Skip to main content
. 2024 Sep 27:15910199241286547. Online ahead of print. doi: 10.1177/15910199241286547

Figure 3.

Figure 3.

Patient 2. (A) The initial selective digital subtraction angiogram from the left vertebral artery, noting the left P1 occlusion. (B and C) The position of the Rebar18 microcatheter (with yellow arrow pointing to the distal tip). Immediately above these images is noted the heart rate (green), blood pressure from radial artery line (white), and manometry waveforms from the Rebar18 microcatheter (blue). Trans-clot manometry shows a MAP of 3 mmHg distal to the clot, in comparison to the MAP of 97 mmHg proximal to the clot. (D) The successful first-pass recanalization with TICI score of 3.