2009. ; 108 ( 5 ): 246 – 249 . [PMC free article] [PubMed] [Google Scholar]
- 10. Peppard PE, Young T, Barnet JH, Palta M, Hagen EW, Hla KM . Increased prevalence of sleep-disordered breathing in adults . Am J Epidemiol. 2013. ; 177 ( 9 ): 1006 – 1014 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Benjafield AV, Ayas NT, Eastwood PR, et al . Estimation of the global prevalence and burden of obstructive sleep apnoea: a literature-based analysis . Lancet Respir Med. 2019. ; 7 ( 8 ): 687 – 698 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Lyons MM, Bhatt NY, Pack AI, Magalang UJ . Global burden of sleep-disordered breathing and its implications . Respirology. 2020. ; 25 ( 7 ): 690 – 702 . [DOI] [PubMed] [Google Scholar]
- 13. Randerath W, Bassetti CL, Bonsignore MR, et al . Challenges and perspectives in obstructive sleep apnoea: report by an ad hoc working group of the sleep disordered breathing group of the European Respiratory Society and the European Sleep Research Society . Eur Respir J. 2018. ; 52 ( 3 ): 1702616 . [DOI] [PubMed] [Google Scholar]
- 14. Green P . Sleep is the new status symbol . The New York Times. https://www.nytimes.com/2017/04/08/fashion/sleep-tips-and-tools.html . Published April 8, 2017. . Accessed March 8, 2024.
- 15. Penzel T, Glos M, Fietze I . New trends and new technologies in sleep medicine: expanding accessibility . Sleep Med Clin. 2021. ; 16 ( 3 ): 475 – 483 . [DOI] [PubMed] [Google Scholar]
- 16. Kwon S, Kim H, Yeo WH . Recent advances in wearable sensors and portable electronics for sleep monitoring . iScience. 2021. ; 24 ( 5 ): 102461 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Schutte-Rodin S, Deak MC, Khosla S, et al . Evaluating consumer and clinical sleep technologies: an American Academy of Sleep Medicine update . J Clin Sleep Med. 2021. ; 17 ( 11 ): 2275 – 2282 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Goldstein C . Current and future roles of consumer sleep technologies in sleep medicine . Sleep Med Clin. 2020. ; 15 ( 3 ): 391 – 408 . [DOI] [PubMed] [Google Scholar]
- 19. Chiang AA, Khosla S . Consumer wearable sleep trackers: are they ready for clinical use? Sleep Med Clin. 2023. ; 18 ( 3 ): 311 – 330 . [DOI] [PubMed] [Google Scholar]
- 20. Jerkin E, Chandrkantan A, Chiang A, et al . Novel devices and applications for diagnosing obstructive sleep apnea. Montage . American Academy of Sleep Medicine; . https://aasm.org/novel-devices-applications-diagnosing-obstructive-sleep-apnea . Published September 2023. . Accessed March 8, 2024. [Google Scholar]
- 21. Kapur VK, Auckley DH, Chowdhuri S, et al . Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an American Academy of Sleep Medicine clinical practice guideline . J Clin Sleep Med. 2017. ; 13 ( 3 ): 479 – 504 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Bianchi MT, Goparaju B . Potential underestimation of sleep apnea severity by at home kits: rescoring in-laboratory polysomnography without sleep staging . J Clin Sleep Med. 2017. ; 13 ( 4 ): 551 – 555 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. U.S. Food & Drug Administration . K182618. SleepImage System 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf18/K182618.pdf . Published August 14, 2019. . Accessed March 8, 2024.
- 24. U.S. Food & Drug Administration . K191031. NightOwl 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf19/K191031.pdf . Published March 6, 2020. . Accessed March 8, 2024.
- 25. U.S. Food & Drug Administration . K210480. AcuPebble SA100 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf21/K210480.pdfz . Published July 6, 2021. . Accessed March 8, 2024.
- 26. U.S. Food & Drug Administration . K203343. Wesper Lab 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf20/K203343.pdf . Published December 21, 2021. . Accessed March 8, 2024.
- 27. U.S. Food & Drug Administration . DEN210015. Sunrise Sleep Disorder Diagnostic Aid De Novo classification notification . https://www.accessdata.fda.gov/cdrh_docs/pdf21/DEN210015.pdf . Published January 7, 2022. . Accessed March 8, 2024.
- 28. U.S. Food & Drug Administration . K220095. ANNE Sleep 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf22/K220095.pdf . Published July 11, 2022. . Accessed March 8, 2024.
- 29. U.S. Food & Drug Administration . K220012. BresoDX1 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf22/K220012.pdf . Published December 21, 2022. . Accessed March 8, 2024.
- 30. U.S. Food & Drug Administration . K222579. Belun Sleep System BLS-100 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf22/K222579.pdf . Published February 23, 2023. . Accessed March 8, 2024.
- 31. U.S. Food & Drug Administration . K231546. Somfit 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf23/K231546.pdf . Published November 30, 2023. . Accessed March 8, 2024.
- 32. Bandyopadhyay A, Bae C, Cheng H, et al . Smart sleep: what to consider when adopting AI-enabled solutions in clinical practice of sleep medicine . J Clin Sleep Med. 2023. ; 19 ( 10 ): 1823 – 1833 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Collop NA, Tracy SL, Kapur V, et al . Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation . J Clin Sleep Med. 2011. ; 7 ( 5 ): 531 – 548 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. U.S. Food & Drug Administration . 510(k) Premarket Notification Search Database . https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm . Updated March 4, 2024. . Accessed March 8, 2024.
- 35. U.S. Food & Drug Administration . Device Classification under Section 513(f)(2)(De Novo) search database . https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm . Updated March 4, 2024. . Accessed March 8, 2024.
- 36. The National Institutes of Health . National Library of Medicine. ClinicalTrials.gov. https://clinicaltrials.gov . Accessed March 8, 2024.
- 37. U.S. Food & Drug Administration . 510(k) clearances . https://www.fda.gov/medical-devices/device-approvals-and-clearances/510k-clearances . Updated November 6, 2023. . Accessed March 8, 2024.
- 38. U.S. Food & Drug Administration . De novo classification request . https://www.fda.gov/medical-devices/premarket-submissions-selecting-and-preparing-correct-submission/de-novo-classification-request . Updated October 4, 2022. . Accessed March 8, 2024.
- 39. U.S. Food & Drug Administration . Software as a medical device (SaMD) . https://www.fda.gov/medical-devices/digital-health-center-excellence/software-medical-device-samd . Updated December 4, 2018. . Accessed March 8, 2024.
- 40. Ryals S, Chiang A, Schutte-Rodin S, et al . Photoplethysmography-new applications for an old technology: a sleep technology review . J Clin Sleep Med. 2023. ; 19 ( 1 ): 189 – 195 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Kyriacou P . Introduction to photoplethysmography . In: Kyriacou P, Allen J , eds. Photoplethysmography: Technology, Signal Analysis, and Applications. Amsterdam, The Netherlands: : Elsevier; ; 2022. : 1 – 15 . [Google Scholar]
- 42. Tamura T, Maeda Y, Sekine M, Yoshida M . Wearable photoplethysmographic sensors – past and present . Electronics. 2014. ; 3 ( 2 ): 282 – 302 . [Google Scholar]
- 43. Ghamari M, Castaneda D, Esparza A, Soltanpur C, Nazeran H . A review on wearable photoplethysmography sensors and their potential future applications in health care . Int J Biosens Bioelectron. 2018. ; 4 ( 4 ): 195 – 202 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Kyriacou P, Chatterjee S . The origin of photoplethysmography . In: Kyriacou P, Allen J , eds. Photoplethysmography: Technology, Signal Analysis, and Applications. Amsterdam, The Netherlands: : Elsevier; ; 2022. : 17 – 42 . [Google Scholar]
- 45. Budhida K, Kyriacou P . Photoplethysmography technology . In: Kyriacou P, Allen J , eds. Photoplethysmography: Technology, Signal Analysis, and Applications. Amsterdam, The Netherlands: : Elsevier; ; 2022. : 43 – 67 . [Google Scholar]
- 46. Schnall RP, Shlitner A, Sheffy J, Kedar R, Lavie P . Periodic, profound peripheral vasoconstriction—a new marker of obstructive sleep apnea . Sleep. 1999. ; 22 ( 7 ): 939 – 946 . [PubMed] [Google Scholar]
- 47. Schnall RP, Sheffy JK, Penzel T . Peripheral arterial tonometry-PAT technology . Sleep Med Rev. 2022. ; 61 : 101566 . [DOI] [PubMed] [Google Scholar]
- 48. Van Pee B, Massie F, Vits S, et al . A multicentric validation study of a novel home sleep apnea test based on peripheral arterial tonometry . Sleep. 2022. ; 45 ( 5 ): zsac028 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49. Collop NA, Anderson WM, Boehlecke B, et al. ; Portable Monitoring Task Force of the American Academy of Sleep Medicine . Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable monitoring task force of the American Academy of Sleep Medicine . J Clin Sleep Med. 2007. ; 3 ( 7 ): 737 – 747 . [PMC free article] [PubMed] [Google Scholar]
- 50. Ayas NT, Pittman S, MacDonald M, White DP . Assessment of a wrist-worn device in the detection of obstructive sleep apnea . Sleep Med. 2003. ; 4 ( 5 ): 435 – 442 . [DOI] [PubMed] [Google Scholar]
- 51. Massie F, Mendes de Almeida D, Dreesen P, Thijs I, Vranken J, Klerkx S . An evaluation of the NightOwl home sleep apnea testing system . J Clin Sleep Med. 2018. ; 14 ( 10 ): 1791 – 1796 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Yeh E, Wong E, Tsai CW, et al . Detection of obstructive sleep apnea using Belun Sleep Platform wearable with neural network-based algorithm and its combined use with STOP-Bang questionnaire . PLoS One. 2021. ; 16 ( 10 ): e0258040 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. ResMed . Get in touch . https://nightowl.resmed.com/signin . Accessed March 8, 2024.
- 54. Lyne CJ, Hamilton GS, Turton ARE, et al . Validation of a single-use and reusable home sleep apnea test based on peripheral arterial tonometry compared to laboratory polysomnography for the diagnosis of obstructive sleep apnea . J Clin Sleep Med. 2023. ; 19 ( 8 ): 1429 – 1435 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Massie F, Van Pee B, Vits S, Verbraecken J, Bergmann J . Phenotyping REM OSA by means of peripheral arterial tone-based home sleep apnea testing and polysomnography: a critical assessment of the sensitivity and specificity of both methods . J Sleep Res. 2022. ; 31 ( 2 ): e13481 . [DOI] [PubMed] [Google Scholar]
- 56. Massie F, Vits S, Khachatryan A, Van Pee B, Verbraecken J, Bergmann J . Central sleep apnea detection by means of finger photoplethysmography . IEEE J Transl Eng Health Med. 2023. ; 11 : 126 – 136 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Sibel . Sleep Diagnostic/ANNE® Sleep. Clinical grade sleep monitoring anywhere . https://sibelhealth.com/sleep-diagnostic . Accessed March 8, 2024.
- 58. Davies C, Lee JY, Walter J, et al . A single-arm, open-label, multicenter, and comparative study of the ANNE sleep system vs polysomnography to diagnose obstructive sleep apnea . J Clin Sleep Med. 2022. ; 18 ( 12 ): 2703 – 2712 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. Argod J, Pépin JL, Lévy P . Differentiating obstructive and central sleep respiratory events through pulse transit time . Am J Respir Crit Care Med. 1998. ; 158 ( 6 ): 1778 – 1783 . [DOI] [PubMed] [Google Scholar]
- 60. Miller DJ, Sargent C, Roach GD . A validation of six wearable devices for estimating sleep, heart rate and heart rate variability in healthy adults . Sensors (Basel). 2022. ; 22 ( 16 ): 6317 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61. Belun . Belun Sleep System . https://beluntech.com/us . Accessed March 8, 2024.
- 62. Strumpf Z, Gu W, Tsai CW, et al . Belun Ring (Belun Sleep System BLS-100): deep learning-facilitated wearable enables obstructive sleep apnea detection, apnea severity categorization, and sleep stage classification in patients suspected of obstructive sleep apnea . Sleep Health. 2023. ; 9 ( 4 ): 430 – 440 . [DOI] [PubMed] [Google Scholar]
- 63. Gu W, Leung L, Kwok KC, Wu IC, Folz RJ, Chiang AA . Belun ring platform: a novel home sleep apnea testing system for assessment of obstructive sleep apnea . J Clin Sleep Med. 2020. ; 16 ( 9 ): 1611 – 1617 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64. Ou YH, Ong J, Thant AT, et al . The Belun sleep platform to diagnose obstructive sleep apnea in patients with hypertension and high cardiovascular risk . J Hypertens. 2023. ; 41 ( 6 ): 1011 – 1017 . [DOI] [PubMed] [Google Scholar]
- 65. Tsai C, Gu W, Yeh E, et al . Correlation of pulse rate variability (PRV) and heart rate variability (HRV) metrics during sleep in subjects suspected of OSA . Sleep. 2023. ; 46 ( Supplement_1 ): A420 – A421 . [Google Scholar]
- 66. The National Institutes of Health . National Library of Medicine. ClinicalTrials.gov. Validation of Belun Ring Gen3 deep learning algorithms with subxiphoid body sensor. https://clinicaltrials.gov/study/NCT06255613?intr=belun&rank=1 . Updated March 6, 2024. . Accessed March 8, 2024.
- 67. SleepImage . Introduction to SleepImage . https://sleepimage.com/wp-content/uploads/Introduction-to-SleepImage.pdf . Accessed March 8, 2024.
- 68. Thomas RJ, Mietus JE, Peng CK, Goldberger AL . An electrocardiogram-based technique to assess cardiopulmonary coupling during sleep . Sleep. 2005. ; 28 ( 9 ): 1151 – 1161 . [DOI] [PubMed] [Google Scholar]
- 69. Thomas RJ, Mietus JE, Peng CK, et al . Differentiating obstructive from central and complex sleep apnea using an automated electrocardiogram-based method . Sleep. 2007. ; 30 ( 12 ): 1756 – 1769 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. Hilmisson H, Berman S, Magnusdottir S . Sleep apnea diagnosis in children using software-generated apnea-hypopnea index (AHI) derived from data recorded with a single photoplethysmogram sensor (PPG): results from the Childhood Adenotonsillectomy Study (CHAT) based on cardiopulmonary coupling analysis . Sleep Breath. 2020. ; 24 ( 4 ): 1739 – 1749 . [DOI] [PubMed] [Google Scholar]
- 71. Al Ashry HS, Hilmisson H, Ni Y, Thomas RJ , APPLES Investigators . Automated apnea-hypopnea index from oximetry and spectral analysis of cardiopulmonary coupling . Ann Am Thorac Soc. 2021. ; 18 ( 5 ): 876 – 883 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72. Lu M, Brenzinger L, Rosenblum L, et al . Comparative study of the SleepImage ring device and polysomnography for diagnosing obstructive sleep apnea . Biomed Eng Lett. 2023. ; 13 ( 3 ): 343 – 352 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Marcus CL, Moore RH, Rosen CL, et al. ; Childhood Adenotonsillectomy Trial (CHAT) . A randomized trial of adenotonsillectomy for childhood sleep apnea . N Engl J Med. 2013. ; 368 ( 25 ): 2366 – 2376 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74. Kushida CA, Nichols DA, Holmes TH, et al . Effects of continuous positive airway pressure on neurocognitive function in obstructive sleep apnea patients: the Apnea Positive Pressure Long-term Efficacy Study (APPLES) . Published correction appears in Sleep. 2016;39(7):1483. Sleep. 2012. ; 35 ( 12 ): 1593 – 1602 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75. Acurable . AcuPebble SA100 . https://acurable.com/en/products/acupebble-SA100 . Accessed March 8, 2024.
- 76. U.S. Food & Drug Administration . K222950. AcuPebble Ox100 510(k) premarket notification . https://www.accessdata.fda.gov/scripts/cdrh/devicesatfda/index.cfm?db=pmn&id=K222950 . Published February 3, 2023. . Accessed March 8, 2024.
- 77. Devani N, Pramono RXA, Imtiaz SA, Bowyer S, Rodriguez-Villegas E, Mandal S . Accuracy and usability of AcuPebble SA100 for automated diagnosis of obstructive sleep apnoea in the home environment setting: an evaluation study . BMJ Open. 2021. ; 11 ( 12 ): e046803 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78. Rodriguez-Villegas E, Chen G, Radcliffe J, Duncan J . A pilot study of a wearable apnoea detection device . BMJ Open. 2014. ; 4 ( 10 ): e005299 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. Sanchez Gomez J, Pramono RXA, Imtiaz SA, Rodriguez-Villegas E, Valido Morales A . Validation of a wearable medical device for automatic diagnosis of OSA against standard PSG . J Clin Med. 2024. ; 13 ( 2 ): 571 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80. BresoTEC . Engineering sleep, advancing health . https://bresotec.com . Accessed March 8, 2024.
- 81. Montazeri Ghahjaverestan N, Fan W, Aguiar C, Yu J, Bradley TD . Respiratory motion and airflow estimation during sleep using tracheal movement and sound . Nat Sci Sleep. 2022. ; 14 : 1213 – 1223 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82. Montazeri Ghahjaverestan N, Aguiar C, Hummel R, Cao X, Yu J, Bradley TD . Sleep apnea detection by tracheal motion and sound, and oximetry via application of deep neural networks . Nat Sci Sleep. 2023. ; 15 : 423 – 432 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83. Alshaer H, Fernie GR, Maki E, Bradley TD . Validation of an automated algorithm for detecting apneas and hypopneas by acoustic analysis of breath sounds . Sleep Med. 2013. ; 14 ( 6 ): 562 – 571 . [DOI] [PubMed] [Google Scholar]
- 84. Alshaer H, Fernie GR, Tseng WH, Bradley TD . Comparison of in-laboratory and home diagnosis of sleep apnea using a cordless portable acoustic device . Sleep Med. 2016. ; 22 : 91 – 96 . [DOI] [PubMed] [Google Scholar]
- 85. Fitzpatrick M, Rac VE, Mitsakakis N, et al . SIESTA - Home sleep study with BresoDx for obstructive sleep apnea: a randomized controlled trial . Sleep Med. 2020. ; 65 : 45 – 53 . [DOI] [PubMed] [Google Scholar]
- 86. Saha S, Kabir M, Montazeri Ghahjaverestan N, et al . Portable diagnosis of sleep apnea with the validation of individual event detection . Sleep Med. 2020. ; 69 : 51 – 57 . [DOI] [PubMed] [Google Scholar]
- 87. Montazeri Ghahjaverestan N, Saha S, Kabir M, Gavrilovic B, Zhu K, Yadollahi A . Sleep apnea severity based on estimated tidal volume and snoring features from tracheal signals . J Sleep Res. 2022. ; 31 ( 2 ): e13490 . [DOI] [PubMed] [Google Scholar]
- 88. Sunrise . Polygraphy and more in a 8g sensor; SleepScience, at the tip of your chin . https://us.hellosunrise.com . Accessed March 8, 2024.
- 89. U.S. Food & Drug Administration . K222262. Sunrise 510(k) premarket notification . https://www.accessdata.fda.gov/scripts/cdrh/devicesatfda/index.cfm?db=pmn&id=K222262 . Published December 22, 2022. . Accessed March 8, 2024.
- 90. Pepin JL, Le-Dong NN, Cuthbert V, et al . Mandibular movements are a reliable noninvasive alternative to esophageal pressure for measuring respiratory effort in patients with sleep apnea syndrome . Nat Sci Sleep. 2022. ; 14 : 635 – 644 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91. Malhotra A, Martinot JB, Pépin JL . Insights on mandibular jaw movements during polysomnography in obstructive sleep apnea . J Clin Sleep Med. 2024. ; 20 ( 1 ): 151 – 163 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92. Pépin JL, Letesson C, Le-Dong NN, et al . Assessment of mandibular movement monitoring with machine learning analysis for the diagnosis of obstructive sleep apnea . JAMA Netw Open. 2020. ; 3 ( 1 ): e1919657 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93. Le-Dong NN, Martinot JB, Coumans N, et al . Machine learning-based sleep staging in patients with sleep apnea using a single mandibular movement signal . Am J Respir Crit Care Med. 2021. ; 204 ( 10 ): 1227 – 1231 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94. Martinot JB, Borel JC, Cuthbert V, et al . Mandibular position and movements: suitability for diagnosis of sleep apnoea . Published correction appears in Respirology. 2018;23(1):117. Respirology. 2017. ; 22 ( 3 ): 567 – 574 . [DOI] [PubMed] [Google Scholar]
- 95. Kelly JL, Ben Messaoud R, Joyeux-Faure M, et al . Diagnosis of sleep apnoea using a mandibular monitor and machine learning analysis: one-night agreement compared to in-home polysomnography . Front Neurosci. 2022. ; 16 : 726880 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96. Martinot JB, Cuthbert V, Le-Dong NN, et al . Clinical validation of a mandibular movement signal based system for the diagnosis of pediatric sleep apnea . Pediatr Pulmonol. 2022. ; 57 ( 8 ): 1904 – 1913 . [DOI] [PubMed] [Google Scholar]
- 97. Wesper . The future of sleep care . https://wesper.co . Accessed March 8, 2024.
- 98. U.S. Food & Drug Administration . K221816. Wesper Lab 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf22/K221816.pdf . Published March 9, 2023. . Accessed March 8, 2024.
- 99. U.S. Food & Drug Administration . K213515. WesperO2 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf21/K213515.pdf . Published June 10, 2022. . Accessed March 8, 2024.
- 100. Zandieh S, Kirschenbaum MA, Greenberg H, Ancoli-Israel S . Keep it simple: a novel technique for measuring airflow using a wireless patch . Sleep Health. 2023. ; 9 ( 1 ): 100 – 107 . [DOI] [PubMed] [Google Scholar]
- 101. Raphelson JR, Ahmed IM, Ancoli-Israel S, et al . Evaluation of a novel device to assess obstructive sleep apnea and body position . J Clin Sleep Med. 2023. ; 19 ( 9 ): 1643 – 1649 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102. Ensodata . Optimize the sleep care continuum . https://www.ensodata.com . Accessed March 8, 2024.
- 103. Massie F, Van Pee B, Bergmann J . Correlations between home sleep apnea tests and polysomnography outcomes do not fully reflect the diagnostic accuracy of these tests . J Clin Sleep Med. 2022. ; 18 ( 3 ): 871 – 876 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104. Depner CM, Cheng PC, Devine JK, et al . Wearable technologies for developing sleep and circadian biomarkers: a summary of workshop discussions . Sleep. 2020. ; 43 ( 2 ): zsz254 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105. de Zambotti M, Goldstein C, Cook J, et al . State of the science and recommendations for using wearable technology in sleep and circadian research . Sleep. 2023. ; 47 ( 4 ): zsad325 . [DOI] [PubMed] [Google Scholar]
- 106. de Zambotti M, Menghini L, Grandner MA, et al . Rigorous performance evaluation (previously, “validation”) for informed use of new technologies for sleep health measurement . Sleep Health. 2022. ; 8 ( 3 ): 263 – 269 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107. Centers for Medicare & Medicaid Services . CMS.gov. Billing and coding: polysomnography A56995 . https://www.cms.gov/medicare-coverage-database/view/article.aspx?articleid=56995 . Published March 16, 2023. . Accessed March 8, 2024.
- 108. Azarbarzin A, Sands SA, Stone KL, et al . The hypoxic burden of sleep apnoea predicts cardiovascular disease-related mortality: the osteoporotic fractures in men study and the sleep heart health study . Published correction appears in Eur Heart J. 2019;40(14):1157. Eur Heart J. 2019. ; 40 ( 14 ): 1149 – 1157 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109. Azarbarzin A, Sands SA, Taranto-Montemurro L, et al . The sleep apnea-specific hypoxic burden predicts incident heart failure . Chest. 2020. ; 158 ( 2 ): 739 – 750 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110. Azarbarzin A, Sands SA, Younes M, et al . The sleep apnea-specific pulse-rate response predicts cardiovascular morbidity and mortality . Am J Respir Crit Care Med. 2021. ; 203 ( 12 ): 1546 – 1555 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111. Solelhac G, Sánchez-de-la-Torre M, Blanchard M, et al . Pulse wave amplitude drops index: a biomarker of cardiovascular risk in obstructive sleep apnea . Am J Respir Crit Care Med. 2023. ; 207 ( 12 ): 1620 – 1632 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112. Phillips CL, Marshall NS . Searching for a vulnerable cardiovascular endotype in obstructive sleep apnea: is the humble pulse wave a useful biomarker? Am J Respir Crit Care Med. 2023. ; 207 ( 12 ): 1553 – 1555 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113. U.S. Food & Drug Administration . K173974. Drowzle sleep apnea prescreening device 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf17/K173974.pdf . Published July 14, 2019. . Accessed January 19, 2024.
- 114. U.S. Food & Drug Administration . K213360. SleepCheckRx 510(k) premarket notification . https://www.accessdata.fda.gov/cdrh_docs/pdf21/K213360.pdf . Published July 5, 2022. . Accessed January 19, 2024.
- 115. U.S. Food & Drug Administration . DEN230041. Sleep apnea feature de novo classification notification . https://www.accessdata.fda.gov/cdrh_docs/pdf21/DEN210015.pdf . Published February 6, 2024. . Accessed March 8, 2024.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.