Abstract
Hinge constructs can be used in revision total knee arthroplasty in cases of major collateral ligament deficiency, global knee instability, repeated failure of constrained implants, or large bone defects. Although the success of this construct in revision total knee arthroplasty varies depending on the type of surgery and bone quality, hinged implants can solve a wide array of complex problems while preserving stable flexion-extension motion for ambulation. The most common reasons for re-revision include patellofemoral complications, aseptic loosening, and infection. Most cases of mechanical failure for these hinged constructs occur at the hinge component. Here, we report the first instance of catastrophic femoral component failure in a modular rotating hinge via a low-energy mechanism.
Keywords: Revision total knee arthroplasty, Modular rotating hinge, Femoral component, Mechanical failure, Stress radiographs
Introduction
Although modern implant designs for primary total knee arthroplasty (TKA) have excellent success regarding outcomes and survivorship, the demand for revision TKA (rTKA) is expected to continue increasing [1]. Each rTKA is unique depending on the indication for surgery, the implant being revised, patient bone quality, and ligamentous stability [2]. Common indications for utilization of a rotating or fixed hinge implant in rTKA are repeated failure of constrained implants, collateral ligament and/or global knee instability, extensor mechanism insufficiency, and significant bone loss [[3], [4], [5], [6], [7], [8], [9]]. The benefit of hinge implants includes coronal stability and preservation of flexion-extension motion for walking [7,8,10]. Initial iterations of the hinge prosthesis exhibited a high incidence of early loosening and required long cemented stems to accommodate the significant stress exerted on the bone-implant interface [8,10]. However, contemporary designs have incorporated a rotating platform, which confers more physiologic knee motion and decreases stress transmission through the bone-implant interface [8,10].
Although these implants are an improvement from previous designs, survivorship and complication rates vary widely depending on surgery type and patient-related factors [[3], [4], [5], [6], [7], [8], [9], [10]]. Overall, the most common complications include infection and aseptic loosening [8,10]. There have been several reported complications involving disengagement of the hinge mechanism; however, there have been no reported mechanical failures of the femoral component to our knowledge [[11], [12], [13], [14], [15]]. Here, we report the case of a patient with catastrophic femoral component failure in a modular rotating hinge via a low-energy mechanism.
Case history
A 73-year-old Jehovah’s Witness female patient with a history of hypertension, gastroesophageal reflux disease, and a previous cerebral vascular accident (without residual effects) presented with pain and instability in her right knee. These symptoms appeared 11 years after her initial TKA. Her primary concerns were significant pain and instability. At this visit, the patient was found to have a range of motion of 0-90 degrees with gross instability and no endpoint on varus stress. Anteroposterior (AP) and lateral radiographs were suggestive of lateral tibial subluxation but no obvious evidence of component loosening, polyethylene insert failure, malpositioning, or fracture (Fig. 1). Intraoperatively, it was noted that her lateral condyle lacked soft tissue attachments and her medial collateral ligament also showed significant laxity; thus, she was revised to a modular rotating hinge (MRH) (Stryker, Kalamazoo, MI). At the time of revision to a MRH, minimal bone loss was noted with removal of the femoral and tibial components. However, poor bone quality resulted in needing a tibial cone for additional stability and adequate fixation. Femoral bone loss was less than 5 mm, thus cement was sufficient without the need for augments. Immediate postoperative films demonstrated well-fixed components with radiolucency noted about the femoral component along the anterior flange and posterior condyle (Fig. 2).
Figure 1.
(a) Anteroposterior (AP) and (b) lateral radiograph of right knee at presentation with instability.
Figure 2.
(a) Anteroposterior (AP) and (b) lateral radiograph of right knee after revision to modular rotating hinge.
Over the course of her first year of follow-up, she made improvements in pain and function returning to her baseline. However, at her 1-year postoperative visit, she presented with new-onset pain and instability after a caregiver fell onto her flexed knee. After this incident, she continued to have pain that localized anteriorly and caused her increased difficulty with walking. AP and lateral radiograph of the knee did not demonstrate any obvious complications; femoral and tibial components appeared well-fixed with no progression of radiolucency (Fig. 3). Consequently, she was instructed to use a cane as needed and to wrap her leg with an ace wrap for gentle compression. She initially improved after this visit until she fell, resulting in significant worsening of her condition. She returned to clinic in a wheelchair, as she now had trouble ambulating. On examination, her straight leg raise was intact with a range of motion of 0-115 degrees. Updated radiographs were unchanged. However, given her worsening pain, infectious workup and computed tomography scan were done. Erythrocyte sedimentation rate (12 mm/hr) was within normal limits, C-reactive protein (1.56 mg/dL) was elevated, and aspiration was benign (cell count: 859 cells/μL; neutrophils: 48%; culture: no growth).
Figure 3.
(a) Anteroposterior (AP) and (b) lateral radiographic view of right knee after onset of pain and instability 1-year postoperatively.
At her subsequent visit, her knee was found to have gross multidirectional instability. She did not report any additional specific events leading to this new instability. Repeat AP and lateral radiographs as well as stress radiographs showed catastrophic failure of the MRH femoral component (Fig. 4) with motion through the lateral femoral condyle. Considering that the tibial component appeared to be well fixed, we planned for expeditious revision of her femoral component only.
Figure 4.
(a) Anteroposterior (AP) and (b) lateral radiograph of right knee as well as (c) valgus and (d) varus stress view showing catastrophic femoral component failure.
Intraoperatively, upon entering the knee joint, the femoral component was found to be completely fragmented across the distal lateral condyle (Fig. 5). There was considerable wear on the extension block poly; however, there was no evidence of damage at the hinge mechanism. The femoral component was easily removed with only mild bone loss on the lateral distal femur likely associated with the component failure (Fig. 6). She was reconstructed with a new MRH femoral component. Postoperatively, the patient was doing well, and her AP and lateral radiographs showed appropriately aligned and well-fixed components with no sign of early complications (Fig. 7). At most recent follow-up, 6 months after revision of her failed MRH femoral component, the patient was pleased with her surgery and was independently ambulating.
Figure 5.
(a) Intraoperative in situ view of the fractured femoral component lateral condyle. (b) Intraoperative in situ view of the fractured femoral component notch and hinge mechanism.Intraoperative view of femoral component failure prior to explant.
Figure 6.
(a) Intraoperative view of femoral component failure after explanation showing the intact hinge mechanism. (b) An oblique view of the fracture line across the lateral condyle of the femoral component. (c) A lateral view of the fracture line across the lateral condyle of the femoral component.
Figure 7.
(a) Anteroposterior (AP) and (b) lateral radiograph of right knee after femur only revision.
Discussion
Generally, hinge constructs are a valuable option for patients with significant instability and bone loss, especially on repeat revisions [8,10]. While MRH constructs have been an improvement on pure hinge constructs regarding aseptic loosening and periprosthetic fracture, aseptic loosening remains one of the most common mechanisms of failure [3,4,6,7,10]. Additionally, other mechanisms of failure have been reported in the literature such as failure of the bushing [11], fracture of tibial post [12], fracture of the tibial yoke [13], disengagement of the antidislocation mechanism [16], and tibiofemoral disengagement of the hinge-post extension [12,14]. However, there are no reports in the literature of lateral condyle fracture of an MRH construct.
To our knowledge, this was the first report of catastrophic lateral femoral condyle failure in an MRH. In this case, there were several instances of trauma to the knee. The initial trauma was poorly described by the patient; however, there appeared to be significant force applied to the knee while in deep flexion. This resulted in pain and instability, but there was no radiographic evidence of problems with the components at this time. The patient subsequently had another fall, likely due to the instability and progressively worsening function of her knee. Still, at this point there was no radiographic evidence of femoral component failure. It was not until her next follow-up that there was global instability on examination and radiographic evidence of femoral component failure.
The true cause for failure at this junction of the prosthesis is ultimately unknown. Failure secondary to fatigue of cobalt-chromium femoral component was possible given the combination of the initial trauma suffered by the patient and insufficient deep lateral condyle support producing a stress riser. Most fatigue fractures reported in the literature occurred 4-10 years after implantation and were often related to material properties, component design, high loading conditions, and/or poor fixation [17,18]. It is possible that additional cement, metal augments, rebar, and/or a femoral cone could have improved deep support for the lateral femoral condyle and decreased metal fatigue; however, considering the trauma was low energy and the time to failure was less than 2 years, this would be an unusual presentation of this mechanism, thus there was likely an additional contributing factor.
For example, the bumper could have been mal-seated, allowing for increased forces across the axle and lateral condyle; however, at the time of revision surgery, this did not appear to be the case, and the entirety of the hinge mechanism traveled with the lateral aspect of the implant where the fracture occurred. This device is predicated on previous MRH designs such as the Modular Rotating Hinge Knee (Howmedica Osteonics; Allendale, NJ), which was predicated from the Kinematic Rotating Hinge Total Knee Pr. (Howmedica Osteonics; Walker, MI) before that [[19], [20], [21]]. Although for approval by the Food and Drug Administration, this MRH passed magnetically induced displacement force tests and magnetically induced torque tests [21], there may have been an issue with casting of the component at the time of its manufacturing leading to an area of weakness within the lateral aspect of the implant, leaving it susceptible to fracture. Ultimately, the cause of this catastrophic failure of the lateral femoral condyle was likely multifactorial.
Summary
Re-revision of an MRH is most commonly due to patellofemoral complications, aseptic loosening, and infection. Most cases of mechanical failure for these hinged constructs occur at the hinge component. This is the first report describing a femoral component failure in a MRH via a low-energy mechanism.
Funding
Duke University provided funding for this study.
Conflicts of interest
M. P. Bolognesi receives royalties from Smith & Nephew, Total Joint Orthopedics, and Zimmer; is an unpaid consultant for Amedica; has stock options in Amedica and Total Joint Orthopedics; receives research support from Biomet, Exactech, Inc., KCI, Zimmer, and DePuy; receives other financial and material support from Acelity and AOA Omega; is an editorial board member of Journal of Arthroplasty and Arthroplasty Today; and is a board/committee member of AAHKS, EOA, and AAOS. T. Seyler receives royalties from Total Joint Orthopedics, Inc., Pattern Health, and Restor3d; is a paid consultant for Smith and Nephew, Total Joint Orthopedics, Inc., and Heraeus; receives research support from Next Science and Zimmer; receives royalties and financial support from Lippincott Williams and Wilkins; and is a board/committee member of the American Association of Hip and Knee Surgeons and Musculoskeletal Infection Society. S. Wellman is a speaker bureau of TJO and Zimmer; is a paid consultant for Smith & Nephew and TJO; has stock options in Joint Development, LLC; receives research support from Biomet, DePuy, A Johnson & Johnson Company, Medacta, Smith & Nephew, Stryker, and Zimmer; receives royalties/financial support from TJO; is an editorial board member of the Journal of Arthroplasty; and is a board/committee member of the American Association of Hip and Knee Surgeons. All other authors declare no potential conflicts of interest.
For full disclosure statements refer to https://doi.org/10.1016/j.artd.2024.101495.
Informed patient consent
The author(s) confirm that written informed consent has been obtained from the involved patient(s) or if appropriate from the parent, guardian, power of attorney of the involved patient(s); and, they have given approval for this information to be published in this case report (series).
CRediT authorship contribution statement
Justin Leal: Writing – review & editing, Writing – original draft, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Rahul K. Goel: Writing – review & editing, Writing – original draft, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Michael P. Bolognesi: Writing – review & editing, Supervision, Methodology, Investigation, Conceptualization. Samuel S. Wellman: Writing – review & editing, Supervision, Methodology, Investigation, Data curation, Conceptualization. Thorsten M. Seyler: Writing – review & editing, Investigation, Data curation, Conceptualization. Sean P. Ryan: Writing – review & editing, Writing – original draft, Supervision, Methodology, Investigation, Formal analysis, Data curation, Conceptualization.
Appendix A. Supplementary data
References
- 1.Kurtz S., Ong K., Lau E., Mowat F., Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89:780–785. doi: 10.2106/JBJS.F.00222. [DOI] [PubMed] [Google Scholar]
- 2.Salari P., Baldini A. Revision knee surgery: the practical approach. EFORT Open Rev. 2021;6:495–500. doi: 10.1302/2058-5241.6.210018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Cottino U., Abdel M.P., Perry K.I., Mara K.C., Lewallen D.G., Hanssen A.D. Long-Term results after total knee arthroplasty with contemporary rotating-hinge prostheses. J Bone Joint Surg Am. 2017;99:324–330. doi: 10.2106/JBJS.16.00307. [DOI] [PubMed] [Google Scholar]
- 4.Farid Y.R., Thakral R., Finn H.A. Intermediate-Term results of 142 single-design, rotating-hinge implants: frequent complications may not preclude salvage of severely affected knees. J Arthroplasty. 2015;30:2173–2180. doi: 10.1016/j.arth.2015.06.033. [DOI] [PubMed] [Google Scholar]
- 5.Joshi N., Navarro-Quilis A. Is there a place for rotating-hinge arthroplasty in knee revision surgery for aseptic loosening? J Arthroplasty. 2008;23:1204–1211. doi: 10.1016/j.arth.2007.10.016. [DOI] [PubMed] [Google Scholar]
- 6.Korim M.T., Esler C.N.A., Reddy V.R.M., Ashford R.U. A systematic review of endoprosthetic replacement for non-tumour indications around the knee joint. Knee. 2013;20:367–375. doi: 10.1016/j.knee.2013.09.001. [DOI] [PubMed] [Google Scholar]
- 7.Matar H.E., Bloch B.V., James P.J. Satisfactory medium- to long-term outcomes of cemented rotating hinge prosthesis in revision total knee arthroplasty: a specialist centre study with minimum four years’ follow-up. Bone Jt Open. 2023;4:776–781. doi: 10.1302/2633-1462.410.BJO-2022-0032.R1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Rodríguez-Merchán E.C. Total knee arthroplasty using hinge joints: indications and results. EFORT Open Rev. 2019;4:121–132. doi: 10.1302/2058-5241.4.180056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Von Hintze J., Niemeläinen M., Sintonen H., Nieminen J., Eskelinen A. Outcomes of the rotating hinge knee in revision total knee arthroplasty with a median follow-up of 6.2 years. BMC Musculoskelet Disord. 2021;22:336. doi: 10.1186/s12891-021-04205-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Pasquier G., Ehlinger M., Mainard D. The role of rotating hinge implants in revision total knee arthroplasty. EFORT Open Rev. 2019;4:269–278. doi: 10.1302/2058-5241.4.180070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Wang C.-J., Wang H.E. Early catastrophic failure of rotating hinge total knee prosthesis. J Arthroplasty. 2000;15:387–391. doi: 10.1016/S0883-5403(00)90877-5. [DOI] [PubMed] [Google Scholar]
- 12.Biswas D., Haughom B., Mayle R.E., Della Valle C.J. Case report: failure of rotating-hinge total knee prosthesis by disengagement of the hinge-post extension. Clin Orthop Relat Res. 2013;471:1389–1392. doi: 10.1007/s11999-012-2736-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Friesenbichler J., Schwarzkopf R., Sadoghi P., Marwin S.E., Glehr M., Maurer-Ertl W., et al. Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature. Int Orthop. 2012;36:993–998. doi: 10.1007/s00264-011-1462-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Ward W.G., Haight D., Ritchie P., Gordon S., Eckardt J.J. Dislocation of rotating hinge knee prostheses: a report of four cases. J Bone Joint Surg Am. 2005;87:1108–1112. doi: 10.2106/JBJS.00837pp. [DOI] [PubMed] [Google Scholar]
- 15.Rand J.A. Revision total knee arthroplasty using the total condylar III prosthesis. J Arthroplasty. 1991;6:279–284. doi: 10.1016/S0883-5403(06)80175-0. [DOI] [PubMed] [Google Scholar]
- 16.Pacha-Vicente D., Malik A., Castellet-Feliu E., Nardi-Vilardaga J. Dislocation of rotating-hinge knee prostheses with antidislocation mechanism. J Arthroplasty. 2008;23:299–303. doi: 10.1016/j.arth.2006.11.020. [DOI] [PubMed] [Google Scholar]
- 17.Chithartha K., Thilak J., Sukesh A.N., Theruvil B. Fatigue fracture of the femoral component in total knee replacement. Knee. 2020;27:1439–1445. doi: 10.1016/j.knee.2020.06.009. [DOI] [PubMed] [Google Scholar]
- 18.Saito S., Tokuhashi Y., Ishii T., Mori S., Hosaka K., Ryu K., et al. Bilateral fatigue fracture of the femoral components in a cruciate-retaining cementless total knee prosthesis. Orthopedics. 2011;34:e688–e691. doi: 10.3928/01477447-20110826-26. [DOI] [PubMed] [Google Scholar]
- 19.Howmedica Corp . K811630_Kinematic rotating hinge total knee PR._510k. FDA; Silver Spring, MD: 1981. [Google Scholar]
- 20.Howmedica Osteonics Corp . FDA; Silver Spring, MD: 1999. K994207_Modular rotating hinge Knee_510k. [Google Scholar]
- 21.Howmedica Osteonics Corp. DBA Stryker Orthopaedics . FDA; Silver Spring, MD: 2023. K223069_Modular rotating hinge knee System_510k. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.