Table I.

MEAN AND STANDARD DEVIATION (S.D.) OF ESTIMATED SOUND SPEED IN THE HOMOGENEOUS SPECKLE SIMULATION. VALUES WERE COMPUTED AROUND THE SIX POINTS HIGHLIGHTED IN FIG. 3 USING 5-MM-BY-5-MM KERNELS. UNITS ARE IN M/S.

x = −1 cm
 z = 4 cm
 x = 0 cm z = 4 cm  x = 1cm z = 4 cm x = −1 cm
 z = 5 cm
 x = 0 cm z = 5 cm  x = 1 cm z = 5 cm
 Direct Mean
 S.d.
 1542.2
 5.4
 1542.1
 4.8
 1540.6
 5.9
 1544.1
 10.1
 1540.2
 7.4
 1542.2
 10.1
 1D Mean
 S.d.
 1542.2
 3.0
 1542.1
 1.9
 1540.9
 3.9
 1542.1
 5.4
 1540.9
 2.6
 1542.1
 5.4
 2D Mean
 S.d.
 1541.9
 2.0
 1542.1
 1.8
 1541.2
 2.9
 1541.8
 2.9
 1541.0
 2.4
 1541.7
 2.6