Table 2.
Number of patients | n=2183 |
---|---|
Number of cases in which the scope was used before this case, n (%) | |
≤2 | 1921 (88.0) |
>2 up to 5 | 262 (12.0) |
Scope size (Fr), n (%) | |
≥8 | 1386 (63.5) |
7.5 | 797 (36.5) |
General anesthesia, n (%) | 2056 (94.2) |
Ureteral access sheath, n (%) | 2138 (97.9) |
Sterilization, n (%)* | |
Flash sterilization | 1749 (90.2) |
Plasma | 1171 (60.5) |
Gas | 5 (0.3) |
Peracetic acid | 392 (20.3) |
ETO | 339 (17.5) |
Scope testing results, n (%) | |
Problem in mechanism noted before/during case, but continued and completed case | 163 (7.5) |
Upward deflection affected | 35 (1.6) |
Downward deflection affected | 142 (6.5) |
Image quality affected on white balancing | 104 (4.7) |
Working channel affected | 3 (0.13) |
Scope changed during case, n (%) | 85 (3.9) |
Image quality poor with flickering | 81 (3.7) |
Deflection mechanism stopped | 25 (1.1) |
Problems passing basket/laser in working channel | 12 (0.5) |
Case abandoned due to scope malfunction | 2 (0.09) |
Case completed using alternate scope for the 85 cases | 84 (98.8) |
Laser, n (%) | |
Thulium fiber laser | 994 (45.9) |
Holmium laser without + with Moses technology | 1171+18 (54.1) |
Basketing, n (%) | 818 (37.5) |
Laser time, median (IQR) | 21 (15–39) |
Total operation time, median (IQR) | 56 (40–66) |
*More than one modality possible (combination of modalities). IQR=Interquartile range, ETO=Ethylene oxide