Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2025 Dec 1.
Published in final edited form as: Nat Methods. 2024 May 14;21(12):2388–2396. doi: 10.1038/s41592-024-02279-6

Top-down mass spectrometry of native proteoforms and their complexes: A community study

Tanja Habeck 1, Kyle A Brown 2, Benjamin Des Soye 3, Carter Lantz 4, Mowei Zhou 5,, Novera Alam 6, Md Amin Hossain 6, Wonhyeuk Jung 4, James E Keener 7, Michael Volny 8, Jesse W Wilson 5, Yujia Ying 9, Jeffrey N Agar 6,10, Paul O Danis 10, Ying Ge 2,10, Neil L Kelleher 3,10, Huilin Li 9, Joseph A Loo 4,10, Michael T Marty 7, Ljiljana Paša-Tolić 5,10, Wendy Sandoval 8, Frederik Lermyte 1,*
PMCID: PMC11561160  NIHMSID: NIHMS2002388  PMID: 38744918

Abstract

The combination of native electrospray ionisation with top-down fragmentation in mass spectrometry allows simultaneous determination of the stoichiometry of noncovalent complexes and identification of their component proteoforms and co-factors. While this approach is powerful, both native mass spectrometry and top-down mass spectrometry are not yet well standardised, and only a limited number of laboratories regularly carry out this type of research. To address this challenge, the Consortium for Top-Down Proteomics (CTDP) initiated a study to develop and test protocols for native mass spectrometry combined with top-down fragmentation of proteins and protein complexes across eleven instruments in nine laboratories. Here we report the summary of the outcomes to provide robust benchmarks and a valuable entry point for the scientific community.

Introduction

A single gene can give rise to many distinct proteoforms.13 The need to accurately identify and structurally characterise these proteoforms has led to significant technological developments in recent years, most notably in so-called ‘top-down’ mass spectrometry (MS) of intact proteins.49 This term refers to measurement of the intact mass of the protein, followed by gas-phase fragmentation to obtain sequence information. In 2021, the Human Proteoform Project was conceived, with the ambitious goal of producing a high-quality atlas of human proteoforms.10 However, in addition to the precise primary structure of a proteoform, it is also critical to characterise its interactions and corresponding noncovalently bound functional complexes with other proteins and/or co-factors. This information is lost in most top-down MS experiments, which are carried out under denaturing solution conditions. Conversely, ‘native’ MS, in which noncovalent complexes are preserved during transfer into the gas phase, offers a powerful method to characterise interactions between proteins and to determine the stoichiometry of complexes.1115 The power of the combination of native ionisation with extensive top-down proteoform characterisation has been convincingly demonstrated in recent years, for example in the identification of novel endogenous complexes, previously unknown co-factors, and regulatory post-translational modifications.1620 However, the use of this powerful integrated approach is still rare.

The limited adoption of this approach, known as ‘native top-down’,21 ‘nativeomics’,18 and ‘complex-down’2225, is primarily due to the perception that both native and top-down MS are challenging, let alone the combination of the two. Although efforts have been made in recent years to make both methods more widely applied,4, 6, 7, 2628 neither has been standardised to the same extent as conventional bottom-up proteomics. Furthermore, even if one overcomes the barrier to entry that is created by the lack of standardisation and successfully carries out the different steps of this type of experiment, i.e., native ionisation (with electrospray ionisation), gas-phase disassembly of the complex, subsequent backbone fragmentation, and analysis of the resulting data (see Figure 1), there is still the issue of assessing the quality of the results. The relatively sparse relevant literature makes it difficult for a newcomer to the field to identify appropriate benchmark data, ideally acquired on a similar instrument with a protein that is conveniently available.

Figure 1. Summary of the protocol for top-down analysis of natively ionised proteins used by study participants.

Figure 1.

Optional steps are shown in dashed lines, steps specific to water-soluble proteins in gold, and those specific to membrane proteins in blue.

Having identified this unmet need for robust protocols and benchmarks, the Consortium for Top-Down Proteomics (CTDP) launched the ‘Native Top-Down Initiative’ in 2020. This is conceptually similar to other recent community initiatives in mass spectrometry-based proteomics and structural biology.2933 A range of laboratories, covering all major instrument families for protein MS, were invited to perform native ionisation with subsequent top-down fragmentation of a set of six standard proteins that covered a broad mass range from 27 to 800 kDa. These included monomeric as well as multimeric proteins, and water-soluble as well as membrane proteins. The corresponding author conceptualised the study and proposed it to the Consortium for Top-Down Proteomics, and five participants are Directors of the Consortium. Additional participants were invited to provide specific expertise or perspectives, including experience with membrane proteins, an industry-based lab, and representation of a laboratory not based in North America or Europe. It is important to note that not all participants had prior experience with this type of experiment; therefore, the protocols used in this study can demonstrably be implemented by users with various levels of expertise. This report will convincingly lower the entry bar and serve as a convenient benchmark.

Results

Study design

Nine laboratories generated data in this study, with a total of eleven instruments: 5 Orbitraps, 4 quadrupole/time-of-flight (QTOF), and 2 Fourier transform ion cyclotron resonance (FTICR) instruments. At the beginning of the study, before any data collection was performed, five of the participants collaboratively created an optimised protocol which was then used by all nine participants. A visual summary of this protocol is shown in Figure 1, and an overview of the model proteins and acquired datasets is shown in Table 1. The protocol itself can be found in the Supplementary Protocol S1. Participants were allowed free choice of the methods for spectrum deconvolution and fragment assignment. The methods they used are summarised in Supplementary Figure S1. For ion activation and fragmentation, they were instructed to use only collision-based methods to facilitate comparison between datasets. At the start of the study, participants were asked to self-report their experience level with native (top-down) MS. At the end, they were asked to self-report the perceived difficulty of different steps of the experiment. An overview of these survey results can be found in Supplementary Figures S1 and S2. Overall, the responses indicated a relatively broad range of experience levels at the start of the study, as well as a consensus about the relative difficulty of most of the analysis steps, with obtaining extensive backbone fragmentation and analysis of membrane proteins being particularly challenging.

Table 1.

Overview of the datasets generated in this study.

Protein Mass (kDa) Subunits Datasets Average obtained cleavage coverage Typical native precursor charge state for activation
Soluble proteins
Carbonic anhydrase 2 29 1 11 (23.5 ± 9.3)% 10+
(CA)
Haemoglobin (HB) 64 4 11 α: (37.1 ± 14.4)%
β: (27.2 ± 15.6)%
16+
Alcohol dehydrogenase 148 4 10 (10.6 ± 4.5)% 25+
(ADH)
GroEL 801 14 3* (4.1 ± 2.4)% 70+
Membrane proteins
Bacteriorhodopsin 27 1 5 (17.3 ± 10.3)% 9+
Aquaporin Z (AqpZ) (detergent: C8E4) 99 4 3** (4.3 ± 2.4)% 16+
Aquaporin Z (AqpZ) (detergent: DDM) 99 4 5*** (4.9 ± 2.9)% 16+
*

In total 7 participants carried out native MS of GroEL; however, only 3 (all using Orbitrap UHMR instruments) generated interpretable fragment spectra from this precursor

**

7 participants successfully carried out native MS of AqpZ in C8E4; however, only 3 (2 with Orbitrap UHMRs, 1 with an FTICR) were able to generate interpretable fragment spectra

***

of the 5 datasets with fragment spectra, only 4 reported observation of the tetramer (3 Orbitrap UHMRs, 1 FTICR), while the fifth reported a trimer. A sixth dataset (acquired on a QTOF) reported trimeric AqpZ, but fragmentation was unsuccessful.

Similar native spectra across laboratories and instruments

Before addressing the fragmentation data, we wished to benchmark the native MS experiment itself. There have been previous efforts to accomplish this;6, 27 however, here we had access to a much larger dataset, with more proteins, more instruments, and more laboratories. To our knowledge, our work represents the first multi-laboratory study of both native MS and native top-down MS of a range of protein complexes. In the literature, there is broad consensus that native MS leads to lower charge states and a narrower charge state distribution compared to MS under denaturing conditions.11, 34 There have been previous efforts to study the relationship between protein mass, conformation, and charge state;3437 however, these studies have typically been carried out by single laboratories, and as such, it is unclear to what extent the observed native spectra depend on instrument- or operator-related factors. Figure 2 shows representative native spectra for three proteins from our set of samples, each acquired on three different instrument types. It is immediately apparent in this figure that the native charge state distributions are relatively consistent between different operators and instruments, although there were some minor instrument-to-instrument variations, particularly for aquaporin Z. This observation is likely be related to the gas-phase removal of the detergent molecules used for solubilisation. A plot of the observed average charge states per protein per instrument type can be found in Supplementary Figure S3. Our results indicate that this property is indeed mostly determined by protein-specific factors. This consistency also suggests that the different nano-electrospray sources produced ions with similar conformations and internal energies. Additionally, the correct stoichiometries were observed for the oligomeric species with different instruments, which supports the increasing adoption of native mass spectrometry as a versatile tool to determine protein complex stoichiometry. Peak width in the native spectra, however, differed significantly between instruments.

Figure 2. Comparison of native mass spectra of different instrument types.

Figure 2.

(a) CA, (b) GroEL, and (c) AqpZ (detergent: C8E4), acquired on FTICR, QTOF, and Orbitrap instruments (data from 5 participating laboratories in total). Additional native spectra can be found in Supplementary Figure S4.

Resolving power and cleavage coverage

For an accurate determination of the mass of a protein, resolving power – which is inversely correlated with peak width – is an important factor. In native MS, significant peak broadening is caused by the formation of nonspecific adducts with water molecules, sodium cations, etc. As a result, the effective resolving power (i.e., peak centre m/z divided by the full width at half maximum) can be orders of magnitude lower than the limit imposed by the mass analyser, and the ability to reduce the presence of the aforementioned clusters is a critical factor. This can be accomplished by thorough desalting prior to ionisation, controlled gas-phase activation for improved desolvation, or a combination of both. This controlled activation has been achieved in different ways, including careful control of pressures and voltages in the source region of the instrument, and – particularly in certain Orbitrap instruments – extended trapping under moderately activating conditions. Some of these strategies have been implemented in the commercially available instruments that were used in this study, and different instrument designs employ different methods. Further details on this topic have been described in several key publications.18, 3843 Figure 3a shows the effective resolving power obtained for each protein, broken down by instrument type. A clear trend for lower effective resolving power can be seen as the mass of the intact protein ion increases, indicating – as one might expect – that larger proteins tend to retain more water and/or salt during ionisation. We note that the theoretical resolving power limit imposed by the mass analyser decreases with m/z in Orbitrap and FTICR instruments. In practice, this factor adds only a minor contribution to the observed peak broadening compared to adduct formation.44 The type of instrument used still plays an important role, as Orbitrap platforms tended to generate narrower peaks than other instrument types, particularly for water-soluble proteins; however, this seems to be due to differences in desolvation prior to entry in the mass analyser. In some datasets, particularly those acquired with FTICR instruments, the obtained effective resolving power was higher for the membrane protein aquaporin Z compared to soluble proteins of similar mass (haemoglobin and ADH). A possible explanation is that participants erred on the side of using very ‘soft’ instrument settings for water-soluble proteins, which prevented unfolding, but also led to incomplete desolvation. The deliberate harsher tuning required for stripping detergent molecules from membrane proteins therefore may have also improved the removal of water and salt. However, two participants reported the observation of trimeric AqpZ – almost certainly formed through unintentional gas-phase dissociation of the complex during detergent removal – when using DDM to assist in solubilisation rather than the expected tetramer, illustrating the risk of overactivation. Conversely, the C8E4 detergent seems to be easier to remove, providing a greater ‘safety margin’ between detergent removal and disruption of the native complex. At the same time, it should be noted that while C8E4 is very good for native MS, it can have a denaturing effect in solution, and stability of complexes should be checked. One workflow that has been suggested is to use DDM for protein purification, followed by a detergent exchange to C8E4 just prior to native MS analysis.45

Figure 3. Evaluation of resolving power and cleavage coverage by protein and instrument type.

Figure 3.

(a) Boxplots of the obtained resolving power by protein and instrument type. Proteins are divided into soluble and membrane proteins, and then arranged by increasing precursor mass. Sample size n for the respective boxes: CA/QTOF 4; CA/Orbitrap 5; CA/FTICR 2; HB/QTOF 3; HB/Orbitrap 4; HB/FTICR 2; ADH/QTOF 3; ADH/Orbitrap 3; ADH/FTICR 2; GroEL/QTOF 2; GroEL/Orbitrap 3; GroEL/FTICR 2; BR/Orbitrap 3; BR/FTICR 2; AqpZ_DDM/Orbitrap 3; AqpZ_DDM/FTICR 1; AqpZ_C8E4/QTOF 1; AqpZ_C8E4/Orbitrap 4; AqpZ_C8E4/FTICR 2. (b) Boxplots of the obtained cleavage coverage by protein and instrument type. Sample size n for the respective boxes: CA/QTOF 4; CA/Orbitrap 5; CA/FTICR 2; HBα/QTOF 4; HBα/Orbitrap 5; HBα/FTICR 2; HBβ/QTOF 4; HBβ/Orbitrap 5; HBβ/FTICR 2; ADH/QTOF 4; ADH/Orbitrap 4; ADH/FTICR 2; GroEL/Orbitrap 3; GroEL/FTICR 2; BR/Orbitrap 3; BR/FTICR 2; AqpZ_DDM/Orbitrap 4; AqpZ_DDM/FTICR 1; AqpZ_C8E4/Orbitrap 2; AqpZ_C8E4/FTICR 1. Proteins are again divided into soluble and membrane proteins, but now sorted by increasing monomer mass. Grey dashed lines in both panels are added to guide the eye and show the trend for lower resolving power and cleavage coverage versus mass. Both boxplots show the minimum and maximum values as whiskers, the mean value is indicated by a dotted line, and the solid line represents the median. The boxes indicate 25% of the upper and lower quartile respectively. Individual data points are indicated as black dots.

Having benchmarked the native MS experiments as described above, we next moved on to the top-down fragmentation studies. Typical precursor charge states used are listed in Table 1, and the protocol can be found as Supplementary Protocol S1. Furthermore, our raw data files, which include instrument settings, are available (see Data Availability Statement below) an we encourage the interested reader to download and use spectra acquired on a similar instrument to their own as a starting point. While good resolving power and accurate precursor mass determination are important, a key to proteoform characterisation is the effective generation and identification of backbone fragments. To compare this factor between datasets, we calculated the cleavage coverage - defined here as the number of observed specific backbone cleavage sites divided by the total number of inter-residue bonds – as a global metric for each spectrum. We then plotted the cleavage coverage for each protein by instrument type, and this is shown in Figure 3b (average values across all datasets are reported in Table 1). As was observed for resolving power, a decrease in cleavage coverage with higher mass was apparent. This can be rationalised as follows: With increasing protein mass, the charge state (and thus the number of mobile protons) and the number of available vibrational modes both increase; however, the latter to a good approximation increases linearly with mass, while the former increases more slowly. A plot of the absolute number of observed cleavages per protein per instrument type can be found in Supplementary Figure S5, and this revealed a lower variability than the cleavage coverage percentage: Typically, around 40–50 cleavages were observed for most water-soluble proteins (fewer for GroEL), and around 20–30 for most membrane proteins. One factor that likely contributes to the lower number of cleavages (both relative and absolute) for membrane proteins is that detergent removal effectively ‘cools’ the ion in the gas phase. Furthermore, membrane proteins typically carry fewer charges than water-soluble proteins of the same mass due to micelle shielding46, 47 (particularly when C8E4 is used) meaning that the same voltage offset leads to lower laboratory-frame collision energies, while the number of mobile protons also decreases. Generally, no instrument type clearly outperformed the others in terms of obtained cleavage coverage. The exception to this was GroEL, for which only Orbitrap instruments, specifically ‘Q Exactive UHMR’ models optimised for the analysis of high-mass ions, yielded interpretable fragment spectra. Representative fragment spectra can be found in Supplementary Figure S6.

Preference for specific fragmentation sites

In addition to global cleavage coverage, the site-selectivity of backbone cleavages is important. Three alternative hypotheses can be postulated a priori: (1) cleavage sites are randomly distributed along the protein sequence; (2) there is a systematic, protein-dependent preference for certain cleavage sites; or (3) site preferences are instrument-dependent. In particular, for collision-based dissociation of native proteins on ‘Q Exactive’ Orbitrap instruments, site-selectivity has previously been linked to both residue-specific factors and to higher-order structure, depending on the specifics of the experiment.48, 49 Here, however, we once again had the opportunity to extend the previous work by including results from different instrument types. The histograms in Figure 4 show the number of times a cleavage site was observed for carbonic anhydrase, haemoglobin, and ADH, ranging from 0 (fragments resulting from this cleavage were never observed) to 11 (fragments resulting from this cleavage were observed by all labs, with all instruments). Equivalent figures for GroEL, bacteriorhodopsin, and aquaporin Z can be found in Supplementary Figure S7. Clear fragmentation ‘hotspots’ can be distinguished, and these are often associated with specific residue types, most notably C-terminal to aspartic acid, as indicated in the plots, and consistent with previous work. 48 Summed across our dataset, the fraction of possible cleavage events C-terminal to aspartic acid was notably higher than for other residue types and the number of ‘missed’ cleavages significantly lower, especially in the terminal regions (See Supplementary Figure S8 and Supplementary Table S5). This charge-remote fragmentation pathway is not dependent on mobile protons,50 explaining why it is common in the rather low-charge state ions created in native ESI. For the membrane proteins bacteriorhodopsin and aquaporin Z, fragmentation was preferentially observed in the transmembrane helices, in agreement with earlier work.51 Remarkably, some fragments were observed in 11 out of 11 datasets, indicating very high inter-laboratory reproducibility, and clearly showing that protein-specific factors are key for determining site-selectivity of fragmentation, regardless of the specific instrument used.

Figure 4. Evaluation of site-specific cleavages in native top-down.

Figure 4.

Number of observations of site-specific cleavage in (a) carbonic anhydrase, (b) alpha subunit and (c) beta subunit of haemoglobin, and (d) alcohol dehydrogenase. The residue-specific cleavage sites are highlighted in orange, and the most common cleavage regions in yellow. N-terminal acetylation of CA and ADH is indicated with a grey box. Equivalent figures for GroEL, bacteriorhodopsin, and aquaporin Z can be found in Supplementary Figure S7.

Another interesting pattern visible in both Figure 3b and Figure 4 is the greater propensity for fragmentation of the alpha subunit of haemoglobin than the beta subunit. Looking at the overall cleavage coverage values across all datasets, these were (37.1 ± 14.4)% and (27.2 ± 15.6)% for the alpha and beta subunit, respectively. While these confidence intervals do overlap (p = 0.13), it is noteworthy that in 10 out of 11 individual datasets, the cleavage coverage for the alpha subunit was higher, strongly suggesting that these observations indeed reflect an inherent difference in fragmentation propensity. From the cleavage site histograms in Figure 4, it is apparent that the main C-terminal fragmentation ‘hotspot’ in the alpha subunit extends further toward the N-terminus compared to the equivalent in the beta subunit, and it also contains significantly more sites where cleavage was observed in 10 or even 11 out of 11 datasets. This can be rationalised by comparing the sequence of both proteins, as in the alpha subunit this 30-residue region contains two proline and two aspartic acid residues, as opposed to one of each in the beta subunit. These regions are highlighted in the monomer structures in Supplementary Figure S9.

We also performed a small-scale comparison between the obtained cleavage coverage with natively ionised versus denatured precursors. For this, three participants, covering all three instrument types (Orbitrap, QTOF, and FTICR), performed top-down CID of denatured carbonic anhydrase and ADH. We selected these specific proteins for two reasons: (1) We already had datasets from all participating labs for each of them under native conditions, and (2) one is a monomer, while the other natively forms a tetramer. During dissociation after native ionisation, the latter ejects a highly-charged monomer, which subsequently fragments, whereas the former can only (partially or fully) unfold and then fragment. For the natively ionised CA and ADH, the average cleavage coverage values achieved by these three participants were (23.7 ± 7.0)% and (10.8 ± 4.0)%, respectively, i.e., in line with the (23.5 ± 9.3)% and (10.6 ± 4.5)% across datasets from all labs. With denatured precursors, the corresponding values were (25.0 ± 5.5)% and (8.8 ± 4.5)%. While our dataset is too small to draw general conclusions about native vs. denatured mode MS, these results indicate – perhaps surprisingly – that at least in CID-based top-down MS of ADH and CA, similar cleavage coverage values can be obtained from natively ionised as from denatured precursors. Charge-remote cleavage C-terminal to aspartic acid was less pronounced for the denatured precursors (see Supplementary Figure S10).

Data analysis in top-down protein mass spectrometry

Software tools used by participants in the analysis of precursor and fragment spectra are summarised in the Methods (Table 2). As MS measures mass-to-charge ratios, the charge states of ions must be determined to assign their mass. This process is quite different for a large, native-like precursor, compared to fragments that often have masses below 10 kDa, and both operations need to be carried out successfully. For fragments, modern MS instruments can often resolve the isotopic distribution, unlike for intact (native) proteins. As isotope peaks are spaced approximately 1 Da apart, measuring the spacing on the m/z axis often makes fragment charge state determination trivial. Algorithms that model the isotopic distribution based on the average composition of proteins can then estimate the monoisotopic mass of the fragment for comparing to candidate assignments, or the entire isotopic distribution can be matched to an in silico generated one.5256 In contrast, for intact protein precursors, conventional instruments are unable to resolve isotope peaks, and other algorithms for charge state deconvolution are required (vide infra).

A particular challenge in top-down MS is that the total signal intensity is divided over a large number of fragments.57 Therefore, fragment signal-to-noise ratios can be rather low, and spectral averaging is often required. Plots showing the degree of averaging used (both in terms of acquisition time and number of scans) in precursor and fragment spectra are shown in Supplementary Figure S11. While there was a slight tendency by some QTOF users to acquire a greater-than-average number of scans, both the mean and range of degrees of spectral averaging were quite comparable between QTOF, Orbitrap, and FTICR instruments. However, there was a clear trend toward less spectral averaging for intact precursors than for fragments (as the intensity is divided among significantly fewer signals). This procedure usually works well in vendor-specific software; however, freely available alternatives that work with vendor-neutral MS data formats such as mzML often do not perform as well as the proprietary software in our experience. An option that was used by a plurality of participants in this study was the MASH software series,5862 which can read in data from a relatively broad range of instruments, as well as generic formats that can be generated after spectral averaging using vendor-specific software. In top-down MS data analysis, it is important to carefully choose the minimum signal-to-noise level and mass error tolerance. Making these parameters too ‘loose’ will result in many false positive assignments, while making them too restrictive will lead to real fragment signals not being identified (i.e., false negatives). Figure 5 illustrates this concept, as well as the use of mass error distributions to assess the quality of fragment assignments. Visual comparison of the calculated (coloured circles) and observed (black solid traces) isotope distributions often makes it obvious whether an assignment is reliable, although the (correct) assignment of a y1798+ fragment is worth highlighting. This ion had a mass of nearly 20 kDa; therefore, its intensity was distributed over a large number of isotopologues, which led to a reduced signal-to-noise ratio, illustrating the particular risk of overly restrictive search settings in top-down protein analysis.

Figure 5. Fragment assignment and data quality assessment.

Figure 5.

(a) Fragment assignments after eTHRASH deconvolution in the MASH software. The blue box shows correct assignments for a range of charge states (coloured dots indicate the calculated isotope distributions). In the red box, incorrect automatically assigned fragments (the result of choosing an inappropriately low signal-to-noise threshold) are shown. Here, either manual rejection in combination with mass error filtering is needed to exclude these fragments, or reprocessing with a higher signal-to-noise threshold. (b) Fragment mass error distributions can be used to assess data quality.70, 71 For real fragment signals, a narrow mass error spread around zero ppm is expected (top panel; blue bars). Poor instrument calibration (middle) shifts the centre of this narrow distribution. Depending on the instrument, narrower distributions (e.g., ±2 – 5 ppm) might be expected. If data quality is poor and mainly noise is assigned (a strong indication of an unsuccessful experiment), evenly distributed error values over a broad range (bottom) are observed.

As researchers become more experienced with their particular instruments and data analysis software, they will generally develop their own criteria for a confident assignment. As a starting point, we recommend looking for three or more sequential isotope peaks with signal-to-noise ratios greater than 3 for fragments larger than 2 kDa, and two or more sequential peaks with signal-to-noise greater than 3 for smaller fragments (as the total intensity in this case is mostly concentrated in the first two isotopologues). Note that it would usually not be expected for the calibration of the instrument to fluctuate significantly within a narrow m/z range; therefore, the spacing between the isotope peaks should be close to the expected value (usually around 1.0034 Da, i.e., the mass difference between 12C and 13C). For very large fragments, spacing between isotope peaks might become less clear or disappear altogether; in this case, signal assignment has to rely purely on mass accuracy. While it is speculative at this point, it could also be imagined that some of the observations from the current study (e.g., typical number of backbone bonds cleaved, higher percentage of aspartic acid residues cleaved compared to other amino acids, missed cleavages mostly in the middle of the sequence) could be incorporated in scoring functions in the future.

In contrast to fragment signals, obtaining isotopic resolution is near-impossible for large, native-like species, as peak broadening due to noncovalent adducts limits resolving power values to a few thousand at best (as shown in Figure 3a). Different deconvolution strategies are thus required for precursors than fragments. Charge detection approaches allow direct measurement of an ion’s charge state;6366 however, this is not (yet) common and in practice it is usually necessary to observe multiple successive charge states and then apply a probabilistic deconvolution algorithm. While several algorithms and software packages exist, a majority of participants in the current study used the freely available UniDec software6769 for deconvolution of precursor spectra (see Supplementary Figure S1). This software provides a mass and intensity list for the main ion series as well as less abundant species, along with a confidence score for each. UniDec is not vendor-specific as a spectrum list can be used as input; however, it also supports upload of raw files from several different instrument manufacturers. UniDec is included in the latest iteration of the MASH software package (MASH Native).62 While the primary focus of this manuscript was on the use of known model proteins, the ultimate goal for many readers will be to identify unknown proteins and complexes. As a proof-of-concept experiment, we used MASH Native’s ‘Discovery Mode’ to analyse the data obtained by one of the participants with an Orbitrap UHMR (Laboratory 5). Except for GroEL, for which cleavage coverage was low (see Figure 3b) the correct proteins were consistently identified by the software (See Supplementary Table S3).

Discussion

Native and top-down mass spectrometry have each in their own right had a significant impact on our understanding of molecular biology, and the combination of both methods promises to open up new scientific vistas. At the same time, however, newcomers to the field experience significant barriers to entry, and more experienced practitioners regularly receive requests for advice on how to get started. Here, we have developed protocols for this type of experiment and demonstrated their implementation across a range of laboratories and instruments. Our post-study surveys (see Supplementary Figure S2) showed that (nano-)ESI and gas-phase activation of membrane proteins were experienced as more challenging than the experiments with water-soluble proteins. The exception is the water-soluble GroEL, for which, as previously mentioned, only high-mass modified Orbitrap instruments were able to induce monomer ejection and extensive backbone fragmentation. This can be explained by the large size and stability of this complex, which results in a high amount of energy being needed for monomer ejection and backbone fragmentation. The high-mass range Orbitraps seem to be more capable of supplying this amount of energy than other, less specialised instrument types.

The protocols described in this study will empower newcomers to the field(s) of native and/or top-down protein MS to successfully ionise monomeric and multimeric, water-soluble and membrane proteins in native mode across a wide mass range, and to obtain backbone fragmentation from these ions. In particular, as native separation methods such as capillary electrophoresis and size-exclusion chromatography are increasingly coupled with top-down approaches, these tools will drive new biological insights through high-throughput analysis of protein complexes. The example data shown in this report, as well as the high-level summaries of obtained resolving power and cleavage coverage per instrument type, provide a valuable benchmark to assess data quality for several easily available protein standards. One potential risk is overreliance on the overall cleavage coverage benchmarks, as in an attempt to reach the target coverage value, newcomers might feel encouraged to inappropriately loosen search parameters in their data analysis. Therefore, these numbers should be used together with the cleavage site frequency data: If similar site-selectivity is observed in conjunction with a similar overall coverage, one can be rather confident that a workflow is reasonably optimised. We also refer to the representative fragment spectra from different instrument types in Supplementary Figure S6 for further guidance.

Finally, we would like to provide some specific recommendations that we found to be helpful in the course of this study. (1) Perform thorough desalting (i.e., multiple rounds) starting with a higher ionic strength and going down to around 200 mM ammonium acetate for the working solution for nano-ESI. (2) Carefully optimise both the nano-ESI voltage and emitter position to obtain a stable spray; avoid trying to quickly acquire data with an unstable spray, as this will compromise data quality. (3) Invest time in setting up a robust data analysis workflow, and, especially in the early stages, manually verify deconvolution results and peak assignments; avoid ‘blind’ reliance on software tools without validation. (4) Avoid applying excessive in-source activation in an attempt to fully desolvate ions, as this can come at the expense of disruption of native protein-protein or protein-ligand interactions and deplete precursor signal. (5) Similarly, avoid the use of excessively high collision energy in top-down fragmentation, as this can lead to abundant formation of internal fragments (which are more challenging to assign reliably); furthermore, some data analysis software struggles if no intact protein signal is detectable in the fragment spectrum.

The combination of native MS with top-down fragmentation is able to provide uniquely detailed information on proteoform-specific biomolecular interactions. In particular, this combination is not only able to identify and characterise proteoforms, but also to link this information to protein structure, complex stoichiometry, and other noncovalent interaction partners. The benchmarks and guidance provided in this report will lead to a more widespread adoption of these methods by the scientific community in the coming years, so that they can be used to answer important biological questions.

Methods

The proteins used for this study were either bought from Sigma or (in the case of AqpZ) supplied by one of the participants. More information and catalogue numbers can be found in Supplementary Table S4. All participants prepared the protein samples according to the given protocol (see Supplementary Protocol S1). In general, all soluble proteins were dissolved in 200 mM ammonium acetate solution, and for membrane proteins the respective detergent was added in 2x critical micelle concentration. After desalting, all labs used nano-ESI with slightly different setups. The applied spray voltage ranged from 0.7 to 2 kV depending on the instrument setup and sample. Extended Data Table 1 shows the used instrument types and data analysis software per lab. More details can be found in Supplementary Table S7, including important tuning parameters for alcohol dehydrogenase (provided as an example and potential starting point) on all instruments used.

The evaluation of the collected data from all participants was performed by one participant with Microsoft Excel LTSC Professional Plus 2021 and MatLab R2022b.

Extended Data

Extended Data Table 1.

Instrument and software used by each participating laboratory

Lab MS instrument Data analysis software
1 Waters Synapt G2Si UniDec*, LcMsSpectator
2 Orbitrap Q-Exactive UHMR UniDec*, MASH Explorer
3 Bruker solariX FTICR DataAnalysis*, MASH Explorer
4 Waters Synapt G2Si MassLynx (manual data analysis)* , UniDec*
5a Bruker solariX FTICR DataAnalysis*, ClipsMS
5b Orbitrap Q-Exactive UFIMR BioPharma Finder*, ClipsMS
6 Orbitrap Q-Exactive UFIMR Uni Dec*, TDValidator
7 Orbitrap Q-Exactive UHMR UniDec*, Freestyle Extract, ProSight Lite
8a Waters Synapt XS UniDec*, MASH Explorer
8b LTQ Orbitrap XL UniDec*, MASH Explorer
9 Agilent 6545XT SLIM QtoF Agilent Bioconfirm 10*, deCharger, MASCOT, ProSight Lite

Supplementary Material

Supporting Information

Acknowledgements

We are grateful to all members of the Board of Directors of the Consortium for Top-Down Proteomics for valuable discussions. We also thank Boris Krichel and Sean McIlwain (University of Wisconsin-Madison) and Christian Hake (TU Darmstadt) for assistance with database searching. The Consortium for Top-Down Proteomics, a 501c3 non-profit corporation, is grateful for the generous support from Seer, Thermo Fisher Scientific, Bruker, SCIEX, Lilly, Pfizer, Newomics, and Agilent. M.T.M. thanks Alexander Makarov and Kyle Fort at Thermo Fisher Scientific for support on the UHMR Q-Exactive HF instrument.

This work was supported by NIH/NIGMS grant R35 GM128624 to M.T.M, NIH grant R01 GM125085 and S10 OD018475 to Y.G., and NIH/NIGMS grant P41 GM108569 to N.L.K., as well as NIH grant R01GM103479, R35GM145286, S10RR028893 and US Department of Energy grant DEFC02–02ER63421 to J.A.L. We acknowledge project award (10.46936/intm.proj.2019.51141/60006696) from the Environmental Molecular Sciences Laboratory, a DOE Office of Science User Facility sponsored by the Biological and Environmental Research program under Contract No. DE-AC0576RL01830 to M.Z. and J.W.W. and the Ruth L. Kirschstein National Research Service Award Program at NIH (GM007185) to C.L. We are grateful for funding by the Hessian Ministry for Science and Arts (HMWK) via the LOEWE project ‘TRABITA’, and by the Deutsche Forschungsgemeinschaft (grant numbers 461372424 and 524226614) to F.L.

The funders had no role in the study design, data collection and analysis, decision to publish or preparation of the manuscript.

Footnotes

Competing interests

N.L.K. is involved in entrepreneurial activities in top-down proteomics and consults for Thermo Fisher Scientific. P.O.D. is the founder and principal of Eastwoods Consulting, providing business advisory services to life science companies. The remaining authors declare no competing interests.

Data availability statement

The mass spectrometry raw data have been deposited to the ProteomeXchange Consortium via the PRIDE partner repository with the dataset identifier PXD047341.72, 73 The protein sequences can be accessed by the respective identifier (see Supplementary Table S4) via uniprot.com.

References

  • 1.Smith LM, Kelleher NL & Consortium for Top Down P Proteoform: a single term describing protein complexity. Nat Methods 10, 186–187 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Aebersold R et al. How many human proteoforms are there? Nat Chem Biol 14, 206–214 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Smith LM & Kelleher NL Proteoforms as the next proteomics currency. Science 359, 1106–1107 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Dang X et al. The first pilot project of the consortium for top-down proteomics: a status report. Proteomics 14, 1130–1140 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chen B, Brown KA, Lin Z & Ge Y Top-Down Proteomics: Ready for Prime Time? Anal Chem 90, 110–127 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Donnelly DP et al. Best practices and benchmarks for intact protein analysis for top-down mass spectrometry. Nat Methods 16, 587–594 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Srzentic K et al. Interlaboratory Study for Characterizing Monoclonal Antibodies by Top-Down and Middle-Down Mass Spectrometry. J Am Soc Mass Spectrom 31, 1783–1802 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Habeck T & Lermyte F Seeing the complete picture: proteins in top-down mass spectrometry. Essays Biochem (2022). [DOI] [PubMed] [Google Scholar]
  • 9.Brown KA, Melby JA, Roberts DS & Ge Y Top-down proteomics: challenges, innovations, and applications in basic and clinical research. Expert Rev Proteomics 17, 719–733 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Smith LM et al. The Human Proteoform Project: Defining the human proteome. Sci Adv 7, eabk0734 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Leney AC & Heck AJ Native Mass Spectrometry: What is in the Name? J Am Soc Mass Spectrom 28, 5–13 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Robinson CV Mass spectrometry: From plasma proteins to mitochondrial membranes. Proc Natl Acad Sci U S A 116, 2814–2820 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Tamara S, den Boer MA & Heck AJR High-Resolution Native Mass Spectrometry. Chem Rev 122, 7269–7326 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Bennett JL, Nguyen GTH & Donald WA Protein-Small Molecule Interactions in Native Mass Spectrometry. Chem Rev 122, 7327–7385 (2022). [DOI] [PubMed] [Google Scholar]
  • 15.Rogawski R & Sharon M Characterizing Endogenous Protein Complexes with Biological Mass Spectrometry. Chem Rev 122, 7386–7414 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Skinner OS et al. Top-down characterization of endogenous protein complexes with native proteomics. Nat Chem Biol 14, 36–41 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Ro SY et al. Native top-down mass spectrometry provides insights into the copper centers of membrane-bound methane monooxygenase. Nat Commun 10, 2675 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Gault J et al. Combining native and ‘omics’ mass spectrometry to identify endogenous ligands bound to membrane proteins. Nat Methods 17, 505–508 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Vimer S et al. Comparative Structural Analysis of 20S Proteasome Ortholog Protein Complexes by Native Mass Spectrometry. ACS Cent Sci 6, 573–588 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Schachner LF et al. Decoding the protein composition of whole nucleosomes with Nuc-MS. Nat Methods 18, 303–308 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Melani RD et al. Mapping Proteoforms and Protein Complexes From King Cobra Venom Using Both Denaturing and Native Top-down Proteomics. Mol Cell Proteomics 15, 2423–2434 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Lermyte F, Tsybin YO, O’Connor PB & Loo JA Top or Middle? Up or Down? Toward a Standard Lexicon for Protein Top-Down and Allied Mass Spectrometry Approaches. J Am Soc Mass Spectrom 30, 1149–1157 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Zhou M et al. Higher-order structural characterisation of native proteins and complexes by top-down mass spectrometry. Chem Sci 11, 12918–12936 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.VanAernum ZL et al. Rapid online buffer exchange for screening of proteins, protein complexes and cell lysates by native mass spectrometry. Nat Protoc 15, 1132–1157 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.McCabe JW et al. Implementing Digital-Waveform Technology for Extended m/z Range Operation on a Native Dual-Quadrupole FT-IM-Orbitrap Mass Spectrometer. J Am Soc Mass Spectrom 32, 2812–2820 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.LeDuc RD et al. ProForma: A Standard Proteoform Notation. J Proteome Res 17, 1321–1325 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Schachner LF et al. Standard Proteoforms and Their Complexes for Native Mass Spectrometry. J Am Soc Mass Spectrom (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Smith LM et al. A five-level classification system for proteoform identifications. Nat Methods (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Iacobucci C et al. First Community-Wide, Comparative Cross-Linking Mass Spectrometry Study. Anal Chem 91, 6953–6961 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Masson GR et al. Recommendations for performing, interpreting and reporting hydrogen deuterium exchange mass spectrometry (HDX-MS) experiments. Nat Methods 16, 595–602 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Allison TM et al. Software Requirements for the Analysis and Interpretation of Native Ion Mobility Mass Spectrometry Data. Anal Chem 92, 10881–10890 (2020). [DOI] [PubMed] [Google Scholar]
  • 32.Allison TM et al. Computational Strategies and Challenges for Using Native Ion Mobility Mass Spectrometry in Biophysics and Structural Biology. Anal Chem 92, 10872–10880 (2020). [DOI] [PubMed] [Google Scholar]
  • 33.Gabelica V et al. Recommendations for reporting ion mobility Mass Spectrometry measurements. Mass Spectrom Rev 38, 291–320 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Konijnenberg A, Butterer A & Sobott F Native ion mobility-mass spectrometry and related methods in structural biology. Biochim Biophys Acta 1834, 1239–1256 (2013). [DOI] [PubMed] [Google Scholar]
  • 35.Konermann L, Ahadi E, Rodriguez AD & Vahidi S Unraveling the mechanism of electrospray ionization. Anal Chem 85, 2–9 (2013). [DOI] [PubMed] [Google Scholar]
  • 36.Hall Z, Politis A, Bush MF, Smith LJ & Robinson CV Charge-state dependent compaction and dissociation of protein complexes: insights from ion mobility and molecular dynamics. J Am Chem Soc 134, 3429–3438 (2012). [DOI] [PubMed] [Google Scholar]
  • 37.Rolland AD, Biberic LS & Prell JS Investigation of Charge-State-Dependent Compaction of Protein Ions with Native Ion Mobility-Mass Spectrometry and Theory. J Am Soc Mass Spectrom 33, 369–381 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Sobott F, Hernandez H, McCammon MG, Tito MA & Robinson CV A tandem mass spectrometer for improved transmission and analysis of large macromolecular assemblies. Anal Chem 74, 1402–1407 (2002). [DOI] [PubMed] [Google Scholar]
  • 39.Sobott F, McCammon MG, Hernandez H & Robinson CV The flight of macromolecular complexes in a mass spectrometer. Philos Trans A Math Phys Eng Sci 363, 379–389; discussion 389–391 (2005). [DOI] [PubMed] [Google Scholar]
  • 40.Rose RJ, Damoc E, Denisov E, Makarov A & Heck AJ High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies. Nat Methods 9, 1084–1086 (2012). [DOI] [PubMed] [Google Scholar]
  • 41.van de Waterbeemd M et al. High-fidelity mass analysis unveils heterogeneity in intact ribosomal particles. Nat Methods 14, 283–286 (2017). [DOI] [PubMed] [Google Scholar]
  • 42.Fort KL et al. Expanding the structural analysis capabilities on an Orbitrap-based mass spectrometer for large macromolecular complexes. Analyst 143, 100–105 (2017). [DOI] [PubMed] [Google Scholar]
  • 43.McGee JP et al. Voltage Rollercoaster Filtering of Low-Mass Contaminants During Native Protein Analysis. J Am Soc Mass Spectrom 31, 763–767 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Snijder J, Rose RJ, Veesler D, Johnson JE & Heck AJ Studying 18 MDa virus assemblies with native mass spectrometry. Angew Chem Int Ed Engl 52, 4020–4023 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Behnke JS & Urner LH Emergence of mass spectrometry detergents for membrane proteomics. Anal Bioanal Chem 415, 3897–3909 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Borysik AJ, Hewitt DJ & Robinson CV Detergent release prolongs the lifetime of native-like membrane protein conformations in the gas-phase. J Am Chem Soc 135, 6078–6083 (2013). [DOI] [PubMed] [Google Scholar]
  • 47.Reading E et al. The role of the detergent micelle in preserving the structure of membrane proteins in the gas phase. Angew Chem Int Ed Engl 54, 4577–4581 (2015). [DOI] [PubMed] [Google Scholar]
  • 48.Ives AN et al. Using 10,000 Fragment Ions to Inform Scoring in Native Top-down Proteomics. J Am Soc Mass Spectrom 31, 1398–1409 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Lantz C et al. Native Top-Down Mass Spectrometry with Collisionally Activated Dissociation Yields Higher-Order Structure Information for Protein Complexes. J Am Chem Soc (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Paizs B & Suhai S Fragmentation pathways of protonated peptides. Mass Spectrom Rev 24, 508–548 (2005). [DOI] [PubMed] [Google Scholar]
  • 51.Skinner OS et al. Fragmentation of integral membrane proteins in the gas phase. Anal Chem 86, 4627–4634 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Horn DM, Zubarev RA & McLafferty FW Automated reduction and interpretation of high resolution electrospray mass spectra of large molecules. J Am Soc Mass Spectrom 11, 320–332 (2000). [DOI] [PubMed] [Google Scholar]
  • 53.Zamdborg L et al. ProSight PTM 2.0: improved protein identification and characterization for top down mass spectrometry. Nucleic Acids Res 35, W701–706 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Mayampurath AM et al. DeconMSn: a software tool for accurate parent ion monoisotopic mass determination for tandem mass spectra. Bioinformatics 24, 1021–1023 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Li L & Tian Z Interpreting raw biological mass spectra using isotopic mass-to-charge ratio and envelope fingerprinting. Rapid Commun Mass Spectrom 27, 1267–1277 (2013). [DOI] [PubMed] [Google Scholar]
  • 56.Liu X et al. Deconvolution and database search of complex tandem mass spectra of intact proteins: a combinatorial approach. Mol Cell Proteomics 9, 2772–2782 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Compton PD, Zamdborg L, Thomas PM & Kelleher NL On the scalability and requirements of whole protein mass spectrometry. Anal Chem 83, 6868–6874 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Guner H et al. MASH Suite: a user-friendly and versatile software interface for high-resolution mass spectrometry data interpretation and visualization. J Am Soc Mass Spectrom 25, 464–470 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Cai W et al. MASH Suite Pro: A Comprehensive Software Tool for Top-Down Proteomics. Mol Cell Proteomics 15, 703–714 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Wu Z et al. MASH Explorer: A Universal Software Environment for Top-Down Proteomics. J Proteome Res 19, 3867–3876 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.McIlwain SJ et al. Enhancing Top-Down Proteomics Data Analysis by Combining Deconvolution Results through a Machine Learning Strategy. J Am Soc Mass Spectrom 31, 1104–1113 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Larson EJ et al. MASH Native: A Unified Solution for Native Top-Down Proteomics Data Processing. Bioinformatics (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Kafader JO et al. Multiplexed mass spectrometry of individual ions improves measurement of proteoforms and their complexes. Nat Methods (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Kafader JO et al. Individual Ion Mass Spectrometry Enhances the Sensitivity and Sequence Coverage of Top-Down Mass Spectrometry. J Proteome Res 19, 1346–1350 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Worner TP et al. Resolving heterogeneous macromolecular assemblies by Orbitrap-based single-particle charge detection mass spectrometry. Nat Methods 17, 395–398 (2020). [DOI] [PubMed] [Google Scholar]
  • 66.McGee JP et al. Isotopic Resolution of Protein Complexes up to 466 kDa Using Individual Ion Mass Spectrometry. Anal Chem 93, 2723–2727 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Marty MT et al. Bayesian deconvolution of mass and ion mobility spectra: from binary interactions to polydisperse ensembles. Anal Chem 87, 4370–4376 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Reid DJ et al. MetaUniDec: High-Throughput Deconvolution of Native Mass Spectra. J Am Soc Mass Spectrom 30, 118–127 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Marty MT A Universal Score for Deconvolution of Intact Protein and Native Electrospray Mass Spectra. Anal Chem 92, 4395–4401 (2020). [DOI] [PubMed] [Google Scholar]
  • 70.Park J et al. Informed-Proteomics: open-source software package for top-down proteomics. Nat Methods 14, 909–914 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Zhou M, Pasa-Tolic L & Stenoien DL Profiling of Histone Post-Translational Modifications in Mouse Brain with High-Resolution Top-Down Mass Spectrometry. J Proteome Res 16, 599–608 (2017). [DOI] [PubMed] [Google Scholar]
  • 72.Deutsch EW et al. The ProteomeXchange consortium in 2020: enabling ‘big data’ approaches in proteomics. Nucleic Acids Res 48, D1145–D1152 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Perez-Riverol Y et al. The PRIDE database resources in 2022: a hub for mass spectrometry-based proteomics evidences. Nucleic Acids Res 50, D543–D552 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supporting Information

Data Availability Statement

The mass spectrometry raw data have been deposited to the ProteomeXchange Consortium via the PRIDE partner repository with the dataset identifier PXD047341.72, 73 The protein sequences can be accessed by the respective identifier (see Supplementary Table S4) via uniprot.com.

RESOURCES