Skip to main content
. 2024 Oct 9;10(20):e39145. doi: 10.1016/j.heliyon.2024.e39145

Table 2.

Ibuprofen validation parameters in plasma (n = 5).

Intra-assay variability
Concentration (μg/mL) Measured conc. (Mean ± SD) Accuracy (%) RSD (%)
0.05 0.05 ± 0.002 100 4.0
0.15 0.15 ± 0.01 100 9.9
7.5 8.2 ± 0.07 109 0.8
35 35 ± 1.3 100 3.8
75 76 ± 1.98 101 2.6
Inter-assay variability

0.05 0.05 ± 0.004 100 8.0
0.15 0.15 ± 0.01 100 6.2
7.5 7.5 ± 0.74 100 9.8
35 35 ± 1.6 100 4.6
75 76 ± 2.0 101 2.7

Recovery Recovery ± SD RSD (%)

0.05 98 ± 10.3 10.6
0.15 97 ± 10.1 10.0
7.5 95 ± 1.9 2.0
35 97 ± 5.3 5.4
75 96 ± 6.3 6.6

n: number of samples; SD: standard deviation; RSD: relative standard deviation.