Normal Value of Perfusion Index in Healthy Neonates Born in Iran

Maryam Saeedi 1, Razieh Sangsari 1, Kayvan Mirnia 1,*, Monireh Ghanbari 1
PMCID: PMC11558613  PMID: 39534994

Abstract

Background:

Perfusion index is a dependable indicator for assessing the perfusion status of newborns. A low perfusion index indicates compromised hemodynamic function. The study aims to investigate perfusion index in asymptomatic newborns aged 35 to 41 weeks who did not require medical support.

Methods:

Healthy neonates born in four major maternity hospitals of Tehran University from 2019 to 2021 were selected. To ensure consistency and reliability in data collection, a detailed manual was developed and distributed, along with comprehensive training sessions for all personnel involved.

Results:

A total of 994 newborns entered the study. Among them, echocardiography was adversely affected in 218 neonates due to abnormal screening pulse oximetry. Of these 218 neonates, 53 were found to have abnormal echocardiography results. The median perfusion index value in healthy neonates was 1.6%, and the median oxygen saturation was 97%. A percentile perfusion index curve was developed for healthy neonates to establish a normal reference range.

Conclusion:

Developing a percentile perfusion index curve specific to healthy neonates provides a useful reference range for healthcare providers to assess perfusion status in this population, but further research is needed to confirm its accuracy.

Keywords: Newborn, Oxygen saturation, Perfusion index, Median PI value, PI normal range

Introduction

Evaluating the health of newborns becomes more effective and convenient with the use of pulse oximetry, a non-invasive tool that monitors neonatal heart rate (HR) and oxygen saturation (SpO2). In addition, the Perfusion Index (PI) exhibited during pulse oximetry monitoring serves as a means to assess pulse intensity by calculating the pulsatile (AC) and non-pulsatile (DC) ratio signals.1,2 Monitoring PI can aid in evaluating neonatal hemodynamics, particularly in critically ill infants.3 A decrease in PI can indicate the severity of illness, as changes in stroke volume and vasomotor tone influence this measure. Lower PI values are associated with reduced perfusion and can serve as an early detection tool for adverse outcomes during the neonatal period.4 Research indicates that lower PI in neonates correlates with diminished superior vena cava flow; serving as a predictor of volume status and suggesting a deterioration of the newborn’s condition.5 Interestingly, newborns with transient tachypnea of the newborn (TTN) may exhibit lower PI values compared to healthy counterparts. Monitoring PI offers insights into subtle changes in perfusion that may be overlooked by static displays.6,7 To accurately assess potential pathological conditions, it is essential to understand the normal range of PI. Studies have shown that PI can vary based on factors such as the neonate’s gender, weight,8,9 chronological age, and gestational age (GA).10 Moreover, PI can range from 0.02% to 20%, depending on the monitoring site and patient’s age.11 However, further research is necessary to establish the median PI in infants, particularly in healthy term neonates within the first 24 hours after birth. This study aims to map peripheral PI distribution in asymptomatic neonates with GA of 35 to 41 weeks who did not need any medical treatment. This study aims to plot PI normal range values in healthy neonates in Iran.

Material and Methods

This observational cross-sectional survey aimed to assess physiological indicators (PI) among asymptomatic, healthy newborns aged 35 to 41 weeks who were 24 hours old at the time of enrollment. To ensure uniformity in the studied population and increase the strength of the study, we selected four major maternity hospitals affiliated with Tehran University. Out of the six maternity hospitals associated with the university, we chose these four from different locations across Tehran. Consistency in data collection practices was maintained across all participating hospitals to enhance the reliability and validity of the research findings. To achieve this goal, we developed and distributed a detailed data collection manual that outlined standardized procedures, definitions, and methodologies for data collection. Additionally, we conducted comprehensive training sessions for all personnel involved in data collection at each hospital. Participants in the study included healthy newborns delivered either via vaginal delivery or elective cesarean section, which did not require any medical intervention at birth. The study population comprised healthy newborns delivered at the four hospitals over a two-year period, from 2019 to 2021. All newborns received standard postnatal care, facilitated by experienced midwives at each hospital, in accordance with unit guidelines. This approach aimed to eliminate confounding factors such as variability in caregiver experience and care quality, which could affect PI measurements.

Newborns with pre- or post-ductal SpO2 levels less than 95%, a pre- to post-ductal SpO2 gradient greater than 3%, or a PI value less than 0.9 underwent echocardiography12 and were excluded from the study. Additionally, neonates presenting with congenital anomalies or those who were discharged before reaching 24 hours of life were also excluded from the study. Healthy neonates were defined as those not requiring resuscitation or medical support after birth, those not born with meconium-stained amniotic fluid, and those who did not die within two days of delivery.

The pediatric resident utilized a Masimo Pulse Oximeter (Radical-7, USA) to document PI, HR, and SpO2. The palm or wrist on the right side was the chosen placement for pre-ductal monitoring, while either foot was used for post-ductal monitoring. SpO2 and HR values were recorded when the pulse oximeter displayed stable waveforms, along with the accompanying PI values. In cases where PI values were invalid ( ≤ 0.02 or ≥ 20%), measurements were repeated until valid values were obtained. Additionally, demographic data such as sex, birth weight (BW), and GA were recorded for each participant. Sample size calculation followed the methodology of Hu and colleagues’ study,10 which reported a standard deviation (SD) of 1.26. With a margin of error (d) set at 0.08, a total sample size of 953 patients was determined for adequate statistical power.

Statistical Analysis

We used SPSS 26.0 for Windows for analysis. The paired t test was used to compare characteristic data. Multiple linear regressions showed PI value relationship with SpO2, while adjusting for sex, birth weight, and GA. Statistical significance was determined as P < 0.05. To illustrate the variation of PI between GAs in a more meaningful way, percentile curves were created. Cole’s Lambda Mu Sigma (LMS) method was applied to plot these curves at the 3rd, 10th, 25th, 50th, 75th, 90th, and 97th percentiles. The LMS method utilizes the Box-Cox transformation of the three parameters to achieve a near-normal distribution of the PI measurement. The L parameter introduces a non-linear change to PI, the M parameter represents the mean of the normal distribution, and the S parameter represents the coefficient of variation. As the covariate shifts, the three parameters must adjust in a consistent manner.13 LMS chart maker is a program to fit smooth centile curves to reference data using the LMS method.10 Non-linear regression combined with penalized likelihood allows the three curves to be modeled as cubic splines, and the amount of smoothing required can be represented by smoothing parameters or the same number of freedoms.14 The process consists of five stages, namely data entry, model fitting, graphical display, model checking, and model saving, with the last four stages being repeated as many times as required.

The software used for this analysis and modeling was the LMS chart maker program developed by Pan H and Cole TJ. version 2.54. The program was obtained from the website https://lmschartmaker.software.informer.com/2.5/ in 2011.

Results

In our study, 994 newborns with a GA of 35‒41 weeks entered the study. Of these newborns, 49.9% were female and 50.1% were male. The average GA was 38.5 ± 1.26 weeks. The mean BW was 3209 ± 449 grams. Among the newborns, 218 had abnormal pulse oximetry screening, pre- or post-ductal SpO2 < 95%, the pre-post ductal SpO2 gradient > 3%, or the PI < 0.9 and an echocardiography was performed. In these cases, 53 newborns had abnormal echocardiography findings. In healthy newborns, both pre- and post-ductal PI had a median value of 1.6%. The median values of both pre- and post-ductal arterial oxygen saturation (SPO2) in healthy newborns were 97%. In neonates with congenital heart defect (CHD), the median pre-ductal and post-ductal PI was 1%, and the median pre-ductal and post-ductal SPO2 was 95% (Table 1). The mean pre-ductal and post-ductal PI was calculated based on the presence of cardiac disease. The mean of PI showed a significant relationship with the presence of CHD (pre-duct P value = 0.001) and (post-duct P value = 0.001), but there was no significant relationship with other variables (Table 2). Pearson’s correlation coefficients were used to determine the relationship between PI and SPO2. This analysis revealed a strong positive relationship between pre-ductal SPO2 and pre-ductal PI in neonates with CHD (r = 1, P = 0.04) and a strong positive relationship between post-ductal SPO2 and PI in healthy neonates (r = 1, P = 0.03). In both cases, higher SPO2 was associated with higher PI (Table 3). The mean pre-ductal and post-ductal PI in healthy neonates were 1.8% and 1.7%, while the mean pre-ductal and post-ductal PI in neonates with CHD were 1.17% and 1.3%, respectively (Table 1). Given the sample size and the obtained information, we created a percentile PI curve in healthy neonates to be used as a normal reference value (Figure 1).

Table 1. Distribution of Pre- and Post-ductal PI and SPO2 .

Right Hand Sat Foot Sat Right Hand PI Foot PI
Healthy N 941 941 941 941
Mean 97.014 97.0478 1.8340 1.7751
Median 97.000 97.0000 1.6000 1.6000
Standard deviation 2.0172 2.19788 0.81782 0.74569
Minimum 88.0 63.00 0.09 0.30
Maximum 100.0 100.00 7.90 7.50
CHD N 53 53 53 53
Mean 95.642 95.7925 1.1732 1.3019
Median 95.000 95.0000 1.0000 1.0000
Standard deviation 3.3345 3.62342 0.68143 0.81700
Minimum 82.0 78.00 0.08 0.10
Maximum 100.0 100.00 4.50 5.20

CHD, Congenital heart defect; PI, Perfusion index; SPO2, Oxygen saturation.

Table 2. Linear Regression for Pre- and Post-ductal PI .

Model Unstandardized Coefficients Standardized Coefficients T Significant 95% CI
B Standard Error Beta
Dependent Variable: Right hand PI
Echocardiography -0.630 0.088 -0.226 -7.141 0.0000 -0.803, -0.457
Gender 0.0530 0.040 0.042 1.308 0.1910 -0.26, 0.132
GA (wk) 0.0150 0.018 0.028 0.799 0.4250 -0.21,0.051
BW (kg) 0.0570 0.049 0.040 1.158 0.2470 -0.39, 0.153
Dependent Variable: Foot PI
Echocardiography -0.459 0.0820 -0.179 -5.614 0.0000 -0.62, -0.299
Gender 0.022 0.0370 0.019 0.5990 0.5490 -0.51, 0.096
GA (wk) 0.018 0.0170 0.037 1.056 0.2910 -0.015, 0.052
BW (kg) 0.047 0.0450 0.036 1.029 0.3040 -0.042, 0.136

B, beta coefficient; T, T-value; PI, Perfusion index.

Table 3. Correlation between SPO2 and PI .

Right Hand Sat Right Hand PI
Healthy Right hand sat Pearson Correlation 1 -0.060
Sig. (2-tailed) 0.064
Right hand PI Pearson Correlation -0.060 1
Sig. (2-tailed) 0.064
CHD Right hand sat Pearson Correlation 1 -0.282*
Sig. (2-tailed) 0.040
Right hand PI Pearson Correlation -0.282* 1
Sig. (2-tailed) 0.040
Healthy Foot sat Pearson Correlation 1 -0.069*
Sig. (2-tailed) 0.034
Foot PI Pearson Correlation -0.069* 1
Sig. (2-tailed) 0.034
CHD Foo. sat Pearson Correlation 1 -0.169
Sig. (2-tailed) 0.227
Foot PI Pearson Correlation -0.169 1
Sig. (2-tailed) 0.227

* Correlation is significant at the 0.05 level (2-tailed).

CHD, Congenital heart defect, PI, Perfusion index.

Figure 1.

Figure 1

Percentile Pre- and Post-ductal PI Curve in Healthy Male and Female Neonates

Discussion

In healthy neonates, skin perfusion is higher than the oxygen demand. However, when illness occurs, the body reallocates cardiac output to ensure that essential organs such as the brain, heart, and adrenal glands receive sufficient oxygen. This period, before perfusion reduction becomes irreversible, is known as the golden time. Identifying neonates during this stage can enable prompt treatment initiation.13,14 Pulse oximetry PI is being increasingly recognized as a useful parameter to enhance sensitivity in detecting significant neonatal complications. Many studies concentrate on utilizing PI to forecast life-threatening illnesses.4,15 Given the potential of PI in diagnosing the severity of a newborn’s disease and screening for heart conditions, it is essential to study the normal range of PI in healthy neonates.

In 2017, Jegatheesan et al conducted a study in the United States that employed a computerized data selection method to find the median PI in asymptomatic neonates who were 24 hours old. Their study yielded a median PI of 1.8%, displaying a tight interquartile range (IQR) between 1.2 and 2.7.16 The nomogram in Granelli and Ostman-Smith study’s, which reported data from asymptomatic newborns at 1 to 120 hours of life, showed a median PI of 1.7% and an IQR of 1.18 to 2.5. These findings are consistent with our own research that showed PI with a median of 1.6% and an IQR of 1.3 to 2.1. These consistent distributions of PI suggest that PI values in asymptomatic neonates exhibit a limited range of distribution during screening time.17 Hawkes and colleagues found that PI had a wider range during the first 5 minutes after birth compared to the period after.18

Our study showed that the median of pre-ductal and post-ductal PI in healthy neonates was 1.6%, and the median of pre-ductal and post-ductal SPO2 in healthy neonates was 97%. Although the Jegatheesan et al study reported a post-ductal PI that was 0.1 lower than the pre-ductal PI,16 and the Granelli and Ostman-Smith’s study showed a post-ductal PI that was 0.02 higher than the pre-ductal PI17, Hua et al reported a median pre-ductal and post-ductal PI of 1.7%.19 These small differences depend on the chronological age and the GA of the studied healthy newborns.

In our neonates with CHD, the median of pre- and post-ductal PI was 1%. The mean of PI had a significant relationship with the presence of heart disease (pre-ductal P value = 0.000) and (post-ductal P value = 0.000). A study conducted by Iuri Corsini and colleagues reported a significant correlation between the PI and left ventricular output in term infants.20 Schena et alconducted a prospective study in 16 hospitals where a total of 42,169 asymptomatic newborns were screened using pre- and post-ductal SPO2 and PI. The study concluded that PI can be useful in identifying cases of coarctation of the aorta that may have been missed by pre- and post-ductal SPO2 measurements.12 In another cross-sectional study in patent ductus arteriosus neonates, a significant difference was observed in pre- and post-ductal PI before and after arterial duct closure (P = 0.004).21 Most of the studies are related to the PI and left heart disorders. The emphasis is on the diagnosis of left heart diseases that cannot be accurately diagnosed solely by pre- and post-ductal SPO2 measurements,22 but in our study there was no preference for left heart disease. A case of total anomalous pulmonary venous return had pre-ductal PI of 1% and post-ductal PI of 0.9%. Although these PIs are higher than the number required for the screening of CHD, they are actually lower than the normal range typically found in healthy newborns. Therefore, this study also emphasizes the inclusion of PI in pre- and post-ductal SPO2 measurements for CHD screening in healthy neonates.

In this study, we designed percentile PI curves in healthy neonates to be used as normal values (Figure 1). In 2020, a study carried out in China provided data on the percentiles of pre- and post-duct PI for neonates with GA of 35 to 41 weeks, with a specific focus on sex differences.8 The presented percentile curves can aid medical professionals in determining the standard PI value depending on the GA and gender. This can potentially lead to the early identification of high-risk neonates.

Limitation

This study faced some limitations that, if addressed, could have led to more thorough results. Notable limitations included possible biases in data collection, the individuals responsible for data collection were not identical, and variations in demographics. Initially, an extended follow-up period for the newborns might have offered valuable information regarding their health, making it a crucial aspect of the study. Additionally, a larger sample size could have been included, which would have increased the statistical power and reliability of the findings.

Conclusion

The newborn-specific PI percentile curve is important for neonatologists as it aids in the early diagnosis of high-risk neonates. However, further research is necessary to validate the accuracy of this curve.

Acknowledgments

We would like to thank the personnel of the four hospitals that carried out the measurement.

Cite this article as: Saeedi M, Sangsari R, Mirnia K, Ghanbari M. Normal value of perfusion index in healthy neonates born in Iran. Arch Iran Med. 2024;27(11):606-610 doi: 10.34172/aim.31293

Footnotes

Competing Interests

The authors declare no conflict of interest.

Ethical Approval

This study was approved by the Tehran University of Medical Sciences’ Ethics Committee. ID: IR.TUMS.CHMC.REC.1398.131

Funding

This research did not receive any specific grant from funding agencies.

References

  • 1.Piasek CZ, Van Bel F, Sola A. Perfusion index in newborn infants: a noninvasive tool for neonatal monitoring. Acta Paediatr. 2014;103(5):468–73. doi: 10.1111/apa.12574. [DOI] [PubMed] [Google Scholar]
  • 2.Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Crit Care Med. 2002;30(6):1210–3. doi: 10.1097/00003246-200206000-00006. [DOI] [PubMed] [Google Scholar]
  • 3.Ibrahim M, Mohamed M. Validity of perfusion index in prediction of circulatory compromise and mortality in neonates. Ain Shams Med J. 2023;74(1):57–64. doi: 10.21608/asmj.2023.298198. [DOI] [Google Scholar]
  • 4.Costa Monteiro S, Correia-Costa L, Proença E. Perfusion index in preterm newborns during the first week of life and association with neonatal morbimortality: a prospective observational study. J Pediatr Neonat Individual Med. 2017;6(2):e060212. doi: 10.7363/060212. [DOI] [Google Scholar]
  • 5.Takahashi S, Kakiuchi S, Nanba Y, Tsukamoto K, Nakamura T, Ito Y. The perfusion index derived from a pulse oximeter for predicting low superior vena cava flow in very low birth weight infants. J Perinatol. 2010;30(4):265–9. doi: 10.1038/jp.2009.159. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Unal S, Ergenekon E, Aktas S, Beken S, Altuntas N, Kazanci E, et al. Perfusion index assessment during transition period of newborns: an observational study. BMC Pediatr. 2016;16(1):164. doi: 10.1186/s12887-016-0701-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Kroese JK, van Vonderen JJ, Narayen IC, Walther FJ, Hooper S, te Pas AB. The perfusion index of healthy term infants during transition at birth. Eur J Pediatr. 2016;175(4):475–9. doi: 10.1007/s00431-015-2650-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Nishimura T, Nakae A, Shibata M, Mashimo T, Fujino Y. Age-related and sex-related changes in perfusion index in response to noxious electrical stimulation in healthy subjects. J Pain Res. 2014;7:91–7. doi: 10.2147/jpr.s57140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Knobel-Dail RB, Tanaka DT, Holditch-Davis D, White J. Perfusion index in very low birth weight premature infants during their first 2 weeks of life. Biol Res Nurs. 2017;19(1):45–52. doi: 10.1177/1099800416656914. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Hu XJ, Ding JX, Wang Y, Niu C, Zhang Y, Zhao QM, et al. Peripheral perfusion index percentiles for healthy newborns by gestational age and sex in China. Sci Rep. 2020;10(1):4213. doi: 10.1038/s41598-020-60741-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Mandala VK, Mendu SB, Bollaboina SK, Kotha R Sr. Role of perfusion index and pulse variability index in the assessment of neonatal hemodynamics: a systematic review. Cureus. 2023;15(10):e48058. doi: 10.7759/cureus.48058. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Schena F, Picciolli I, Agosti M, Zuppa AA, Zuccotti G, Parola L, et al. Perfusion index and pulse oximetry screening for congenital heart defects. J Pediatr 2017;183:74-9.e1. 10.1016/j.jpeds.2016.12.076. [DOI] [PubMed]
  • 13.Piasek CZ, Van Bel F, Sola A. Perfusion index in newborn infants: a noninvasive tool for neonatal monitoring. Acta Paediatr. 2014;103(5):468–73. doi: 10.1111/apa.12574. [DOI] [PubMed] [Google Scholar]
  • 14.Badurdeen S, Roberts C, Blank D, Miller S, Stojanovska V, Davis P, et al. Haemodynamic instability and brain injury in neonates exposed to hypoxia–ischaemia. Brain Sci. 2019;9(3):49. doi: 10.3390/brainsci9030049. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Costa Monteiro S, Correia-Costa L, Proença E. Perfusion index in preterm newborns: predictive value for morbimortality and association with Apgar score at five minutes and CRIB-II score. J Pediatr Neonat Individual Med. 2019;8(1):e080104. doi: 10.7363/080104. [DOI] [Google Scholar]
  • 16.Jegatheesan P, Nudelman M, Goel K, Song D, Govindaswami B. Perfusion index in healthy newborns during critical congenital heart disease screening at 24 hours: retrospective observational study from the USA. BMJ Open. 2017;7(12):e017580. doi: 10.1136/bmjopen-2017-017580. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Granelli A, Ostman-Smith I. Noninvasive peripheral perfusion index as a possible tool for screening for critical left heart obstruction. Acta Paediatr. 2007;96(10):1455–9. doi: 10.1111/j.1651-2227.2007.00439.x. [DOI] [PubMed] [Google Scholar]
  • 18.Hawkes GA, O’Toole JM, Kenosi M, Ryan CA, Dempsey EM. Perfusion index in the preterm infant immediately after birth. Early Hum Dev. 2015;91(8):463–5. doi: 10.1016/j.earlhumdev.2015.05.003. [DOI] [PubMed] [Google Scholar]
  • 19.Hua W, Niu C, Xuan Y, Zhao QM, Ren Y, Hu X, et al. Peripheral perfusion index in well newborns at 6 to 72 h of life at different altitudes: a multi-center study in China. Eur J Pediatr. 2023;182(2):907–15. doi: 10.1007/s00431-022-04725-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Corsini I, Cecchi A, Coviello C, Dani C. Perfusion index and left ventricular output correlation in healthy term infants. Eur J Pediatr. 2017;176(8):1013–8. doi: 10.1007/s00431-017-2920-1. [DOI] [PubMed] [Google Scholar]
  • 21.Sangsari R, Dalili H, Kadivar M, Saeedi M, Mirnia K, Fathi A, et al. Evaluation of the relationship between perfusion index and the improvement of patent ductus arteriosus. Iran J Pediatr. 2023;33(2):e134709. doi: 10.5812/ijp-134709. [DOI] [Google Scholar]
  • 22.Siefkes H, Kair L, Tancredi DJ, Vasquez B, Garcia L, Bedford-Mu C, et al. Oxygen saturation and perfusion index-based enhanced critical congenital heart disease screening. Am J Perinatol. 2020;37(2):158–65. doi: 10.1055/s-0039-1685445. [DOI] [PubMed] [Google Scholar]

Articles from Archives of Iranian Medicine are provided here courtesy of Academy of Medical Sciences of the I.R. of Iran