TABLE 1.

Summary of method validation results.

Criteria 3‐HG 2‐HG GA
Limit of quantification (LOQ) 1 μmol/L 5 μmol/L 6 μmol/L
Linearity (r 2)—0 to 400 μmol/L (3‐HG 0–100 μmol/L) 0.996 0.992 0.995
Measurement range limits 1–500 μmol/L a 5–1000 μmol/L a 6–1000 μmol/L a
Precision (intra‐batch) b
Low value 13.3% 6.4% 3.5%
High value 5.5% 4.2% 5.3%
Precision (inter‐day) c
Low value 10.2% 13.7% 14.0%
High value 14.3% 13.8% 14.5%
Accuracy 84% 85% 97%
Recovery d
Low value 85% 97% 82%
Medium value 91% 113% 102%
High value 97% 116% 116%

Note: Accuracy was evaluated using samples from a quantitative external quality program (ERNDIM).

a

Highest tested concentration.

b

Low value 3 μM (3‐HG), 20 μM (2‐HG), 10 μM (GA). High value 200 μM for all three acids. Results expressed as relative standard deviation (RSD%).

c

Low value 5 μM (3‐HG), 27 μM (2‐HG), 14 μM (GA). High value 124 μM (3‐HG), 190 μM (2‐HG), 216 μM (GA). Results expressed as relative standard deviation (RSD%).

d

Low value 5 μM (3‐HG), 10 μM (2‐HG and GA). Median value 40 μM (3‐HG), 80 μM (2‐HG and GA). High value 100 μM (3‐HG), 200 μM (2‐HG and GA).