5 |
Delayed or improper dissection of cortices |
Usage of inappropriate tools can cause difficulties |
Use curved-end forceps |
11 |
The pellet does not seem red or is covered with a white layer |
The pellet was not properly resuspended |
Mix the pellet from Step 9 completely with 15% (wt/vol) dextran–DPBS by pipetting |
|
The pellet does not appear at the bottom of tube |
The concentration of dextran solution was incorrect |
Confirm that the dextran completely dissolves. Do not exceed the correct final volume |
12 |
Debris is too close to the pellet. It is difficult to get a clean pellet |
Dextran was not mixed thoroughly with brain lysate or a small volume of dextran solution was used |
Increase the volume of dextran and mix the pellet thoroughly by pipetting. If necessary, wipe out debris from the walls of the tube with a Kimwipe |
14 |
The pellet becomes trapped inside the pipette tip |
The microvessels can easily adhere to the surface of plastic |
Pre-wet the pipette tip by pipetting 1× DPBS in and out repeatedly |
17 |
The pellet is very small or invisible |
The microvessels remain in the cell strainer or BSA was not added to the medium |
Use enough volume to retrieve the microvessels from the strainer. Add BSA (0.5% (wt/vol)) to pellet the microvessels |