. 2024 Nov 11;23:114. doi: 10.1186/s12938-024-01309-w

Table 2.

Accuracy of pulse rate of Corsano CardioWatch 287–2 compared to Vivalink ECG patch

Pooled (n = 7183) Female (n = 2593) Male (n = 4590) Fitzpatrick I-IV (n = 4993) Fitzpatrick V-VI (n = 2190)
Arms [bpm] 2.51 2.59 2.46 2.54 2.45
Bias (+ 95% CI) [%] 0.05 (−0.01, 0.11) −0.10 (−0.21, 0.00) 0.14 (0.07, 0.21) 0.00 (−0.07, 0.07) 0.16 (0.06, 0.26)
95% LoA (+ 95% CI) [%] Lower −4.87 (−4.93, −4.81) −5.19 (−5.28, −5.09) −4.68 (−4.75, −4.61) −4.98 (−5.05, −4.91) −4.62 (−4.72, −4.52)
95% LoA (+ 95% CI) [%] Upper 4.97 (4.91, 5.02) 4.97 (4.87, 5.07) 4.96 (4.88, 5.02) 4.98 (4.91, 5.05) 4.94 (4.84, 5.05)

n number of samples, Arms Average root mean square, CI confidence interval, LoA Limits of Agreement