Skip to main content
. 2024 Oct 29;29(21):5111. doi: 10.3390/molecules29215111

Table 2.

Average recovery rates (R¯%) and relative standard deviations (RSD) of four IPR laboratory tests for PCBs, along with the acceptance criteria for R% and RSD% according to EPA Method 1668B.

Laboratory Test Results EPA 1668B Accepted Range
IPR R¯% IPR RSD IPR R¯% IPR RSD
60L 77 19 43–106 63
127L 84 10 75–102 23
159L 88 11 78–117 30